Jiajia Wang, Yan Shen, Xingtao Zhou, Zhiqiang Yu, Jiaxu Hong, Qihua Le
{"title":"Evaluation of tear film function by Oculus Keratograph 5M and IDRA ocular surface analyser.","authors":"Jiajia Wang, Yan Shen, Xingtao Zhou, Zhiqiang Yu, Jiaxu Hong, Qihua Le","doi":"10.1007/s10792-024-03262-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this study was to evaluate the function of tear film with Oculus Keratograph 5M (Oculus K5M) and IDRA ocular surface analyser (IDRA), analyse their consistency and explore the potential of IDRA in the diagnosis of dry eye disease (DED).</p><p><strong>Methods: </strong>This cross-sectional study enrolled 36 participants (DED group, 14 eyes; non-DED group, 22 eyes). The parameters of tear film function, including the first noninvasive breakup time (fNIBUT), average NIBUT (aNIBUT), tear meniscus height (TMH), lipid layer thickness (LLT), lipid layer colour (LLC), lipid layer uniformity (LLU), morphology of meibomian glands (MGs) and MG loss, were obtained with Oculus K5M and IDRA. The consistency of parameter measurements between the two devices was evaluated.</p><p><strong>Results: </strong>All the parameters except LLT, which can be measured only by IDRA, were not significantly different between the two instruments in DED eyes. However, IDRA reported lower values of fNIBUT, aNIBUT and TMH as well as higher MG loss scores in non-DED eyes than Oculus K5M did (p < 0.001, < 0.001, = 0.002, and = 0.002, respectively). Further regression analysis revealed that aNIBUT and LLT measured by IDRA were the optimal parameters for diagnosing DED (OR = 0.567 and 0.845, p = 0.057 and 0.043, respectively), and their combination had the strongest diagnostic potential (AUC = 0.841, sensitivity = 85.7%, and specificity = 77.3%).</p><p><strong>Conclusion: </strong>As a user-friendly noninvasive device, the tear film function parameters measured by IDRA were highly consistent with those measured by Oculus K5M in DED patients. The combination of aNIBUT and LLT measured by IDRA had the best diagnostic accuracy for DED.</p>","PeriodicalId":14473,"journal":{"name":"International Ophthalmology","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10792-024-03262-y","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The aim of this study was to evaluate the function of tear film with Oculus Keratograph 5M (Oculus K5M) and IDRA ocular surface analyser (IDRA), analyse their consistency and explore the potential of IDRA in the diagnosis of dry eye disease (DED).
Methods: This cross-sectional study enrolled 36 participants (DED group, 14 eyes; non-DED group, 22 eyes). The parameters of tear film function, including the first noninvasive breakup time (fNIBUT), average NIBUT (aNIBUT), tear meniscus height (TMH), lipid layer thickness (LLT), lipid layer colour (LLC), lipid layer uniformity (LLU), morphology of meibomian glands (MGs) and MG loss, were obtained with Oculus K5M and IDRA. The consistency of parameter measurements between the two devices was evaluated.
Results: All the parameters except LLT, which can be measured only by IDRA, were not significantly different between the two instruments in DED eyes. However, IDRA reported lower values of fNIBUT, aNIBUT and TMH as well as higher MG loss scores in non-DED eyes than Oculus K5M did (p < 0.001, < 0.001, = 0.002, and = 0.002, respectively). Further regression analysis revealed that aNIBUT and LLT measured by IDRA were the optimal parameters for diagnosing DED (OR = 0.567 and 0.845, p = 0.057 and 0.043, respectively), and their combination had the strongest diagnostic potential (AUC = 0.841, sensitivity = 85.7%, and specificity = 77.3%).
Conclusion: As a user-friendly noninvasive device, the tear film function parameters measured by IDRA were highly consistent with those measured by Oculus K5M in DED patients. The combination of aNIBUT and LLT measured by IDRA had the best diagnostic accuracy for DED.
期刊介绍:
International Ophthalmology provides the clinician with articles on all the relevant subspecialties of ophthalmology, with a broad international scope. The emphasis is on presentation of the latest clinical research in the field. In addition, the journal includes regular sections devoted to new developments in technologies, products, and techniques.