Multimodal image-guided surgical robot versus 3D-printed template for brachytherapy of malignant tumours in the skull base and deep facial region: a clinical comparative study.

N Jin, F Meng, L Zhu, L Xing, Q Lin, H Zhang
{"title":"Multimodal image-guided surgical robot versus 3D-printed template for brachytherapy of malignant tumours in the skull base and deep facial region: a clinical comparative study.","authors":"N Jin, F Meng, L Zhu, L Xing, Q Lin, H Zhang","doi":"10.1016/j.ijom.2024.09.002","DOIUrl":null,"url":null,"abstract":"<p><p>This study compared a multimodal image-guided robot and three-dimensionally (3D) printed templates for implanting iodine-125 (I<sup>125</sup>) radioactive seeds in patients with malignant tumours in the skull base and deep facial region. Seventeen patients who underwent I<sup>125</sup> radioactive seed implantation between December 2018 and December 2019 were included. The operation time, intraoperative blood loss, and accuracy of seed implantation were compared between the multimodal image-guided robot-assisted implantation (experimental) group (n = 7) and 3D-printed template-assisted implantation (control) group (n = 10). In total, 291 seeds were implanted in the experimental group and 436 in the control group; the mean error of seed implantation accuracy was 1.95 ± 0.13 mm and 1.90 ± 0.08 mm, respectively (P = 0.309). The preparation time was 26.13 ± 5.28 min in the experimental group and 0 min in the control group, while the average operation time was 34.44 ± 6.39 min versus 43.70 ± 6.06 min, respectively. The intraoperative blood loss was 4.96 ± 1.76 ml (experimental) versus 8.97 ± 2.99 ml (control) (P = 0.123). Multimodal image-guided robot-assisted I<sup>125</sup> radioactive seed implantation met the clinical requirements for treating malignant tumours in the skull base and deep facial regions.</p>","PeriodicalId":94053,"journal":{"name":"International journal of oral and maxillofacial surgery","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of oral and maxillofacial surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.ijom.2024.09.002","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This study compared a multimodal image-guided robot and three-dimensionally (3D) printed templates for implanting iodine-125 (I125) radioactive seeds in patients with malignant tumours in the skull base and deep facial region. Seventeen patients who underwent I125 radioactive seed implantation between December 2018 and December 2019 were included. The operation time, intraoperative blood loss, and accuracy of seed implantation were compared between the multimodal image-guided robot-assisted implantation (experimental) group (n = 7) and 3D-printed template-assisted implantation (control) group (n = 10). In total, 291 seeds were implanted in the experimental group and 436 in the control group; the mean error of seed implantation accuracy was 1.95 ± 0.13 mm and 1.90 ± 0.08 mm, respectively (P = 0.309). The preparation time was 26.13 ± 5.28 min in the experimental group and 0 min in the control group, while the average operation time was 34.44 ± 6.39 min versus 43.70 ± 6.06 min, respectively. The intraoperative blood loss was 4.96 ± 1.76 ml (experimental) versus 8.97 ± 2.99 ml (control) (P = 0.123). Multimodal image-guided robot-assisted I125 radioactive seed implantation met the clinical requirements for treating malignant tumours in the skull base and deep facial regions.

用于颅底和面部深部恶性肿瘤近距离放射治疗的多模态图像引导手术机器人与三维打印模板:一项临床比较研究。
本研究比较了多模态图像引导机器人和三维(3D)打印模板在颅底和面部深部恶性肿瘤患者中植入碘-125(I125)放射性粒子的情况。17名患者在2018年12月至2019年12月期间接受了I125放射性粒子植入术。比较了多模态图像引导机器人辅助植入组(实验组)(n = 7)和3D打印模板辅助植入组(对照组)(n = 10)的手术时间、术中失血量和种子植入的准确性。实验组共植入 291 颗种子,对照组共植入 436 颗种子;种子植入精度的平均误差分别为 1.95 ± 0.13 毫米和 1.90 ± 0.08 毫米(P = 0.309)。实验组的准备时间为(26.13 ± 5.28)分钟,对照组为 0 分钟;平均手术时间分别为(34.44 ± 6.39)分钟和(43.70 ± 6.06)分钟。术中失血量为 4.96 ± 1.76 毫升(实验组)对 8.97 ± 2.99 毫升(对照组)(P = 0.123)。多模态图像引导机器人辅助 I125 放射性粒子植入术符合治疗颅底和面部深部恶性肿瘤的临床要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信