{"title":"Determination of the force transmission error in a single-sinker magnetic suspension densimeter due to the fluid-specific effect and its correction for use with gas mixtures containing oxygen","authors":"Daniel Lozano-Martín, María E. Mondéjar, José J. Segovia, César R. Chamorro","doi":"arxiv-2409.07140","DOIUrl":null,"url":null,"abstract":"Density measurements from single-sinker magnetic suspension densimeters need\nto be corrected to compensate for the magnetic effects of the measuring cell\nmaterials and the fluid on the coupling transmission system. While the magnetic\neffect of the densimeter materials can be easily determined, the fluid effect\nrequires the calculation of an apparatus-specific constant, $\\epsilon_{\\rho}$.\nIn this work, the apparatus-specific constant of the single-sinker magnetic\nsuspension densimeter at the University of Valladolid has been determined by\nusing two alternative methods. The first method, which uses density data for\nthe same fluid and conditions and different sinkers, yielded a value of\n$\\epsilon_{\\rho}$ = $4.6\\cdot10^{-5}$. The second method, obtained from\nmeasurements with pure oxygen, yielded a value of $\\epsilon_{\\rho}$ =\n$8.822\\cdot10^{-5}$. The second value is considered as more reliable, as the\nfirst method presents inherent limitations in this case.","PeriodicalId":501304,"journal":{"name":"arXiv - PHYS - Chemical Physics","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"arXiv - PHYS - Chemical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/arxiv-2409.07140","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Density measurements from single-sinker magnetic suspension densimeters need to be corrected to compensate for the magnetic effects of the measuring cell materials and the fluid on the coupling transmission system. While the magnetic effect of the densimeter materials can be easily determined, the fluid effect requires the calculation of an apparatus-specific constant, $\epsilon_{\rho}$. In this work, the apparatus-specific constant of the single-sinker magnetic suspension densimeter at the University of Valladolid has been determined by using two alternative methods. The first method, which uses density data for the same fluid and conditions and different sinkers, yielded a value of $\epsilon_{\rho}$ = $4.6\cdot10^{-5}$. The second method, obtained from measurements with pure oxygen, yielded a value of $\epsilon_{\rho}$ = $8.822\cdot10^{-5}$. The second value is considered as more reliable, as the first method presents inherent limitations in this case.
确定单沉降片磁悬浮密度计因流体特异性效应而产生的力传递误差,并对其在含氧混合气体中的使用进行修正
单沉降片磁悬浮密度计的密度测量需要进行校正,以补偿测量池材料和液体对耦合传输系统的磁效应。密度计材料的磁效应很容易确定,而流体效应则需要计算仪器的特定常数 $\epsilon_{\rho}$。第一种方法使用相同流体和条件以及不同沉降器的密度数据,得出的数值为$\epsilon_{\rho}$ = $4.6\cdot10^{-5}$。第二种方法是用纯氧测量得到的,其值为:$epsilon_{\rho}$ =$8.822\cdot10^{-5}$。第二个值被认为更可靠,因为第一种方法在这种情况下存在固有的局限性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信