Evaluation of Electrical Cardiometry to Assess Fluid Responsiveness in Patients with Acute Circulatory Failure: A Comparative Study with Transthoracic Echocardiography.

IF 1.5 Q3 CRITICAL CARE MEDICINE
Shashikant Sharma, Rashmi Ramachandran, Vimi Rewari, Anjan Trikha
{"title":"Evaluation of Electrical Cardiometry to Assess Fluid Responsiveness in Patients with Acute Circulatory Failure: A Comparative Study with Transthoracic Echocardiography.","authors":"Shashikant Sharma, Rashmi Ramachandran, Vimi Rewari, Anjan Trikha","doi":"10.5005/jp-journals-10071-24753","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>Acute circulatory failure is commonly encountered in critically ill patients, that requires fluid administration as the first line of treatment. However, only 50% of patients are fluid-responsive. Identification of fluid responders is essential to avoid the harmful effects of overzealous fluid therapy. Electrical cardiometry (EC) is a non-invasive bedside tool and has proven to be as good as transthoracic echocardiography (TTE) to track changes in cardiac output. We aimed to look for an agreement between EC and TTE for tracking changes in cardiac output in adult patients with acute circulatory failure before and after the passive leg-raising maneuver.</p><p><strong>Materials and methods: </strong>Prospective comparative study, conducted at a Tertiary Care Teaching Hospital.</p><p><strong>Results: </strong>We recruited 125 patients with acute circulatory failure and found 42.4% (53 out of 125) to be fluid-responsive. The Bland-Altman plot analysis showed a mean difference of 2.08 L/min between EC and TTE, with a precision of 3.8 L/min. The limits of agreement (defined as bias ± 1.96SD), were -1.7 L/min and 5.8 L/min, respectively. The percentage of error between EC and TTE was 56% with acceptable limits of 30%.</p><p><strong>Conclusion: </strong>The percentage error beyond the acceptable limit suggests the non-interchangeability of the two techniques. More studies with larger sample sizes are required to establish the interchangeability of EC with TTE for tracking changes in cardiac output in critically ill patients with acute circulatory failure.</p><p><strong>How to cite this article: </strong>Sharma S, Ramachandran R, Rewari V, Trikha A. Evaluation of Electrical Cardiometry to Assess Fluid Responsiveness in Patients with Acute Circulatory Failure: A Comparative Study with Transthoracic Echocardiography. Indian J Crit Care Med 2024;28(7):650-656.</p>","PeriodicalId":47664,"journal":{"name":"Indian Journal of Critical Care Medicine","volume":"28 7","pages":"650-656"},"PeriodicalIF":1.5000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11234120/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Critical Care Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5005/jp-journals-10071-24753","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CRITICAL CARE MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Aim: Acute circulatory failure is commonly encountered in critically ill patients, that requires fluid administration as the first line of treatment. However, only 50% of patients are fluid-responsive. Identification of fluid responders is essential to avoid the harmful effects of overzealous fluid therapy. Electrical cardiometry (EC) is a non-invasive bedside tool and has proven to be as good as transthoracic echocardiography (TTE) to track changes in cardiac output. We aimed to look for an agreement between EC and TTE for tracking changes in cardiac output in adult patients with acute circulatory failure before and after the passive leg-raising maneuver.

Materials and methods: Prospective comparative study, conducted at a Tertiary Care Teaching Hospital.

Results: We recruited 125 patients with acute circulatory failure and found 42.4% (53 out of 125) to be fluid-responsive. The Bland-Altman plot analysis showed a mean difference of 2.08 L/min between EC and TTE, with a precision of 3.8 L/min. The limits of agreement (defined as bias ± 1.96SD), were -1.7 L/min and 5.8 L/min, respectively. The percentage of error between EC and TTE was 56% with acceptable limits of 30%.

Conclusion: The percentage error beyond the acceptable limit suggests the non-interchangeability of the two techniques. More studies with larger sample sizes are required to establish the interchangeability of EC with TTE for tracking changes in cardiac output in critically ill patients with acute circulatory failure.

How to cite this article: Sharma S, Ramachandran R, Rewari V, Trikha A. Evaluation of Electrical Cardiometry to Assess Fluid Responsiveness in Patients with Acute Circulatory Failure: A Comparative Study with Transthoracic Echocardiography. Indian J Crit Care Med 2024;28(7):650-656.

评估急性循环衰竭患者输液反应性的心电测量法:与经胸超声心动图的比较研究。
目的:急性循环衰竭是危重病人的常见病,需要将输液作为第一线治疗措施。然而,只有 50% 的患者对输液有反应。要避免过度输液治疗的有害影响,识别输液反应者至关重要。心电图(EC)是一种无创床边工具,在跟踪心输出量变化方面已被证明与经胸超声心动图(TTE)一样出色。我们的目的是寻找在被动抬腿动作前后,EC 和 TTE 在追踪急性循环衰竭成人患者心输出量变化方面的一致性:在一家三级教学医院进行的前瞻性比较研究:我们招募了 125 名急性循环衰竭患者,发现 42.4%(125 人中有 53 人)对液体有反应。Bland-Altman 图分析显示,EC 和 TTE 的平均差异为 2.08 升/分钟,精确度为 3.8 升/分钟。一致性极限(定义为偏差 ± 1.96SD)分别为-1.7 升/分钟和 5.8 升/分钟。EC 和 TTE 之间的误差百分比为 56%,可接受范围为 30%:结论:超出可接受范围的误差百分比表明这两种技术不可互换。需要更多样本量更大的研究来确定心电图与 TTE 在跟踪急性循环衰竭重症患者心输出量变化方面的互换性:Sharma S, Ramachandran R, Rewari V, Trikha A. Evaluation of Electrical Cardiometry to Assess Fluid Responsiveness in Patients with Acute Circulatory Failure:与经胸超声心动图的比较研究》。Indian J Crit Care Med 2024;28(7):650-656.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.50
自引率
10.00%
发文量
299
期刊介绍: Indian Journal of Critical Care Medicine (ISSN 0972-5229) is specialty periodical published under the auspices of Indian Society of Critical Care Medicine. Journal encourages research, education and dissemination of knowledge in the fields of critical and emergency medicine.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信