Within-participants reliability and measurement error of magnetization transfer imaging determinations within the healthy cervical spinal cord

IF 2.5 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
H. Al-shaari , C.J. Heales , J. Fulford
{"title":"Within-participants reliability and measurement error of magnetization transfer imaging determinations within the healthy cervical spinal cord","authors":"H. Al-shaari ,&nbsp;C.J. Heales ,&nbsp;J. Fulford","doi":"10.1016/j.radi.2024.04.027","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>To assess the within-participant reliability and measurement error in the determination of MTR in the healthy human cervical spinal cord.</p></div><div><h3>Methods and materials</h3><p>A total of twenty healthy controls (10 male, mean ± sd age: 33.9 ± 3.5 years, 10 females, mean ± sd age: 47.5 ± 14.4 years), with no family history of any neurological disorders or a contraindication to MRI scanning were recruited over a period of two months. Each participant was scanned twice with a 3T MRI scanner using standard MTI sequences. Spinal Cord Toolbox (v5.4) was used for image post-processing. Data were first segmented and then registered to a template and then MTR was computed. The within-participant coefficients of variation (CV%), single and average within-participants intraclass correlation coefficients (ICC) and Bland–Altman plots were determined for MT values over the volume between the 2nd and 5th cervical vertebrae for the total WM and for specific WM regions: dorsal column (DC), ventral column (VC) and lateral column (LC).</p></div><div><h3>Results</h3><p>MTR showed poor to excellent within-participant reliability for the total WM, DC, VC and LC with single/average ICC values of 0.03/0.06, 0.10/0.18, 0.39/0.75, and 0.001/0.002, respectively, and the CV% reported an acceptable variation with values less than 10%. The Bland–Altman plots showed good within-participant agreement between the scan-rescan values.</p></div><div><h3>Conclusion</h3><p>This study demonstrates that clinical trials using MTI technique are feasible and shows that quantitative MTI can monitor tissue changes in degenerative WM patients.</p></div><div><h3>Implications for practice</h3><p>MTI with its MTR index provide broad assessment of the integrity of white matter tissue and are being studied widely in brain as a diagnostic tool for the assessment of different neurological diseases.</p></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-05-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1078817424001056/pdfft?md5=9be60c292b80b94d2223dd6c511bb548&pid=1-s2.0-S1078817424001056-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424001056","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

To assess the within-participant reliability and measurement error in the determination of MTR in the healthy human cervical spinal cord.

Methods and materials

A total of twenty healthy controls (10 male, mean ± sd age: 33.9 ± 3.5 years, 10 females, mean ± sd age: 47.5 ± 14.4 years), with no family history of any neurological disorders or a contraindication to MRI scanning were recruited over a period of two months. Each participant was scanned twice with a 3T MRI scanner using standard MTI sequences. Spinal Cord Toolbox (v5.4) was used for image post-processing. Data were first segmented and then registered to a template and then MTR was computed. The within-participant coefficients of variation (CV%), single and average within-participants intraclass correlation coefficients (ICC) and Bland–Altman plots were determined for MT values over the volume between the 2nd and 5th cervical vertebrae for the total WM and for specific WM regions: dorsal column (DC), ventral column (VC) and lateral column (LC).

Results

MTR showed poor to excellent within-participant reliability for the total WM, DC, VC and LC with single/average ICC values of 0.03/0.06, 0.10/0.18, 0.39/0.75, and 0.001/0.002, respectively, and the CV% reported an acceptable variation with values less than 10%. The Bland–Altman plots showed good within-participant agreement between the scan-rescan values.

Conclusion

This study demonstrates that clinical trials using MTI technique are feasible and shows that quantitative MTI can monitor tissue changes in degenerative WM patients.

Implications for practice

MTI with its MTR index provide broad assessment of the integrity of white matter tissue and are being studied widely in brain as a diagnostic tool for the assessment of different neurological diseases.

健康颈脊髓内磁化传递成像测定的参与者内部可靠性和测量误差
方法和材料共招募了 20 名健康对照者(10 名男性,平均年龄(±sd):33.9±3.5 岁;10 名女性,平均年龄(±sd):47.5±14.4 岁),他们均无任何神经系统疾病家族史或核磁共振成像扫描禁忌症,为期两个月。每位受试者使用标准 MTI 序列在 3T 核磁共振成像扫描仪上扫描两次。脊髓工具箱(v5.4)用于图像后处理。首先对数据进行分割,然后将其注册到模板上,然后计算 MTR。对整个 WM 和特定 WM 区域(背侧柱 (DC)、腹侧柱 (VC) 和侧柱 (LC))的第 2 至第 5 颈椎之间体积的 MT 值确定了参与者内变异系数 (CV%)、单个和平均参与者类内相关系数 (ICC) 和 Bland-Altman 图。结果MTR显示,总WM、DC、VC和LC的参与者内部可靠性从差到优,单个/平均ICC值分别为0.03/0.06、0.10/0.18、0.39/0.75和0.001/0.002,CV%报告的变异值小于10%,可以接受。结论这项研究表明,使用 MTI 技术进行临床试验是可行的,并表明定量 MTI 可以监测退行性 WM 患者的组织变化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiography
Radiography RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍: Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信