Desarrollo y validación de un procedimiento alternativo para realizar los controles de pureza radioquímica de radiofármacos de 99mTc con un contador Geiger Müller
IF 1.6 4区 医学Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
{"title":"Desarrollo y validación de un procedimiento alternativo para realizar los controles de pureza radioquímica de radiofármacos de 99mTc con un contador Geiger Müller","authors":"M.M. Gonzalez , A.L. Poma , N. Cuello , A.L. Soldati","doi":"10.1016/j.remn.2024.500006","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and objectives</h3><p>In a hospital radiopharmacy with 2a operational level, including the preparation of radiopharmaceuticals from prepared and approved reagent kits, it is common to have a single activimeter or dose calibrator for labeling and fractionation, and to perform the quality controls of the <sup>99m</sup>Tc-radiopharmaceuticals. In certain cases, the accumulation of radioactive material or accidental contamination of the work area causes the background to exceed the limits to carry out the radiochemical purity analyses and it is necessary to look for viable alternatives. In this work, a Geiger Müller detector (equipped with a probe for measuring surface contamination) frequently used for radioprotection purposes, was validated as an alternative and its performance was compared against the activimeter for <sup>99m</sup>Tc-radiopharmaceuticals.</p></div><div><h3>Materials and methods</h3><p>Using [<sup>99m</sup>Tc]pertechnetate, systematic studies of error analyses and detector response to activity concentration, activity and measurement time were carried out in liquid matrices and in paper. The results were compared against an activimeter calibrated for [<sup>99m</sup>Tc]Tc.</p></div><div><h3>Results</h3><p>The developed method was used to determine the radiochemical purity of the compounds [<sup>99m</sup>Tc]Tc-MDP and [<sup>99m</sup>Tc]Tc-MIBI by ascending paper chromatography tests, obtaining comparable values to those measured with an activimeter in the same system (within 1% uncertainty) and using the method of vial partitioning in a dedicated equipment.</p></div><div><h3>Conclusions</h3><p>This work demonstrates that a Geiger Müller detector with a probe for measuring surface contamination can be adequately used to replace other equipment in the control of radiochemical purity in the hospital radiopharmacy.</p></div>","PeriodicalId":48986,"journal":{"name":"Revista Espanola De Medicina Nuclear E Imagen Molecular","volume":"43 3","pages":"Article 500006"},"PeriodicalIF":1.6000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Espanola De Medicina Nuclear E Imagen Molecular","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2253654X24000106","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background and objectives
In a hospital radiopharmacy with 2a operational level, including the preparation of radiopharmaceuticals from prepared and approved reagent kits, it is common to have a single activimeter or dose calibrator for labeling and fractionation, and to perform the quality controls of the 99mTc-radiopharmaceuticals. In certain cases, the accumulation of radioactive material or accidental contamination of the work area causes the background to exceed the limits to carry out the radiochemical purity analyses and it is necessary to look for viable alternatives. In this work, a Geiger Müller detector (equipped with a probe for measuring surface contamination) frequently used for radioprotection purposes, was validated as an alternative and its performance was compared against the activimeter for 99mTc-radiopharmaceuticals.
Materials and methods
Using [99mTc]pertechnetate, systematic studies of error analyses and detector response to activity concentration, activity and measurement time were carried out in liquid matrices and in paper. The results were compared against an activimeter calibrated for [99mTc]Tc.
Results
The developed method was used to determine the radiochemical purity of the compounds [99mTc]Tc-MDP and [99mTc]Tc-MIBI by ascending paper chromatography tests, obtaining comparable values to those measured with an activimeter in the same system (within 1% uncertainty) and using the method of vial partitioning in a dedicated equipment.
Conclusions
This work demonstrates that a Geiger Müller detector with a probe for measuring surface contamination can be adequately used to replace other equipment in the control of radiochemical purity in the hospital radiopharmacy.