Clinical validation of the Ion Torrent Oncomine Myeloid Assay GX v2 on the Genexus Integrated Sequencer as a stand-alone assay for single-nucleotide variants, insertions/deletions, and fusion genes: Challenges, performance, and perspectives.
{"title":"Clinical validation of the Ion Torrent Oncomine Myeloid Assay GX v2 on the Genexus Integrated Sequencer as a stand-alone assay for single-nucleotide variants, insertions/deletions, and fusion genes: Challenges, performance, and perspectives.","authors":"Kritika Krishnamurthy, Jiani Chai, Xiaowei Liu, Yanhua Wang, Rizwan Naeem, D Yitzchak Goldstein","doi":"10.1093/ajcp/aqae063","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Myeloid neoplasms require comprehensive characterization of genetic abnormalities, including single-nucleotide variants, small insertions and deletions, and fusions and translocations for management. The Oncomine Myeloid Assay GX v2 (Thermo Fisher Scientific) analyzes 17 full genes, 28 hotspot genes, 30 fusion driver genes, and 5 expression genes.</p><p><strong>Methods: </strong>The validation set included 192 DNA samples, 28 RNA samples, and 9 cell lines and contrived controls. The DNA and RNA were extracted from both peripheral blood and bone marrow. Library preparation, templating, and sequencing was performed on the fully automated Genexus Integrated Sequencer (Thermo Fisher Scientific). The sequencing data were analyzed by manual curation, default Oncomine filters and the Oncomine Reporter (Thermo Fisher Scientific).</p><p><strong>Results: </strong>Of the 600 reference pathogenic DNA variants targeted by the assay, concordance was seen in 98.3% of unfiltered variant call format files. Precision and reproducibility were 100%, and the lower limit of detection was 2% variant allele frequency for DNA. Inability to detect variants in long homopolymer regions intrinsic to the Ion Torrent chemistry led to 7 missed variants; 100% concordance was seen with reference RNA samples.</p><p><strong>Conclusions: </strong>This extensive clinical validation of the Oncomine Myeloid Assay GX v2 on the Genexus Integrated Sequencer with its built-in bioinformatics pipeline and Ion Torrent Oncomine Reporter shows robust performance in terms of variant calling accuracy, precision, and reproducibility, with the advantage of a rapid turnaround time of 2 days. The greatest limitation is the inability to detect variants in long homopolymer regions.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/ajcp/aqae063","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: Myeloid neoplasms require comprehensive characterization of genetic abnormalities, including single-nucleotide variants, small insertions and deletions, and fusions and translocations for management. The Oncomine Myeloid Assay GX v2 (Thermo Fisher Scientific) analyzes 17 full genes, 28 hotspot genes, 30 fusion driver genes, and 5 expression genes.
Methods: The validation set included 192 DNA samples, 28 RNA samples, and 9 cell lines and contrived controls. The DNA and RNA were extracted from both peripheral blood and bone marrow. Library preparation, templating, and sequencing was performed on the fully automated Genexus Integrated Sequencer (Thermo Fisher Scientific). The sequencing data were analyzed by manual curation, default Oncomine filters and the Oncomine Reporter (Thermo Fisher Scientific).
Results: Of the 600 reference pathogenic DNA variants targeted by the assay, concordance was seen in 98.3% of unfiltered variant call format files. Precision and reproducibility were 100%, and the lower limit of detection was 2% variant allele frequency for DNA. Inability to detect variants in long homopolymer regions intrinsic to the Ion Torrent chemistry led to 7 missed variants; 100% concordance was seen with reference RNA samples.
Conclusions: This extensive clinical validation of the Oncomine Myeloid Assay GX v2 on the Genexus Integrated Sequencer with its built-in bioinformatics pipeline and Ion Torrent Oncomine Reporter shows robust performance in terms of variant calling accuracy, precision, and reproducibility, with the advantage of a rapid turnaround time of 2 days. The greatest limitation is the inability to detect variants in long homopolymer regions.