{"title":"Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices","authors":"Yue Feng MD, MSc, Tore Nitter MD, PhD, Geir Bertelsen MD, PhD, Aleksandar Stojanovic MD, PhD","doi":"10.1111/ceo.14423","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster 700, and MS-39.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components <i>J</i><sub>0</sub> and <i>J</i><sub>45</sub>. The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within-subject standard deviation (<i>S</i><sub>w</sub>). The agreement among the four devices was assessed by pairwise comparisons and Bland–Altman plots.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>For all devices, the repeatability of TCA measurements showed <i>S</i><sub>w</sub> ≤0.23 D for TCA magnitude, ≤0.14 D for <i>J</i><sub>0</sub>, and ≤0.12 D for <i>J</i><sub>45</sub>. There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (<i>S</i><sub>w</sub>) of axis measurements had a statistically significant negative correlation with the TCA magnitude (<i>p</i> < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements (<i>S</i><sub>w</sub>: 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 ± 0.033 mm (<i>p</i> < 0.001).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.</p>\n </section>\n </div>","PeriodicalId":55253,"journal":{"name":"Clinical and Experimental Ophthalmology","volume":"52 8","pages":"800-810"},"PeriodicalIF":4.9000,"publicationDate":"2024-07-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/ceo.14423","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/ceo.14423","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster 700, and MS-39.
Methods
Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J0 and J45. The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within-subject standard deviation (Sw). The agreement among the four devices was assessed by pairwise comparisons and Bland–Altman plots.
Results
For all devices, the repeatability of TCA measurements showed Sw ≤0.23 D for TCA magnitude, ≤0.14 D for J0, and ≤0.12 D for J45. There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (Sw) of axis measurements had a statistically significant negative correlation with the TCA magnitude (p < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements (Sw: 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 ± 0.033 mm (p < 0.001).
Conclusions
All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.
期刊介绍:
Clinical & Experimental Ophthalmology is the official journal of The Royal Australian and New Zealand College of Ophthalmologists. The journal publishes peer-reviewed original research and reviews dealing with all aspects of clinical practice and research which are international in scope and application. CEO recognises the importance of collaborative research and welcomes papers that have a direct influence on ophthalmic practice but are not unique to ophthalmology.