Behavior of biometric parameters during clinical hours

IF 1.2 4区 医学 Q3 OPHTHALMOLOGY
M. Atar-Vardi , B. Dubinsky-Pertzov , A. Shemer , I. Hecht , E. Pras , A. Einan-Lifshitz , L. Or
{"title":"Behavior of biometric parameters during clinical hours","authors":"M. Atar-Vardi ,&nbsp;B. Dubinsky-Pertzov ,&nbsp;A. Shemer ,&nbsp;I. Hecht ,&nbsp;E. Pras ,&nbsp;A. Einan-Lifshitz ,&nbsp;L. Or","doi":"10.1016/j.jfo.2024.104296","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To evaluate and quantify variation of biometric parameters - axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT) and white-to-white (WTW).</div></div><div><h3>Methods</h3><div>A population-based retrospective cohort study was performed on patients who underwent a biometry test prior to cataract surgery using the IOL Master 700 (Carl <em>Zeiss</em> Meditec, Jena, Germany) between the years 2017–2021. Differences in these parameters were evaluated between scans executed at different times of the day.</div></div><div><h3>Results</h3><div>21,975 examinations of 8611 patients were included. Mean age was 70.50<!--> <!-->±<!--> <!-->12.56<!--> <!-->years. The mean time of the biometry exams was 10:52<!--> <!-->±<!--> <!-->1:23 AM. Measurements of AL, ACD, CCT and WTW were tested hourly and grouped between 7:00–9:00 AM and 12:00–03:00 PM. All the parameters showed a diurnal increase with a significance of <em>P</em> <!-->&lt;<!--> <!-->0.001 (AL from 23.64<!--> <!-->±<!--> <!-->1.5 to 24.01<!--> <!-->±<!--> <!-->1.76<!--> <!-->mm; ACD from 3.29<!--> <!-->±<!--> <!-->0.67 to 3.35<!--> <!-->±<!--> <!-->0.64<!--> <!-->mm; CCT from 0.52<!--> <!-->±<!--> <!-->0.04 to 0.53<!--> <!-->±<!--> <!-->0.04<!--> <!-->μm and WTW from11.83<!--> <!-->±<!--> <!-->0.46 to 11.90<!--> <!-->±<!--> <!-->0.51<!--> <!-->mm). The most significant change was seen in AL. The difference between time groups remained significant in a generalized linear mixed model (<em>P</em> <!-->&lt;<!--> <!-->0.001).</div></div><div><h3>Conclusions</h3><div>There are fluctuations in AL, ACD, CCT, WTW measurements during office hours. These results raise questions about the significance of timing of the biometry exam and the effect on the ELP calculation.</div></div><div><h3>But</h3><div>Évaluer et quantifier la variation des paramètres biométriques - longueur axiale, Profondeur de la chambre antérieure, Épaisseur centrale de la cornée et les distances blanc à blanc.</div></div><div><h3>Méthodes</h3><div>Nous avons réalisé une étude de cohorte rétrospective, portant sur une population de patients ayant bénéficié de tests biométriques utilisant IOI master 700 (Carl Zeiss Meditec, Jena, Allemagne) avant chirurgie de la cataracte, entre 2017 et 2021. Les écarts entre les paramètres mentionnés ont été évalués au moyen de tests réalisés à différents moments de la journée.</div></div><div><h3>Résultats</h3><div>21 975 examens de 8611 patients ont été inclus. L’âge moyen était de 70,50<!--> <!-->±<!--> <!-->12,56 ans. L’heure moyenne de réalisation de l’examen de biométrie était de 10h52<!--> <!-->±<!--> <!-->1h23. Les mesures de la longueur axiale, de la profondeur de la chambre antérieure, de l’épaisseur cornéenne centrale et du blanc à blanc ont été testées chaque heure et regroupées entre 7h00–9h00 et 12h00–15h00. Tous les paramètres ont montré une augmentation diurne avec un écart de <em>p</em> <!-->&lt;<!--> <!-->0,001 (longueur axiale de 23,64<!--> <!-->±<!--> <!-->1,5 à 24,01<!--> <!-->±<!--> <!-->1,76<!--> <!-->mm; Profondeur de la chambre antérieure de 3,29<!--> <!-->±<!--> <!-->0,67 à 3,35<!--> <!-->±<!--> <!-->0,64<!--> <!-->mm ; épaisseur centrale de la cornée de 0,52<!--> <!-->±<!--> <!-->0,04 à 0,53<!--> <!-->±<!--> <!-->0,04<!--> <!-->μm et les distances blanc à blanc de 11,83<!--> <!-->±<!--> <!-->0,46 à 11,90<!--> <!-->±<!--> <!-->0,51<!--> <!-->mm). Le changement le plus significatif a été observé dans longueur axiale. La différence entre les groupes horaires est restée significative dans un modèle linéaire généralisé mixte (<em>p</em> <!-->&lt;<!--> <!-->0,001).</div></div><div><h3>Conclusions</h3><div>Il existe des fluctuations dans les mesures de la longueur axiale, de la profondeur de la chambre antérieure, de l’épaisseur cornéenne centrale et les distances blanc à blanc effectuées pendant les heures de bureau. Ces résultats soulèvent des questions sur l’importance de l’horaire de l’examen biométrique et son incidence sur le calcul de la position effective de la lentille.</div></div>","PeriodicalId":14777,"journal":{"name":"Journal Francais D Ophtalmologie","volume":"47 9","pages":"Article 104296"},"PeriodicalIF":1.2000,"publicationDate":"2024-09-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal Francais D Ophtalmologie","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0181551224002419","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

