Are some neutral liners more neutral than others? An ex vivo morphological analysis of acetabular liners classified as "neutral".

IF 2.5 2区 医学 Q1 ORTHOPEDICS
José Á Ochoa, Perttu S Neuvonen, Jari Hyttinen, Jari Viik, Antti P Eskelinen
{"title":"Are some neutral liners more neutral than others? An ex vivo morphological analysis of acetabular liners classified as \"neutral\".","authors":"José Á Ochoa, Perttu S Neuvonen, Jari Hyttinen, Jari Viik, Antti P Eskelinen","doi":"10.2340/17453674.2024.41946","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong> In contemporary total hip replacement (THR), dislocation is one of the most common complications. At our institution, the cause of an increase in the dislocation rate was recently reported to be reduced head coverage of a newly introduced neutral liner. We therefore aimed to ascertain whether differences exist in articulating head coverage between the various neutral liners used in contemporary THR. A secondary aim was to utilize coverage measurements to develop a new liner coverage classification.</p><p><strong>Methods: </strong> The articulating head coverage of 25 modular neutral polyethylene liners used in 6 uncemented cup designs from 4 major manufacturers was evaluated. The measurements were performed in a metrology laboratory and a mathematical model was developed to calculate coverage of the articulating surfaces. Further, 1 \"elevated rim\" liner and 1 \"face changing liner\" were included to develop a new liner coverage classification.</p><p><strong>Results: </strong> The articulating head coverage among the studied liners ranged from 167.7° to 194.8°, corresponding to a variation of 27.1°. The variations with different cup and head sizes within each design were smaller (from 1.0° to 5.6°) than those between different designs. Each of the liner designs offered distinct coverage, even though they were all classified as neutral. Based on measurements, a set of descriptive parameters to discriminate different liners in terms of coverage was created.</p><p><strong>Conclusion: </strong> We showed that all neutral liners are not equal - instead, they clearly varied in terms of their actual coverage design. We suggest our set of descriptive parameters called \"hemispheric coverage index values\" be used in discriminating the differences in liner coverage.</p>","PeriodicalId":6916,"journal":{"name":"Acta Orthopaedica","volume":"95 ","pages":"586-591"},"PeriodicalIF":2.5000,"publicationDate":"2024-10-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11468231/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Orthopaedica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2340/17453674.2024.41946","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background and purpose:  In contemporary total hip replacement (THR), dislocation is one of the most common complications. At our institution, the cause of an increase in the dislocation rate was recently reported to be reduced head coverage of a newly introduced neutral liner. We therefore aimed to ascertain whether differences exist in articulating head coverage between the various neutral liners used in contemporary THR. A secondary aim was to utilize coverage measurements to develop a new liner coverage classification.

Methods:  The articulating head coverage of 25 modular neutral polyethylene liners used in 6 uncemented cup designs from 4 major manufacturers was evaluated. The measurements were performed in a metrology laboratory and a mathematical model was developed to calculate coverage of the articulating surfaces. Further, 1 "elevated rim" liner and 1 "face changing liner" were included to develop a new liner coverage classification.

Results:  The articulating head coverage among the studied liners ranged from 167.7° to 194.8°, corresponding to a variation of 27.1°. The variations with different cup and head sizes within each design were smaller (from 1.0° to 5.6°) than those between different designs. Each of the liner designs offered distinct coverage, even though they were all classified as neutral. Based on measurements, a set of descriptive parameters to discriminate different liners in terms of coverage was created.

Conclusion:  We showed that all neutral liners are not equal - instead, they clearly varied in terms of their actual coverage design. We suggest our set of descriptive parameters called "hemispheric coverage index values" be used in discriminating the differences in liner coverage.

中性内衬是否比其他内衬更中性?对归类为 "中性 "的髋臼衬垫进行体外形态学分析。
背景和目的:在当代全髋关节置换术(THR)中,脱位是最常见的并发症之一。在我院,最近有报道称脱位率增加的原因是新引进的中性衬垫的关节头覆盖面减小。因此,我们旨在确定当代 THR 中使用的各种中性衬垫在关节头覆盖方面是否存在差异。我们的另一个目的是利用覆盖率测量结果来制定新的衬垫覆盖率分类: 方法:对 4 家主要制造商生产的 6 种非骨水泥髋臼杯设计中使用的 25 种模块化中性聚乙烯衬垫的关节头覆盖范围进行了评估。测量在计量实验室进行,并建立了一个数学模型来计算铰接面的覆盖率。此外,还纳入了 1 个 "凸缘 "衬垫和 1 个 "换面衬垫",以制定新的衬垫覆盖率分类: 结果:所研究的衬垫的衔接头覆盖范围从 167.7° 到 194.8°,相差 27.1°。与不同设计之间的差异(从 1.0°到 5.6°)相比,每种设计中不同杯体和杯头大小的差异较小。尽管每种衬垫设计都被归类为中性设计,但它们都提供了不同的覆盖范围。根据测量结果,我们创建了一套描述性参数来区分不同衬垫的覆盖范围: 我们的研究表明,所有的中性衬垫都不相同,相反,它们在实际覆盖设计方面存在明显差异。我们建议在区分衬垫覆盖率的差异时使用我们的一套描述性参数,即 "半球覆盖率指数值"。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Acta Orthopaedica
Acta Orthopaedica 医学-整形外科
CiteScore
6.40
自引率
8.10%
发文量
105
审稿时长
4-8 weeks
期刊介绍: Acta Orthopaedica (previously Acta Orthopaedica Scandinavica) presents original articles of basic research interest, as well as clinical studies in the field of orthopedics and related sub disciplines. Ever since the journal was founded in 1930, by a group of Scandinavian orthopedic surgeons, the journal has been published for an international audience. Acta Orthopaedica is owned by the Nordic Orthopaedic Federation and is the official publication of this federation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信