{"title":"Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.","authors":"Irene Abicca, Domenico Schiano-Lomoriello, Marta Gilardi, Daniela Giannini, Valentin Dinu, Catarina Coutinho, Giacomo Savini","doi":"10.3928/1081597X-20240514-03","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus.</p><p><strong>Methods: </strong>Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired <i>t</i>-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices.</p><p><strong>Results: </strong>Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry.</p><p><strong>Conclusions: </strong>The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. <b>[<i>J Refract Surg</i>. 2024;40(7):e445-e452.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"40 7","pages":"e445-e452"},"PeriodicalIF":2.9000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20240514-03","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus.
Methods: Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices.
Results: Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry.
Conclusions: The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. [J Refract Surg. 2024;40(7):e445-e452.].
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.