To evaluate and quantify variation of biometric parameters - axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT) and white-to-white (WTW).

Methods

A population-based retrospective cohort study was performed on patients who underwent a biometry test prior to cataract surgery using the IOL Master 700 (Carl Zeiss Meditec, Jena, Germany) between the years 2017–2021. Differences in these parameters were evaluated between scans executed at different times of the day.

Results

21,975 examinations of 8611 patients were included. Mean age was 70.50 ± 12.56 years. The mean time of the biometry exams was 10:52 ± 1:23 AM. Measurements of AL, ACD, CCT and WTW were tested hourly and grouped between 7:00–9:00 AM and 12:00–03:00 PM. All the parameters showed a diurnal increase with a significance of P < 0.001 (AL from 23.64 ± 1.5 to 24.01 ± 1.76 mm; ACD from 3.29 ± 0.67 to 3.35 ± 0.64 mm; CCT from 0.52 ± 0.04 to 0.53 ± 0.04 μm and WTW from11.83 ± 0.46 to 11.90 ± 0.51 mm). The most significant change was seen in AL. The difference between time groups remained significant in a generalized linear mixed model (P < 0.001).

Conclusions

There are fluctuations in AL, ACD, CCT, WTW measurements during office hours. These results raise questions about the significance of timing of the biometry exam and the effect on the ELP calculation.

But

Évaluer et quantifier la variation des paramètres biométriques - longueur axiale, Profondeur de la chambre antérieure, Épaisseur centrale de la cornée et les distances blanc à blanc.

Méthodes

Nous avons réalisé une étude de cohorte rétrospective, portant sur une population de patients ayant bénéficié de tests biométriques utilisant IOI master 700 (Carl Zeiss Meditec, Jena, Allemagne) avant chirurgie de la cataracte, entre 2017 et 2021. Les écarts entre les paramètres mentionnés ont été évalués au moyen de tests réalisés à différents moments de la journée.

Résultats

21 975 examens de 8611 patients ont été inclus. L’âge moyen était de 70,50 ± 12,56 ans. L’heure moyenne de réalisation de l’examen de biométrie était de 10h52 ± 1h23. Les mesures de la longueur axiale, de la profondeur de la chambre antérieure, de l’épaisseur cornéenne centrale et du blanc à blanc ont été testées chaque heure et regroupées entre 7h00–9h00 et 12h00–15h00. Tous les paramètres ont montré une augmentation diurne avec un écart de p < 0,001 (longueur axiale de 23,64 ± 1,5 à 24,01 ± 1,76 mm; Profondeur de la chambre antérieure de 3,29 ± 0,67 à 3,35 ± 0,64 mm ; épaisseur centrale de la cornée de 0,52 ± 0,04 à 0,53 ± 0,04 μm et les distances blanc à blanc de 11,83 ± 0,46 à 11,90 ± 0,51 mm). Le changement le plus significatif a été observé dans longueur axiale. La différence entre les groupes horaires est restée significative dans un modèle linéaire généralisé mixte (p < 0,001).

