{"title":"[Effect of Whole-body Continuous Scanning Speed of Bone Scintigraphy on the Detectability of Vertebral Lesions].","authors":"Tetsumaru Kobayashi, Hajime Ichikawa, Toyohiro Kato, Hirotaka Nagura, Syohei Michino, Yoshinao Misu, Hideki Shimada, Yoichi Watanabe","doi":"10.6009/jjrt.2024-1494","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate the influence of the scanning speed of whole-body scans on the detectability of positive vertebral bone images in bone scintigraphy.</p><p><strong>Methods: </strong>We used SIM<sup>2</sup> bone phantom to obtain planar images equivalent to scanning speeds of 15, 17, and 20 cm/min. Receiver operating characteristic (ROC) analysis to evaluate lesion detectability and average count (Ct)/pixel, contrast ratio, and contrast-to-noise ratio (CNR) of the normal vertebral body and the simulated tumor site were measured.</p><p><strong>Results: </strong>The average area under the ROC curves (AUC) was 0.936, 0.929, and 0.915 at speeds of 15, 17, and 20 cm/min, respectively. The average AUC at 20 cm/min was significantly lower than that at 15 cm/min (p<0.05) . However, no other significant differences were found (p=0.448, 0.139). The average Ct/pixel and CNR decreased at 15, 17, and 20 cm/min. The contrast ratio did not change.</p><p><strong>Conclusion: </strong>The results showed that increasing the scan speed from 15 cm/min to 17 cm/min had no effect on the detection of vertebral lesions. Thus, it is possible to reduce the scan time, albeit slightly.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"1198-1204"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1494","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/3 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study aimed to evaluate the influence of the scanning speed of whole-body scans on the detectability of positive vertebral bone images in bone scintigraphy.
Methods: We used SIM2 bone phantom to obtain planar images equivalent to scanning speeds of 15, 17, and 20 cm/min. Receiver operating characteristic (ROC) analysis to evaluate lesion detectability and average count (Ct)/pixel, contrast ratio, and contrast-to-noise ratio (CNR) of the normal vertebral body and the simulated tumor site were measured.
Results: The average area under the ROC curves (AUC) was 0.936, 0.929, and 0.915 at speeds of 15, 17, and 20 cm/min, respectively. The average AUC at 20 cm/min was significantly lower than that at 15 cm/min (p<0.05) . However, no other significant differences were found (p=0.448, 0.139). The average Ct/pixel and CNR decreased at 15, 17, and 20 cm/min. The contrast ratio did not change.
Conclusion: The results showed that increasing the scan speed from 15 cm/min to 17 cm/min had no effect on the detection of vertebral lesions. Thus, it is possible to reduce the scan time, albeit slightly.