Nurullah Dag, Gulnur Igci, Fatma Hilal Yagin, Muhammed Salih Hanci, Ramazan Kutlu
{"title":"Interobserver Reproducibility of Ultrasound Attenuation Imaging Technology in Liver Fat Quantification.","authors":"Nurullah Dag, Gulnur Igci, Fatma Hilal Yagin, Muhammed Salih Hanci, Ramazan Kutlu","doi":"10.1002/jcu.23877","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aims to investigate the interobserver variability in the quantitative assessment of liver fat content using ultrasound attenuation imaging technology (USAT).</p><p><strong>Methods: </strong>This prospective, single-center study included 96 adult patients who were either diagnosed with or suspected of having metabolic dysfunction-associated steatotic liver disease. Independent observers, blinded to each other's assessments, evaluated hepatic steatosis visually and through USAT measurements. Separate measurements were taken at five intercostal and subcostal sites, and the median values of these measurements were recorded. The correlation between USAT measurements and visual steatosis grades was examined using Spearman's correlation test. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were used to evaluate the interobserver variability of USAT measurements.</p><p><strong>Results: </strong>Interobserver agreement for USAT measurements was excellent for the intercostal examination and good for the subcostal examination (p < 0.001). Body mass index did not significantly affect the level of interobserver agreement. Interobserver variability in Bland-Altman plots of USAT measurements was within the 95% limits of agreement. USAT measurements correlated very strongly with the visual degree of hepatic steatosis, both intercostal and subcostal (p < 0.001). USAT measurements were also significantly different between different visual degrees of hepatic steatosis (p < 0.001).</p><p><strong>Conclusion: </strong>In the assessment of hepatic steatosis, USAT measurements obtained from the intercostal space showed excellent agreement in terms of interobserver reproducibility.</p>","PeriodicalId":15386,"journal":{"name":"Journal of Clinical Ultrasound","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-10-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Ultrasound","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jcu.23877","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ACOUSTICS","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study aims to investigate the interobserver variability in the quantitative assessment of liver fat content using ultrasound attenuation imaging technology (USAT).
Methods: This prospective, single-center study included 96 adult patients who were either diagnosed with or suspected of having metabolic dysfunction-associated steatotic liver disease. Independent observers, blinded to each other's assessments, evaluated hepatic steatosis visually and through USAT measurements. Separate measurements were taken at five intercostal and subcostal sites, and the median values of these measurements were recorded. The correlation between USAT measurements and visual steatosis grades was examined using Spearman's correlation test. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were used to evaluate the interobserver variability of USAT measurements.
Results: Interobserver agreement for USAT measurements was excellent for the intercostal examination and good for the subcostal examination (p < 0.001). Body mass index did not significantly affect the level of interobserver agreement. Interobserver variability in Bland-Altman plots of USAT measurements was within the 95% limits of agreement. USAT measurements correlated very strongly with the visual degree of hepatic steatosis, both intercostal and subcostal (p < 0.001). USAT measurements were also significantly different between different visual degrees of hepatic steatosis (p < 0.001).
Conclusion: In the assessment of hepatic steatosis, USAT measurements obtained from the intercostal space showed excellent agreement in terms of interobserver reproducibility.
期刊介绍:
The Journal of Clinical Ultrasound (JCU) is an international journal dedicated to the worldwide dissemination of scientific information on diagnostic and therapeutic applications of medical sonography.
The scope of the journal includes--but is not limited to--the following areas: sonography of the gastrointestinal tract, genitourinary tract, vascular system, nervous system, head and neck, chest, breast, musculoskeletal system, and other superficial structures; Doppler applications; obstetric and pediatric applications; and interventional sonography. Studies comparing sonography with other imaging modalities are encouraged, as are studies evaluating the economic impact of sonography. Also within the journal''s scope are innovations and improvements in instrumentation and examination techniques and the use of contrast agents.
JCU publishes original research articles, case reports, pictorial essays, technical notes, and letters to the editor. The journal is also dedicated to being an educational resource for its readers, through the publication of review articles and various scientific contributions from members of the editorial board and other world-renowned experts in sonography.