Evaluation of Vessel Tracking Software for Prostatic Artery Embolization.

IF 2.8 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Peter Schott, Tiago Bilhim, Aaron Fischman, Paul Lohle, Patrick Freyhardt, Nuno Vasco Costa, Alexander Venmans, Marcus Katoh
{"title":"Evaluation of Vessel Tracking Software for Prostatic Artery Embolization.","authors":"Peter Schott, Tiago Bilhim, Aaron Fischman, Paul Lohle, Patrick Freyhardt, Nuno Vasco Costa, Alexander Venmans, Marcus Katoh","doi":"10.1007/s00270-024-03841-7","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate Vessel Tracking software for determining the prostatic arteries feeding the prostate gland during prostatic artery embolization (PAE) using Conebeam-CT (CBCT).</p><p><strong>Materials and methods: </strong>EmboGuide is a software developed to assist interventional radiologists in performing embolization of hypervascular tumors in the liver. In this study, a single-center retrospective image collection of 120 intraprocedural CBCT of 60 patients with benign prostatic hyperplasia treated using PAE between May 2017 and January 2019 was evaluated. All patients received 1 intraprocedural CBCT per side for evaluation of vessel anatomy. The \"reference standard\" of the vascular anatomy was defined by segmentation of the prostatic gland and marking of the prostatic artery in conjunction with pre-embolization DSA series. The datasets were then anonymized. Three interventional radiologists with experience in PAE from different centers reviewed the images and used the automatic feeder detection to determine the prostatic artery. Finally, two clinical experts compared the feeding vessels indicated in the \"reference standard\" and those identified by the readers. Objectives of the study were to evaluate the clinical performance of the software based on sensitivity and the agreement between interventional radiologists.</p><p><strong>Results: </strong>Sensitivity was estimated as 0.968 with a 90% confidence interval. Overall agreement was estimated as 0.801 with a 90% confidence interval. On both objectives (Sensitivity and Agreement), specifications were met.</p><p><strong>Conclusions: </strong>The results of this study suggest that EmboGuide can be used to determine the prostate arteries in PAE. The findings could be used to expand the intended use of EmboGuide to include PAE.</p>","PeriodicalId":9591,"journal":{"name":"CardioVascular and Interventional Radiology","volume":null,"pages":null},"PeriodicalIF":2.8000,"publicationDate":"2024-08-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"CardioVascular and Interventional Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00270-024-03841-7","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate Vessel Tracking software for determining the prostatic arteries feeding the prostate gland during prostatic artery embolization (PAE) using Conebeam-CT (CBCT).

Materials and methods: EmboGuide is a software developed to assist interventional radiologists in performing embolization of hypervascular tumors in the liver. In this study, a single-center retrospective image collection of 120 intraprocedural CBCT of 60 patients with benign prostatic hyperplasia treated using PAE between May 2017 and January 2019 was evaluated. All patients received 1 intraprocedural CBCT per side for evaluation of vessel anatomy. The "reference standard" of the vascular anatomy was defined by segmentation of the prostatic gland and marking of the prostatic artery in conjunction with pre-embolization DSA series. The datasets were then anonymized. Three interventional radiologists with experience in PAE from different centers reviewed the images and used the automatic feeder detection to determine the prostatic artery. Finally, two clinical experts compared the feeding vessels indicated in the "reference standard" and those identified by the readers. Objectives of the study were to evaluate the clinical performance of the software based on sensitivity and the agreement between interventional radiologists.

Results: Sensitivity was estimated as 0.968 with a 90% confidence interval. Overall agreement was estimated as 0.801 with a 90% confidence interval. On both objectives (Sensitivity and Agreement), specifications were met.

Conclusions: The results of this study suggest that EmboGuide can be used to determine the prostate arteries in PAE. The findings could be used to expand the intended use of EmboGuide to include PAE.

Abstract Image

评估前列腺动脉栓塞的血管跟踪软件
目的:评估Vessel Tracking软件在使用Conebeam-CT(CBCT)进行前列腺动脉栓塞术(PAE)时确定前列腺供血动脉的情况:EmboGuide 是一款用于协助介入放射医师对肝脏高血管肿瘤进行栓塞治疗的软件。在本研究中,对 2017 年 5 月至 2019 年 1 月间使用 PAE 治疗的 60 例良性前列腺增生患者的 120 例术中 CBCT 的单中心回顾性图像收集进行了评估。所有患者每侧接受 1 次术中 CBCT,用于评估血管解剖。血管解剖的 "参考标准 "是通过前列腺腺体的分割和前列腺动脉的标记以及栓塞前的 DSA 序列来确定的。然后对数据集进行匿名处理。来自不同中心的三位具有 PAE 经验的介入放射科医生对图像进行了审查,并使用自动馈线检测来确定前列腺动脉。最后,两位临床专家比较了 "参考标准 "和读者确定的供血血管。研究目的是根据灵敏度和介入放射科医生之间的一致意见评估软件的临床性能:灵敏度估计为 0.968,置信区间为 90%。总体一致性估计为 0.801,置信区间为 90%。两个目标(灵敏度和一致性)均符合规范:本研究结果表明,EmboGuide 可用于确定 PAE 中的前列腺动脉。研究结果可用于扩大 EmboGuide 的预期用途,将 PAE 包括在内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.50
自引率
13.80%
发文量
306
审稿时长
3-8 weeks
期刊介绍: CardioVascular and Interventional Radiology (CVIR) is the official journal of the Cardiovascular and Interventional Radiological Society of Europe, and is also the official organ of a number of additional distinguished national and international interventional radiological societies. CVIR publishes double blinded peer-reviewed original research work including clinical and laboratory investigations, technical notes, case reports, works in progress, and letters to the editor, as well as review articles, pictorial essays, editorials, and special invited submissions in the field of vascular and interventional radiology. Beside the communication of the latest research results in this field, it is also the aim of CVIR to support continuous medical education. Articles that are accepted for publication are done so with the understanding that they, or their substantive contents, have not been and will not be submitted to any other publication.
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信