[Effect of Whole-body Continuous Scanning Speed of Bone Scintigraphy on the Detectability of Vertebral Lesions].

Nihon Hoshasen Gijutsu Gakkai zasshi Pub Date : 2024-11-20 Epub Date: 2024-10-03 DOI:10.6009/jjrt.2024-1494
Tetsumaru Kobayashi, Hajime Ichikawa, Toyohiro Kato, Hirotaka Nagura, Syohei Michino, Yoshinao Misu, Hideki Shimada, Yoichi Watanabe
{"title":"[Effect of Whole-body Continuous Scanning Speed of Bone Scintigraphy on the Detectability of Vertebral Lesions].","authors":"Tetsumaru Kobayashi, Hajime Ichikawa, Toyohiro Kato, Hirotaka Nagura, Syohei Michino, Yoshinao Misu, Hideki Shimada, Yoichi Watanabe","doi":"10.6009/jjrt.2024-1494","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to evaluate the influence of the scanning speed of whole-body scans on the detectability of positive vertebral bone images in bone scintigraphy.</p><p><strong>Methods: </strong>We used SIM<sup>2</sup> bone phantom to obtain planar images equivalent to scanning speeds of 15, 17, and 20 cm/min. Receiver operating characteristic (ROC) analysis to evaluate lesion detectability and average count (Ct)/pixel, contrast ratio, and contrast-to-noise ratio (CNR) of the normal vertebral body and the simulated tumor site were measured.</p><p><strong>Results: </strong>The average area under the ROC curves (AUC) was 0.936, 0.929, and 0.915 at speeds of 15, 17, and 20 cm/min, respectively. The average AUC at 20 cm/min was significantly lower than that at 15 cm/min (p<0.05) . However, no other significant differences were found (p=0.448, 0.139). The average Ct/pixel and CNR decreased at 15, 17, and 20 cm/min. The contrast ratio did not change.</p><p><strong>Conclusion: </strong>The results showed that increasing the scan speed from 15 cm/min to 17 cm/min had no effect on the detection of vertebral lesions. Thus, it is possible to reduce the scan time, albeit slightly.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"1198-1204"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1494","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/3 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study aimed to evaluate the influence of the scanning speed of whole-body scans on the detectability of positive vertebral bone images in bone scintigraphy.

Methods: We used SIM2 bone phantom to obtain planar images equivalent to scanning speeds of 15, 17, and 20 cm/min. Receiver operating characteristic (ROC) analysis to evaluate lesion detectability and average count (Ct)/pixel, contrast ratio, and contrast-to-noise ratio (CNR) of the normal vertebral body and the simulated tumor site were measured.

Results: The average area under the ROC curves (AUC) was 0.936, 0.929, and 0.915 at speeds of 15, 17, and 20 cm/min, respectively. The average AUC at 20 cm/min was significantly lower than that at 15 cm/min (p<0.05) . However, no other significant differences were found (p=0.448, 0.139). The average Ct/pixel and CNR decreased at 15, 17, and 20 cm/min. The contrast ratio did not change.

Conclusion: The results showed that increasing the scan speed from 15 cm/min to 17 cm/min had no effect on the detection of vertebral lesions. Thus, it is possible to reduce the scan time, albeit slightly.

[骨闪烁成像全身连续扫描速度对椎骨病变可探测性的影响]。
目的:本研究旨在评估全身扫描速度对骨闪烁成像中椎骨阳性图像可探测性的影响:方法:我们使用 SIM2 骨模型获得相当于 15、17 和 20 厘米/分钟扫描速度的平面图像。通过接收者操作特征(ROC)分析评估病变可探测性,并测量正常椎体和模拟肿瘤部位的平均计数(Ct)/像素、对比度和对比噪声比(CNR):速度为 15、17 和 20 厘米/分钟时,ROC 曲线下的平均面积(AUC)分别为 0.936、0.929 和 0.915。20 厘米/分钟时的平均 AUC 明显低于 15 厘米/分钟时(pConclusions:结果表明,将扫描速度从 15 厘米/分钟提高到 17 厘米/分钟对椎体病变的检测没有影响。因此,尽管扫描时间略有缩短,但还是有可能的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信