{"title":"Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm.","authors":"Yi Zhou, Jia-Hui Xia, Wei Zhang, Yan Li, Ji-Guang Wang","doi":"10.1097/MBP.0000000000000706","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm.</p><p><strong>Methods: </strong>An approximately equal number of patients with persistent atrial fibrillation and individuals with sinus rhythm were recruited from outpatients and inpatients of Ruijin Hospital, Shanghai, China. BP was measured three times consecutively with a 30-s interval with the OMRON HEM-7361T automatic electronic BP monitor for atrial fibrillation detection. A hand-held single lead electrocardiogram device was used for simultaneous electrocardiogram recordings.</p><p><strong>Results: </strong>The device accurately identified atrial fibrillation in 100 (99.0%) of the 101 patients, with only 1 patient incorrectly classified as non-atrial fibrillation. The device correctly identified 99 (95.2%) of the 104 participants with sinus rhythm as non-atrial fibrillation, with five participants incorrectly classified as atrial fibrillation. The device had a positive predictive value of 95.2%, negative predictive value of 99.0%, and overall accuracy of 97.1%. Among the six misclassified participants, one with atrial fibrillation had a heart rate of 65 beats/min, and four of the five participants with sinus rhythm had cardiac arrhythmias (atrial or ventricular premature beat in one participants, sinus tachycardia in one participant, and both arrhythmias in one participant).</p><p><strong>Conclusion: </strong>The OMRON HEM-7361T BP monitor is accurate in the differentiation between atrial fibrillation and sinus rhythm. Whether the device is sufficiently accurate in the differentiation between atrial fibrillation and other cardiac arrhythmias remains under investigation.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000706","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/20 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm.
Methods: An approximately equal number of patients with persistent atrial fibrillation and individuals with sinus rhythm were recruited from outpatients and inpatients of Ruijin Hospital, Shanghai, China. BP was measured three times consecutively with a 30-s interval with the OMRON HEM-7361T automatic electronic BP monitor for atrial fibrillation detection. A hand-held single lead electrocardiogram device was used for simultaneous electrocardiogram recordings.
Results: The device accurately identified atrial fibrillation in 100 (99.0%) of the 101 patients, with only 1 patient incorrectly classified as non-atrial fibrillation. The device correctly identified 99 (95.2%) of the 104 participants with sinus rhythm as non-atrial fibrillation, with five participants incorrectly classified as atrial fibrillation. The device had a positive predictive value of 95.2%, negative predictive value of 99.0%, and overall accuracy of 97.1%. Among the six misclassified participants, one with atrial fibrillation had a heart rate of 65 beats/min, and four of the five participants with sinus rhythm had cardiac arrhythmias (atrial or ventricular premature beat in one participants, sinus tachycardia in one participant, and both arrhythmias in one participant).
Conclusion: The OMRON HEM-7361T BP monitor is accurate in the differentiation between atrial fibrillation and sinus rhythm. Whether the device is sufficiently accurate in the differentiation between atrial fibrillation and other cardiac arrhythmias remains under investigation.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.