{"title":"Assessment of different treatment alternatives for patients with total maxillectomy.","authors":"Ilgın Arı, Gülin Acar, Emre Tosun, Mehmet Muhtaroğulları","doi":"10.1111/jopr.13947","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the biomechanical advantages of combining zygoma and partial subperiosteal implants for maxillary reconstruction in severely atrophic maxillae, offering potential solutions to the challenges posed by traditional reconstruction methods.</p><p><strong>Material and methods: </strong>A finite element analysis used a craniofacial model simulating a totally resected maxilla. Four treatment scenarios (SCs) were evaluated: SC-1, the quad zygoma approach; SC-2, two zygoma implants (ZIs) and a one-piece subperiosteal implant (SI); SC-3, two ZI and a two-piece SI; and SC-4, four ZI and a one-piece SI. Stress distributions on bone, implants, abutments, and metal frameworks were compared under occlusal forces.</p><p><strong>Results: </strong>For the simulated bone regions under vertical and oblique forces, SC-4 values were higher than SC-1, while SC-2 and SC-3 values were comparable for p<sub>max</sub> and p<sub>min</sub>. In addition, SC-2 and SC-3 values were lower than those of SC-1 and SC-4. The most balanced von Mises stress values on the ZIsand were observed in SC-3 under vertical and oblique forces. Furthermore, lower von Mises stress values on the abutments were seen in SC-3 under oblique and vertical forces. Although the lower von Mises stress values on the metal frameworks at the lateral incisor and first premolar side were seen in SC-3, lower von Mises values were observed in SC-4 in the first molar region.</p><p><strong>Conclusions: </strong>Overall, this study suggests that combining zygoma and partial subperiosteal implants may be a promising approach for reconstructing severely atrophic maxillae. These implants may offer improved biomechanical properties compared to ZIs alone.</p>","PeriodicalId":49152,"journal":{"name":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","volume":null,"pages":null},"PeriodicalIF":3.4000,"publicationDate":"2024-09-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jopr.13947","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To assess the biomechanical advantages of combining zygoma and partial subperiosteal implants for maxillary reconstruction in severely atrophic maxillae, offering potential solutions to the challenges posed by traditional reconstruction methods.
Material and methods: A finite element analysis used a craniofacial model simulating a totally resected maxilla. Four treatment scenarios (SCs) were evaluated: SC-1, the quad zygoma approach; SC-2, two zygoma implants (ZIs) and a one-piece subperiosteal implant (SI); SC-3, two ZI and a two-piece SI; and SC-4, four ZI and a one-piece SI. Stress distributions on bone, implants, abutments, and metal frameworks were compared under occlusal forces.
Results: For the simulated bone regions under vertical and oblique forces, SC-4 values were higher than SC-1, while SC-2 and SC-3 values were comparable for pmax and pmin. In addition, SC-2 and SC-3 values were lower than those of SC-1 and SC-4. The most balanced von Mises stress values on the ZIsand were observed in SC-3 under vertical and oblique forces. Furthermore, lower von Mises stress values on the abutments were seen in SC-3 under oblique and vertical forces. Although the lower von Mises stress values on the metal frameworks at the lateral incisor and first premolar side were seen in SC-3, lower von Mises values were observed in SC-4 in the first molar region.
Conclusions: Overall, this study suggests that combining zygoma and partial subperiosteal implants may be a promising approach for reconstructing severely atrophic maxillae. These implants may offer improved biomechanical properties compared to ZIs alone.
期刊介绍:
The Journal of Prosthodontics promotes the advanced study and practice of prosthodontics, implant, esthetic, and reconstructive dentistry. It is the official journal of the American College of Prosthodontists, the American Dental Association-recognized voice of the Specialty of Prosthodontics. The journal publishes evidence-based original scientific articles presenting information that is relevant and useful to prosthodontists. Additionally, it publishes reports of innovative techniques, new instructional methodologies, and instructive clinical reports with an interdisciplinary flair. The journal is particularly focused on promoting the study and use of cutting-edge technology and positioning prosthodontists as the early-adopters of new technology in the dental community.