Conclusions

Il existe des fluctuations dans les mesures de la longueur axiale, de la profondeur de la chambre antérieure, de l’épaisseur cornéenne centrale et les distances blanc à blanc effectuées pendant les heures de bureau. Ces résultats soulèvent des questions sur l’importance de l’horaire de l’examen biométrique et son incidence sur le calcul de la position effective de la lentille.
门诊时间生物计量参数的变化
目的评估并量化生物测量参数--轴长(AL)、前房深度(ACD)、中央角膜厚度(CCT)和白比白(WTW)--的变化。方法对2017-2021年间白内障手术前使用IOL Master 700(卡尔蔡司医疗技术公司,德国耶拿,德国)进行生物测量测试的患者进行基于人群的回顾性队列研究。结果共纳入8611名患者的21975次检查。平均年龄为 70.50±12.56 岁。生物测量检查的平均时间为上午 10:52 ± 1:23。AL、ACD、CCT 和 WTW 的测量每小时进行一次,分组时间为上午 7:00-9:00 和下午 12:00-03:00。所有参数都显示出昼夜间的增加,显著性为 P < 0.001(AL 从 23.64 ± 1.5 mm 增加到 24.01 ± 1.76 mm;ACD 从 3.29 ± 0.67 mm 增加到 3.35 ± 0.64 mm;CCT 从 0.52 ± 0.04 μm 增加到 0.53 ± 0.04 μm;WTW 从 11.83 ± 0.46 mm 增加到 11.90 ± 0.51 mm)。AL 的变化最为明显。结论上班时间的 AL、ACD、CCT 和 WTW 测量值存在波动。方法我们对 2017 年至 2021 年期间白内障手术前使用 IOI master 700(卡尔蔡司医疗技术公司,德国耶拿,Jena)进行生物测量测试的患者进行了一项回顾性队列研究。通过在一天中的不同时间进行测试,评估了上述参数之间的偏差。结果共纳入了来自 8611 名患者的 21975 次检查。平均年龄为 70.50±12.56 岁。生物测量检查的平均时间为 10:52 ± 1:23。轴长、前房深度、角膜中央厚度和白内障的测量每小时进行一次,并在上午 7 点至 9 点和中午 12 点至下午 3 点之间进行分组。所有参数均显示出昼夜变化,差异为 p < 0.001(轴长从 23.64 ± 1.5 mm 增加到 24.01 ± 1.76 mm;前房深度从 3.29 ± 0.67 mm 增加到 3.35 ± 0.64 mm;角膜中央厚度从 0.52 ± 0.04 μm 增加到 0.53 ± 0.04 μm;白对白距离从 11.83 ± 0.46 mm 增加到 11.90 ± 0.51 mm)。轴向长度的变化最为明显。在广义线性混合模型中,每小时组间的差异仍然显著(p <0.001)。结论在上班时间进行的轴长、前房深度、角膜中央厚度和白对白距离的测量存在波动。这些结果提出了关于生物计量检查时间安排的重要性及其对有效晶状体位置计算的影响的问题。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.10
自引率
8.30%
发文量
317
审稿时长
49 days
期刊介绍: The Journal français d''ophtalmologie, official publication of the French Society of Ophthalmology, serves the French Speaking Community by publishing excellent research articles, communications of the French Society of Ophthalmology, in-depth reviews, position papers, letters received by the editor and a rich image bank in each issue. The scientific quality is guaranteed through unbiased peer-review, and the journal is member of the Committee of Publication Ethics (COPE). The editors strongly discourage editorial misconduct and in particular if duplicative text from published sources is identified without proper citation, the submission will not be considered for peer review and returned to the authors or immediately rejected.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信