Ka Yin Chan , Xulin Liao , Biyue Guo , Jimmy S.H. Tse , Peter H. Li , Allen M.Y. Cheong , William Ngo , Thomas C. Lam
{"title":"Ocular surface parameters repeatability and agreement —A comparison between Keratograph 5M and IDRA","authors":"Ka Yin Chan , Xulin Liao , Biyue Guo , Jimmy S.H. Tse , Peter H. Li , Allen M.Y. Cheong , William Ngo , Thomas C. Lam","doi":"10.1016/j.clae.2024.102281","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To evaluate the repeatability and agreement in dry eye measurements using Oculus Keratograph 5M (K5M) and SBM Sistemi IDRA (IDRA).</div></div><div><h3>Methods</h3><div>A total of 108 participants were enrolled and 108 eyes were evaluated. Tear meniscus height (TMH) and first and average non-invasive break-up time (NIBUT) were measured using the K5M and IDRA (order randomly assigned). TMH was measured using the built-in caliper tool while NIBUT was computed by the automatic algorithm of the instruments.</div></div><div><h3>Results</h3><div>The Bland Altman plots analysis showed a good agreement between the two instruments for TMH (95 % Limits of Agreement (LoA), −0.17 to 0.16), but not the first NIBUT (95 % LoA, −8.13 to 14.79) and average NIBUT (95 % LoA, −7.89 to 10.32). The values of the first and average NIBUT measured using IDRA were significantly shorter than in K5M (difference = median (IQR) −2.75 (−6.48- −0.28)s, <em>p</em> < 0.001 and difference = median (IQR) −1.65 (−3.97–1.89)s, <em>p</em> = 0.008 respectively). The TMH (<em>p</em> = 0.037) and NIBUT average (<em>p</em> = 0.033) measured by K5M, as well as the TMH (<em>p</em> = 0.040) measured by IDRA, exhibited unstable measurements across the three measurement times. The remaining parameters exhibited stability with three repeated measurements.</div></div><div><h3>Conclusion</h3><div>The NIBUT measurements are not interchangeable between IDRA and K5M, while the TMH was little difference between the two instruments. It is important to exercise caution when using different ocular surface analyzers to minimize errors in comparing multiple measurements.</div></div>","PeriodicalId":49087,"journal":{"name":"Contact Lens & Anterior Eye","volume":"47 6","pages":"Article 102281"},"PeriodicalIF":4.1000,"publicationDate":"2024-08-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Contact Lens & Anterior Eye","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1367048424001735","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose
To evaluate the repeatability and agreement in dry eye measurements using Oculus Keratograph 5M (K5M) and SBM Sistemi IDRA (IDRA).
Methods
A total of 108 participants were enrolled and 108 eyes were evaluated. Tear meniscus height (TMH) and first and average non-invasive break-up time (NIBUT) were measured using the K5M and IDRA (order randomly assigned). TMH was measured using the built-in caliper tool while NIBUT was computed by the automatic algorithm of the instruments.
Results
The Bland Altman plots analysis showed a good agreement between the two instruments for TMH (95 % Limits of Agreement (LoA), −0.17 to 0.16), but not the first NIBUT (95 % LoA, −8.13 to 14.79) and average NIBUT (95 % LoA, −7.89 to 10.32). The values of the first and average NIBUT measured using IDRA were significantly shorter than in K5M (difference = median (IQR) −2.75 (−6.48- −0.28)s, p < 0.001 and difference = median (IQR) −1.65 (−3.97–1.89)s, p = 0.008 respectively). The TMH (p = 0.037) and NIBUT average (p = 0.033) measured by K5M, as well as the TMH (p = 0.040) measured by IDRA, exhibited unstable measurements across the three measurement times. The remaining parameters exhibited stability with three repeated measurements.
Conclusion
The NIBUT measurements are not interchangeable between IDRA and K5M, while the TMH was little difference between the two instruments. It is important to exercise caution when using different ocular surface analyzers to minimize errors in comparing multiple measurements.
期刊介绍:
Contact Lens & Anterior Eye is a research-based journal covering all aspects of contact lens theory and practice, including original articles on invention and innovations, as well as the regular features of: Case Reports; Literary Reviews; Editorials; Instrumentation and Techniques and Dates of Professional Meetings.