A Wireless and Wearable Multimodal Sensor to Non-Invasively Monitor Transabdominal Placental Oxygen Saturation and Maternal Physiological Signals.

IF 4.9 3区 工程技术 Q1 CHEMISTRY, ANALYTICAL
Thien Nguyen, Soongho Park, Asma Sodager, Jinho Park, Dahiana M Gallo, Guoyang Luo, Roberto Romero, Amir Gandjbakhche
{"title":"A Wireless and Wearable Multimodal Sensor to Non-Invasively Monitor Transabdominal Placental Oxygen Saturation and Maternal Physiological Signals.","authors":"Thien Nguyen, Soongho Park, Asma Sodager, Jinho Park, Dahiana M Gallo, Guoyang Luo, Roberto Romero, Amir Gandjbakhche","doi":"10.3390/bios14100481","DOIUrl":null,"url":null,"abstract":"<p><p>Poor placental development and placental defects can lead to adverse pregnancy outcomes such as pre-eclampsia, fetal growth restriction, and stillbirth. This study introduces two sensors, which use a near-infrared spectroscopy (NIRS) technique to measure placental oxygen saturation transabdominally. The first one, an NIRS sensor, is a wearable device consisting of multiple NIRS channels. The second one, a Multimodal sensor, which is an upgraded version of the NIRS sensor, is a wireless and wearable device, integrating a motion sensor and multiple NIRS channels. A pilot clinical study was conducted to assess the feasibility of the two sensors in measuring transabdominal placental oxygenation in 36 pregnant women (n = 12 for the NIRS sensor and n = 24 for the Multimodal sensor). Among these subjects, 4 participants had an uncomplicated pregnancy, and 32 patients had either maternal pre-existing conditions/complications, neonatal complications, and/or placental pathologic abnormalities. The study results indicate that the patients with maternal complicated conditions (69.5 ± 5.4%), placental pathologic abnormalities (69.4 ± 4.9%), and neonatal complications (68.0 ± 5.1%) had statistically significantly lower transabdominal placental oxygenation levels than those with an uncomplicated pregnancy (76.0 ± 4.4%) (<i>F</i> (3,104) = 6.6, <i>p</i> = 0.0004). Additionally, this study shows the capability of the Multimodal sensor in detecting the maternal heart rate and respiratory rate, fetal movements, and uterine contractions. These findings demonstrate the feasibility of the two sensors in the real-time continuous monitoring of transabdominal placental oxygenation to detect at-risk pregnancies and guide timely clinical interventions, thereby improving pregnancy outcomes.</p>","PeriodicalId":48608,"journal":{"name":"Biosensors-Basel","volume":"14 10","pages":""},"PeriodicalIF":4.9000,"publicationDate":"2024-10-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11506160/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosensors-Basel","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3390/bios14100481","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Poor placental development and placental defects can lead to adverse pregnancy outcomes such as pre-eclampsia, fetal growth restriction, and stillbirth. This study introduces two sensors, which use a near-infrared spectroscopy (NIRS) technique to measure placental oxygen saturation transabdominally. The first one, an NIRS sensor, is a wearable device consisting of multiple NIRS channels. The second one, a Multimodal sensor, which is an upgraded version of the NIRS sensor, is a wireless and wearable device, integrating a motion sensor and multiple NIRS channels. A pilot clinical study was conducted to assess the feasibility of the two sensors in measuring transabdominal placental oxygenation in 36 pregnant women (n = 12 for the NIRS sensor and n = 24 for the Multimodal sensor). Among these subjects, 4 participants had an uncomplicated pregnancy, and 32 patients had either maternal pre-existing conditions/complications, neonatal complications, and/or placental pathologic abnormalities. The study results indicate that the patients with maternal complicated conditions (69.5 ± 5.4%), placental pathologic abnormalities (69.4 ± 4.9%), and neonatal complications (68.0 ± 5.1%) had statistically significantly lower transabdominal placental oxygenation levels than those with an uncomplicated pregnancy (76.0 ± 4.4%) (F (3,104) = 6.6, p = 0.0004). Additionally, this study shows the capability of the Multimodal sensor in detecting the maternal heart rate and respiratory rate, fetal movements, and uterine contractions. These findings demonstrate the feasibility of the two sensors in the real-time continuous monitoring of transabdominal placental oxygenation to detect at-risk pregnancies and guide timely clinical interventions, thereby improving pregnancy outcomes.

用于无创监测经腹胎盘血氧饱和度和产妇生理信号的无线可穿戴多模态传感器。
胎盘发育不良和胎盘缺陷可导致不良妊娠结局,如先兆子痫、胎儿生长受限和死胎。本研究介绍了两种利用近红外光谱(NIRS)技术经腹测量胎盘血氧饱和度的传感器。第一种是近红外传感器,是一种由多个近红外通道组成的可穿戴设备。第二种是多模态传感器,它是近红外传感器的升级版,是一种无线可穿戴设备,集成了运动传感器和多个近红外光谱通道。为了评估这两种传感器测量经腹胎盘氧合的可行性,我们对 36 名孕妇(近红外传感器为 12 人,多模态传感器为 24 人)进行了试点临床研究。在这些受试者中,4 名受试者为无并发症妊娠,32 名患者患有母体原有疾病/并发症、新生儿并发症和/或胎盘病理异常。研究结果表明,与无并发症妊娠的患者(76.0 ± 4.4%)相比,患有产妇并发症(69.5 ± 5.4%)、胎盘病理异常(69.4 ± 4.9%)和新生儿并发症(68.0 ± 5.1%)的患者经腹胎盘氧合水平明显较低(F (3,104) = 6.6,P = 0.0004)。此外,该研究还显示了多模态传感器检测母体心率和呼吸频率、胎动和子宫收缩的能力。这些研究结果证明了这两种传感器在实时连续监测经腹胎盘氧合方面的可行性,可用于检测高危妊娠并指导及时的临床干预,从而改善妊娠结局。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Biosensors-Basel
Biosensors-Basel Biochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
6.60
自引率
14.80%
发文量
983
审稿时长
11 weeks
期刊介绍: Biosensors (ISSN 2079-6374) provides an advanced forum for studies related to the science and technology of biosensors and biosensing. It publishes original research papers, comprehensive reviews and communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced. Electronic files and software regarding the full details of the calculation or experimental procedure, if unable to be published in a normal way, can be deposited as supplementary electronic material.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信