Technical Note: Significant Positive Bias in Lower Concentrations but Negative Bias in Higher Concentrations in Testosterone Measurement Using Beckman Access Immunoassay Compared to a Liquid Chromatography-Tandem Mass Spectrometry Reference Method.

IF 1.1 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY
Amitava Dasgupta, Kelsey Woodard, Bheemraj Ramoo, Clinton Frazee, Uttam Garg
{"title":"<i>Technical Note:</i> Significant Positive Bias in Lower Concentrations but Negative Bias in Higher Concentrations in Testosterone Measurement Using Beckman Access Immunoassay Compared to a Liquid Chromatography-Tandem Mass Spectrometry Reference Method.","authors":"Amitava Dasgupta, Kelsey Woodard, Bheemraj Ramoo, Clinton Frazee, Uttam Garg","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Testosterone is the principal male sex hormone and is secreted primarily by the testes. In most clinical laboratories testosterone is routinely measured for diagnosis of male hypogonadism or androgen excess in females using FDA approved immunoassays. We compared testosterone values measured by Beckman access immunoassay with those measured by a reference LC-MS/MS method.</p><p><strong>Methods: </strong>Testosterone was measured in 36 patients using left over serum or plasma specimens by both Beckman immunoassay on the DXI 800 analyzer and a reference LC-MS/MS method.</p><p><strong>Results: </strong>We observed overall significant negative bias of approximately 31.9 % when testosterone values obtained by the reference LC-MS/MS method were plotted in the x-axis and the corresponding testosterone values using the immunoassay in the y-axis, as regression equation was y=0.6887x+38.81 (n=36). The corresponding Deming regression was y=0.6639x+34.7163. However, in eight specimens with low testosterone concentrations, immunoassays significantly overestimated testosterone concentrations.</p><p><strong>Conclusions: </strong>Immunoassays may underestimate the true testosterone concentration in males but overestimate in females with low testosterone concentration. Therefore, for diagnosis of hypogonadism in males and androgen excess in females, testosterone values obtained by Beckman Access immunoassay on the DXI 800 analyzer should be interpreted with caution.</p>","PeriodicalId":8228,"journal":{"name":"Annals of clinical and laboratory science","volume":"54 2","pages":"258-261"},"PeriodicalIF":1.1000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of clinical and laboratory science","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Testosterone is the principal male sex hormone and is secreted primarily by the testes. In most clinical laboratories testosterone is routinely measured for diagnosis of male hypogonadism or androgen excess in females using FDA approved immunoassays. We compared testosterone values measured by Beckman access immunoassay with those measured by a reference LC-MS/MS method.

Methods: Testosterone was measured in 36 patients using left over serum or plasma specimens by both Beckman immunoassay on the DXI 800 analyzer and a reference LC-MS/MS method.

Results: We observed overall significant negative bias of approximately 31.9 % when testosterone values obtained by the reference LC-MS/MS method were plotted in the x-axis and the corresponding testosterone values using the immunoassay in the y-axis, as regression equation was y=0.6887x+38.81 (n=36). The corresponding Deming regression was y=0.6639x+34.7163. However, in eight specimens with low testosterone concentrations, immunoassays significantly overestimated testosterone concentrations.

Conclusions: Immunoassays may underestimate the true testosterone concentration in males but overestimate in females with low testosterone concentration. Therefore, for diagnosis of hypogonadism in males and androgen excess in females, testosterone values obtained by Beckman Access immunoassay on the DXI 800 analyzer should be interpreted with caution.

技术说明:与液相色谱-串联质谱参考方法相比,使用贝克曼通路免疫测定法测定睾酮时,低浓度存在明显的正偏差,而高浓度存在明显的负偏差。
目的:睾酮是男性的主要性激素,主要由睾丸分泌:睾酮是男性的主要性激素,主要由睾丸分泌。在大多数临床实验室中,睾酮是用于诊断男性性腺功能减退症或女性雄激素过多症的常规检测指标,使用的是美国食品及药物管理局(FDA)批准的免疫测定法。我们将贝克曼免疫测定法测得的睾酮值与 LC-MS/MS 参考方法测得的睾酮值进行了比较:使用 DXI 800 分析仪上的贝克曼免疫测定法和一种 LC-MS/MS 参考方法对 36 名患者的残留血清或血浆标本进行了睾酮测量:如果将 LC-MS/MS 参考方法获得的睾酮值绘制成 x 轴,将免疫测定法获得的相应睾酮值绘制成 y 轴,回归方程为 y=0.6887x+38.81 (n=36)。相应的戴明回归方程为 y=0.6639x+34.7163。然而,在 8 份睾酮浓度较低的标本中,免疫测定法明显高估了睾酮浓度:结论:免疫测定可能会低估男性睾酮的真实浓度,但会高估低睾酮浓度女性的睾酮浓度。因此,在诊断男性性腺功能减退症和女性雄激素过多症时,应谨慎解释贝克曼 Access 免疫测定法在 DXI 800 分析仪上获得的睾酮值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of clinical and laboratory science
Annals of clinical and laboratory science 医学-医学实验技术
CiteScore
1.60
自引率
0.00%
发文量
112
审稿时长
6-12 weeks
期刊介绍: The Annals of Clinical & Laboratory Science welcomes manuscripts that report research in clinical science, including pathology, clinical chemistry, biotechnology, molecular biology, cytogenetics, microbiology, immunology, hematology, transfusion medicine, organ and tissue transplantation, therapeutics, toxicology, and clinical informatics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信