Improving the Application of Prothrombin Time-Derived Fibrinogen Assay Through Value Transfer from the Von Clauss Method.

IF 0.7 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY
Yan Zhang, Xiaoyong Duan, Xiao Li, Yunfeng Ling, Fan Wu, Sheng Wang, Xingtai Liu, Ying Chen, Hongbin Tong, Dongmei Li
{"title":"Improving the Application of Prothrombin Time-Derived Fibrinogen Assay Through Value Transfer from the Von Clauss Method.","authors":"Yan Zhang, Xiaoyong Duan, Xiao Li, Yunfeng Ling, Fan Wu, Sheng Wang, Xingtai Liu, Ying Chen, Hongbin Tong, Dongmei Li","doi":"10.7754/Clin.Lab.2024.231026","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aimed to improve the accuracy of the fibrinogen (Fib) prothrombin time-derived (PT-der) method. To achieve this, a value transfer method was introduced for calibration, and its effectiveness was assessed.</p><p><strong>Methods: </strong>The PT-der Fib assay was calibrated by pooled samples (assigned by the von Clauss method) in three different ways: 1) multipoint calibration using an automatic dilution system, 2) multipoint calibration using a manual dilution method, and 3) manual calibration with multiple concentrations. Three calibration equations (1, 2, and 3) were obtained and an optimal equation was selected by comparing the detection results of the von Clauss method with the PT-der method. Subsequently, the optimal equation was assessed for an accuracy limit, and linear analysis and reference interval verification were performed following the guidelines (EP15-A and EP6-A) issued by the CLSI.</p><p><strong>Results: </strong>Compared with the other two equations (equation 1 and 2), equation 3, available from manual calibration with multiple concentrations, showed a better performance for the PT-der determination in a primary cohort (n = 208), and a good agreement (99% of the results between 1.52 and 6.30 g/L were interchangeable) was validated (n = 3226). The reference interval was also verified in almost all healthy individuals (39/40). However, the discrep-ancy between the two methods was observed in several specific conditions, such as hyperfibrinolysis.</p><p><strong>Conclusions: </strong>Manual calibration with multiple concentrations is better for the Fib PT-der method assay. As a rapid, accurate, and economical test, the performance of the Fib PT-der method has been verified and may be more applicable than before.</p>","PeriodicalId":10384,"journal":{"name":"Clinical laboratory","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical laboratory","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7754/Clin.Lab.2024.231026","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: This study aimed to improve the accuracy of the fibrinogen (Fib) prothrombin time-derived (PT-der) method. To achieve this, a value transfer method was introduced for calibration, and its effectiveness was assessed.

Methods: The PT-der Fib assay was calibrated by pooled samples (assigned by the von Clauss method) in three different ways: 1) multipoint calibration using an automatic dilution system, 2) multipoint calibration using a manual dilution method, and 3) manual calibration with multiple concentrations. Three calibration equations (1, 2, and 3) were obtained and an optimal equation was selected by comparing the detection results of the von Clauss method with the PT-der method. Subsequently, the optimal equation was assessed for an accuracy limit, and linear analysis and reference interval verification were performed following the guidelines (EP15-A and EP6-A) issued by the CLSI.

Results: Compared with the other two equations (equation 1 and 2), equation 3, available from manual calibration with multiple concentrations, showed a better performance for the PT-der determination in a primary cohort (n = 208), and a good agreement (99% of the results between 1.52 and 6.30 g/L were interchangeable) was validated (n = 3226). The reference interval was also verified in almost all healthy individuals (39/40). However, the discrep-ancy between the two methods was observed in several specific conditions, such as hyperfibrinolysis.

Conclusions: Manual calibration with multiple concentrations is better for the Fib PT-der method assay. As a rapid, accurate, and economical test, the performance of the Fib PT-der method has been verified and may be more applicable than before.

通过 Von Clauss 方法的价值转移改进凝血酶原时间衍生纤维蛋白原测定法的应用。
研究背景本研究旨在提高纤维蛋白原(Fib)凝血酶原时间衍生法(PT-der)的准确性。为此,我们引入了数值转移法进行校准,并对其有效性进行了评估:PT-der 纤维测定法通过三种不同的方法对集合样本(按 von Clauss 方法分配)进行校准:1)使用自动稀释系统进行多点校准;2)使用手动稀释法进行多点校准;3)使用多种浓度进行手动校准。通过比较 von Clauss 方法和 PT-der 方法的检测结果,得到了三个校准方程(1、2 和 3),并选出了一个最佳方程。随后,根据 CLSI 发布的指南(EP15-A 和 EP6-A),对最佳方程进行了准确度限评估,并进行了线性分析和参考区间验证:与其他两个公式(公式 1 和 2)相比,公式 3 可通过多浓度手动校准获得,在初级队列(n = 208)中的 PT-der 测定中表现更佳,并验证了良好的一致性(99% 的结果在 1.52 至 6.30 g/L 之间可互换)(n = 3226)。几乎所有健康人(39/40)的参考区间也得到了验证。然而,在几种特殊情况下,如纤维蛋白溶解亢进,两种方法之间出现了差异:结论:多浓度手动校准更适合 Fib PT-der 方法检测。作为一种快速、准确、经济的检测方法,Fib PT-der 方法的性能已得到验证,可能比以前更适用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical laboratory
Clinical laboratory 医学-医学实验技术
CiteScore
1.50
自引率
0.00%
发文量
494
审稿时长
3 months
期刊介绍: Clinical Laboratory is an international fully peer-reviewed journal covering all aspects of laboratory medicine and transfusion medicine. In addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies. The journal publishes original articles, review articles, posters, short reports, case studies and letters to the editor dealing with 1) the scientific background, implementation and diagnostic significance of laboratory methods employed in hospitals, blood banks and physicians'' offices and with 2) scientific, administrative and clinical aspects of transfusion medicine and 3) in addition to transfusion medicine topics Clinical Laboratory represents submissions concerning tissue transplantation and hematopoietic, cellular and gene therapies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信