Evaluation of automatic tube current modulation in a CT scanner using a customised homogeneous phantom.

IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Urshella Hishaam, Jeyasingam Jeyasugiththan, Sameera Viswakula, D M Satharasinghe, T Amalaraj, M Costa, A M C Kumarihami, Aruna Pallewatte, Steve Peterson
{"title":"Evaluation of automatic tube current modulation in a CT scanner using a customised homogeneous phantom.","authors":"Urshella Hishaam, Jeyasingam Jeyasugiththan, Sameera Viswakula, D M Satharasinghe, T Amalaraj, M Costa, A M C Kumarihami, Aruna Pallewatte, Steve Peterson","doi":"10.1088/2057-1976/ad857a","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective.</i>The introduction of automatic tube current modulation (ATCM) has resulted in complex relationships between scanner parameters, patient body habitus, radiation dose, and image quality. ATCM adjusts tube current based on x-ray attenuation variations in the scan region, and overall patient dose depends on a combination of factors. This work aims to develop mathematical models that predict CT radiation dose and image noise in terms of attenuating diameter and all relevant scanner parameters.<i>Approach.</i>A homogenous phantom, equipped with the features to conduct discrete and continuous adaption tests, was developed to model ATCM in a Philips CT scanner. Scanner parameters were varied based on theoretical dose relationships, and a MATLAB script was developed to extract data from DICOM images. R statistical software was employed for data analysis, plotting, and regression modelling.<i>Main Results.</i>Phantom data provided the following insights: Median tube current decreased by 81% as tube potential varied from 80 kVp to 140 kVp. Doubling the DoseRight Index (DRI) from 12 to 24, at 24 cm diameter, produced a 294% increase in mA and a 46% decrease in noise. Mean mA increased by 53% whilst mean noise increased by 5.7% as helical pitch increased from 0.6 to 0.925. Changing rotation time from 0.33s to 0.75s gave a 56% reduction in mean mA and no change in image noise. Increasing detector collimation (<i>n × T</i>) resulted in higher tube currents and lower output image noise values, as<i>n</i>and<i>T</i>were varied independently. Interpreting these results to apply transformations relevant to each independent variable produced models for tube current and noise with adjusted R-squared values of 0.965 and 0.912, respectively.<i>Significance.</i>The models developed more accurately predict radiation dose and image quality for specific patients and scanner settings. They provide imaging professionals with a practical tool to optimize scan protocols according to patient diameters and clinical objectives.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":"10 6","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-10-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Physics & Engineering Express","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/2057-1976/ad857a","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective.The introduction of automatic tube current modulation (ATCM) has resulted in complex relationships between scanner parameters, patient body habitus, radiation dose, and image quality. ATCM adjusts tube current based on x-ray attenuation variations in the scan region, and overall patient dose depends on a combination of factors. This work aims to develop mathematical models that predict CT radiation dose and image noise in terms of attenuating diameter and all relevant scanner parameters.Approach.A homogenous phantom, equipped with the features to conduct discrete and continuous adaption tests, was developed to model ATCM in a Philips CT scanner. Scanner parameters were varied based on theoretical dose relationships, and a MATLAB script was developed to extract data from DICOM images. R statistical software was employed for data analysis, plotting, and regression modelling.Main Results.Phantom data provided the following insights: Median tube current decreased by 81% as tube potential varied from 80 kVp to 140 kVp. Doubling the DoseRight Index (DRI) from 12 to 24, at 24 cm diameter, produced a 294% increase in mA and a 46% decrease in noise. Mean mA increased by 53% whilst mean noise increased by 5.7% as helical pitch increased from 0.6 to 0.925. Changing rotation time from 0.33s to 0.75s gave a 56% reduction in mean mA and no change in image noise. Increasing detector collimation (n × T) resulted in higher tube currents and lower output image noise values, asnandTwere varied independently. Interpreting these results to apply transformations relevant to each independent variable produced models for tube current and noise with adjusted R-squared values of 0.965 and 0.912, respectively.Significance.The models developed more accurately predict radiation dose and image quality for specific patients and scanner settings. They provide imaging professionals with a practical tool to optimize scan protocols according to patient diameters and clinical objectives.

使用定制均质模型评估 CT 扫描仪中的自动导管电流调制。
目的:自动电子管电流调节(ATCM)的引入导致扫描仪参数、患者体型、辐射剂量和图像质量之间的关系变得复杂。ATCM 会根据扫描区域的 X 射线衰减变化调整管电流,而患者的总体剂量则取决于多种因素。这项工作旨在建立数学模型,根据衰减直径和所有相关的扫描仪参数预测 CT 辐射剂量和图像噪声。根据理论剂量关系改变扫描仪参数,并开发 MATLAB 脚本从 DICOM 图像中提取数据。采用 R 统计软件进行数据分析、绘图和回归建模:当电子管电位从 80 kVp 变化到 140 kVp 时,电子管电流中值下降了 81%。直径为 24 厘米时,剂量正确指数(DRI)从 12 倍增加到 24 倍,电流增加了 294%,噪声减少了 46%。当螺旋间距从 0.6 增加到 0.925 时,平均毫安增加了 53%,而平均噪音增加了 5.7%。将旋转时间从 0.33 秒改为 0.75 秒后,平均毫安降低了 56%,图像噪声没有变化。增加探测器准直度(n × T)会导致更高的电子管电流和更低的输出图像噪声值,而 n 和 T 是独立变化的。通过对这些结果进行解释,应用与每个自变量相关的变换,得出的管电流和噪声模型的调整 R 方值分别为 0.965 和 0.912。这些模型为成像专业人员提供了一种实用工具,可根据患者直径和临床目标优化扫描方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Biomedical Physics & Engineering Express
Biomedical Physics & Engineering Express RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
2.80
自引率
0.00%
发文量
153
期刊介绍: BPEX is an inclusive, international, multidisciplinary journal devoted to publishing new research on any application of physics and/or engineering in medicine and/or biology. Characterized by a broad geographical coverage and a fast-track peer-review process, relevant topics include all aspects of biophysics, medical physics and biomedical engineering. Papers that are almost entirely clinical or biological in their focus are not suitable. The journal has an emphasis on publishing interdisciplinary work and bringing research fields together, encompassing experimental, theoretical and computational work.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信