Simulator Verification Is Potentially Beneficial for the Fitting of Softband Bone Conduction Hearing Devices in Young Children.

IF 1.9 3区 医学 Q3 CLINICAL NEUROLOGY
Otology & Neurotology Pub Date : 2024-08-01 Epub Date: 2024-06-26 DOI:10.1097/MAO.0000000000004245
Le Sun, Lu Ping, Xinmiao Fan, Jian Wang, Xiaowei Chen
{"title":"Simulator Verification Is Potentially Beneficial for the Fitting of Softband Bone Conduction Hearing Devices in Young Children.","authors":"Le Sun, Lu Ping, Xinmiao Fan, Jian Wang, Xiaowei Chen","doi":"10.1097/MAO.0000000000004245","DOIUrl":null,"url":null,"abstract":"<p><strong>Hypothesis: </strong>The current study employed a skull-simulator verification method to assess whether the output of softband bone conduction hearing devices (BCHDs) at the manufacturer's default settings deviated widely from the target determined by the fitting formula.</p><p><strong>Background: </strong>Real ear analysis is utilized for the verification of the fitting of air conduction hearing devices (ACHDs) in a variety of institutions. This procedure, however, has not been used in the fitting of BCHDs, largely due to the difficulty of testing the output of these devices to temporal bones. Despite the availability of skull simulators, they have not been utilized clinically to measure BCHD output.</p><p><strong>Materials and methods: </strong>This prospective, single-center study enrolled 42 subjects, aged 3 months to 10 years, with microtia-atresia-associated mild-to-severe bilateral conductive hearing loss. Hearing sensitivity was evaluated behaviorally by pure tone audiometry (PTA) in 22 subjects 4 years or older (the PTA group), and by auditory brainstem response (ABR) in 20 subjects younger than 4 years (the ABR group). Following 6 months of subjects wearing the prescribed softband BCHDs, their dial level (DL) thresholds were reassessed while using their own BCHDs, configured with zero gain across all frequencies, functioning solely as a bone vibrator. These DL thresholds were inputted into the fitting formula, desired sensation level-bone conduction devices (DSL-BCD) for children, to obtain the target values of BCHD output. The simulator output of the BCHD programmed at the manufacturer's default setting was measured in response to speech presented at 55, 65, and 80 dB SPL, followed by gain adjustment based on the differences between the simulator output and the target. Aided speech intelligibility index (SII) was measured before and after the gain adjustment.</p><p><strong>Results: </strong>The softband BCHDs at the manufacturer's settings generally had lower output than the prescribed target values. This difference was larger at low frequencies and low levels. Across the 12 points tested (four frequencies from 500 to 4000 Hz multiplied by three levels), 22 (52.3%) and 42 (100%) BCHDs had deviations of +7 and +5 dB, respectively, at one point or more. The gain adjustments reduced the deviation and improved the SII values at the two lower levels of speech presented.</p><p><strong>Conclusion: </strong>The simulator output of softband bone conduction hearing devices (BCHDs) with the manufacturer's settings may exhibit significant deviations from the formula. Objective output verification should be considered a beneficial step in BCHD fitting and is recommended when applicable.</p>","PeriodicalId":19732,"journal":{"name":"Otology & Neurotology","volume":null,"pages":null},"PeriodicalIF":1.9000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Otology & Neurotology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MAO.0000000000004245","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/26 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Hypothesis: The current study employed a skull-simulator verification method to assess whether the output of softband bone conduction hearing devices (BCHDs) at the manufacturer's default settings deviated widely from the target determined by the fitting formula.

Background: Real ear analysis is utilized for the verification of the fitting of air conduction hearing devices (ACHDs) in a variety of institutions. This procedure, however, has not been used in the fitting of BCHDs, largely due to the difficulty of testing the output of these devices to temporal bones. Despite the availability of skull simulators, they have not been utilized clinically to measure BCHD output.

