{"title":"Validation of the validity of the new wrist-type fully automatic blood pressure monitor: DBP-2242 according to ISO 81060-2:2018+AMD 1:2020.","authors":"Yuehong Jin, Xiaoming Bei, Ying Jin","doi":"10.1097/MBP.0000000000000716","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Evaluation of the accuracy of the wrist-type fully automatic blood pressure (BP) monitor: DBP-2242 in adolescent and adult populations according to ISO 81060-2 : 2018+Amd.1 : 2020.</p><p><strong>Methods: </strong>BP measurements were taken from the subjects using the same-arm sequential method, and BP data measured by a mercury sphygmomanometer was used as the standard.</p><p><strong>Results: </strong>This study analyzed 262 sets of data from 89 subjects. According to criterion 1, the mean difference between the SBP of the test and reference devices was -0.57 ± 7.31 mmHg, and the DBP was -2.27 ± 7.17 mmHg, which is in accordance with the requirements. According to criterion 2, the average difference between the SBP was -0.57 ± 6.25 mmHg and the DBP was -2.27 ± 5.99 mmHg, which is in accordance with the requirements.</p><p><strong>Conclusion: </strong>Wrist-type fully automatic BP monitor: DBP-2242 complies with ISO 81060-2 : 2018+Amd.1 : 2020 and can be used for BP measurement in adolescent and adult populations.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":"317-320"},"PeriodicalIF":1.2000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11524616/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000716","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/16 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Evaluation of the accuracy of the wrist-type fully automatic blood pressure (BP) monitor: DBP-2242 in adolescent and adult populations according to ISO 81060-2 : 2018+Amd.1 : 2020.
Methods: BP measurements were taken from the subjects using the same-arm sequential method, and BP data measured by a mercury sphygmomanometer was used as the standard.
Results: This study analyzed 262 sets of data from 89 subjects. According to criterion 1, the mean difference between the SBP of the test and reference devices was -0.57 ± 7.31 mmHg, and the DBP was -2.27 ± 7.17 mmHg, which is in accordance with the requirements. According to criterion 2, the average difference between the SBP was -0.57 ± 6.25 mmHg and the DBP was -2.27 ± 5.99 mmHg, which is in accordance with the requirements.
Conclusion: Wrist-type fully automatic BP monitor: DBP-2242 complies with ISO 81060-2 : 2018+Amd.1 : 2020 and can be used for BP measurement in adolescent and adult populations.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.