Materials and methods: This prospective, single-center study enrolled 42 subjects, aged 3 months to 10 years, with microtia-atresia-associated mild-to-severe bilateral conductive hearing loss. Hearing sensitivity was evaluated behaviorally by pure tone audiometry (PTA) in 22 subjects 4 years or older (the PTA group), and by auditory brainstem response (ABR) in 20 subjects younger than 4 years (the ABR group). Following 6 months of subjects wearing the prescribed softband BCHDs, their dial level (DL) thresholds were reassessed while using their own BCHDs, configured with zero gain across all frequencies, functioning solely as a bone vibrator. These DL thresholds were inputted into the fitting formula, desired sensation level-bone conduction devices (DSL-BCD) for children, to obtain the target values of BCHD output. The simulator output of the BCHD programmed at the manufacturer's default setting was measured in response to speech presented at 55, 65, and 80 dB SPL, followed by gain adjustment based on the differences between the simulator output and the target. Aided speech intelligibility index (SII) was measured before and after the gain adjustment.

Results: The softband BCHDs at the manufacturer's settings generally had lower output than the prescribed target values. This difference was larger at low frequencies and low levels. Across the 12 points tested (four frequencies from 500 to 4000 Hz multiplied by three levels), 22 (52.3%) and 42 (100%) BCHDs had deviations of +7 and +5 dB, respectively, at one point or more. The gain adjustments reduced the deviation and improved the SII values at the two lower levels of speech presented.

Conclusion: The simulator output of softband bone conduction hearing devices (BCHDs) with the manufacturer's settings may exhibit significant deviations from the formula. Objective output verification should be considered a beneficial step in BCHD fitting and is recommended when applicable.

模拟器验证对幼儿佩戴软带骨导助听器有潜在益处。
假设:本研究采用头骨模拟器验证方法,评估软带骨导听力设备(BCHD)在制造商默认设置下的输出是否与验配公式确定的目标有较大偏差:背景:在许多机构中,真耳分析被用于验证气导听力设备(ACHD)的验配。背景:在许多机构中,真耳分析被用于验证气导助听器(ACHD)的验配,但在 BCHD 的验配中还没有使用过,这主要是由于很难测试这些助听器对颞骨的输出。尽管有头骨模拟器,但临床上还没有使用它们来测量 BCHD 的输出:这项前瞻性的单中心研究共招募了 42 名年龄在 3 个月至 10 岁之间的受试者,他们都患有与小耳畸形相关的轻度至重度双侧传导性听力损失。22 名 4 岁或以上的受试者(PTA 组)通过纯音测听(PTA),20 名 4 岁以下的受试者(ABR 组)通过听性脑干反应(ABR)对听力敏感性进行了行为评估。受试者佩戴规定的软带北京听力障碍消除器 6 个月后,使用他们自己的北京听力障碍消除器重新评估他们的拨码电平 (DL) 阈值。将这些DL阈值输入 "儿童所需感觉水平-骨传导设备(DSL-BCD)"的拟合公式中,即可获得BCD输出的目标值。根据 55、65 和 80 dB SPL 下的语音,测量按制造商默认设置编程的骨传导设备模拟器输出,然后根据模拟器输出与目标值之间的差异进行增益调整。在增益调整前后测量了辅助语音清晰度指数(SII):结果:按照制造商的设置,软带 BCHD 的输出通常低于规定的目标值。这种差异在低频和低电平时更大。在测试的 12 个点(500 至 4000 Hz 的四个频率乘以三个电平)中,分别有 22 个(52.3%)和 42 个(100%)BCD 在一个点或更多点的偏差达到 +7 和 +5 dB。增益调整降低了偏差,并改善了两个较低语音级别的 SII 值:结论:软带骨导听力设备(BCD)的模拟器输出与制造商的设置可能会出现明显的公式偏差。客观输出验证应被视为骨导听力设备验配的一个有益步骤,建议在适用时进行验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Otology & Neurotology
Otology & Neurotology 医学-耳鼻喉科学
CiteScore
3.80
自引率
14.30%
发文量
509
审稿时长
3-6 weeks
期刊介绍: ​​​​​Otology & Neurotology publishes original articles relating to both clinical and basic science aspects of otology, neurotology, and cranial base surgery. As the foremost journal in its field, it has become the favored place for publishing the best of new science relating to the human ear and its diseases. The broadly international character of its contributing authors, editorial board, and readership provides the Journal its decidedly global perspective.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信