{"title":"Is 3D-printed self-stabilizing endoprosthesis reconstruction without supplemental fixation following total sacrectomy a viable approach for sacral tumours?","authors":"Taojun Gong, Minxun Lu, Yitian Wang, Zhuangzhuang Li, Xuanhong He, Yi Luo, Yong Zhou, Chongqi Tu, Li Min","doi":"10.1007/s00586-024-08292-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The spinopelvic reconstruction poses significant challenges following total sacrectomy in patients with malignant or aggressive benign bone tumours encompassing the entire sacrum. In this study, we aim to assess the functional outcomes and complications of an integrated 3D-printed sacral endoprostheses featuring a self-stabilizing design, eliminating the requirement for supplemental fixation.</p><p><strong>Methods: </strong>We retrospectively analyzed patients with sacral tumours who underwent total sacrectomy followed by reconstruction with 3D-printed self-stabilizing endoprosthesis. Clinically, we evaluated functional outcomes using the 1993 version of the musculoskeletal tumour society (MSTS-93) score. Perioperative and postoperative complications were also documented.</p><p><strong>Results: </strong>10 patients met final inclusion criteria. The median age was 49 years (range, 31-64 years). The median follow-up time was 26.5 months (range, 15-47 months). Median postoperative functional MSTS-93 was 22.5 (range, 13-25). The median operation time was 399.5 min (305-576 min), and the median intraoperative blood loss was and 3200 ml (2400-7800 ml). Complications include wound dehiscence in one patient, bowel, bladder, and sexual dysfunction in four patients, cerebrospinal fluid leak in one patient, and tumour recurrence in one patient. There were no mechanical complications related to the endoprosthesis at the last follow-up.</p><p><strong>Conclusion: </strong>The utilization of 3D-printed self-stabilizing endoprosthesis proved to be a viable approach, yielding satisfactory short-term outcomes in patients undergoing total sacral reconstruction without supplemental fixation.</p>","PeriodicalId":12323,"journal":{"name":"European Spine Journal","volume":" ","pages":"4316-4324"},"PeriodicalIF":2.6000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Spine Journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00586-024-08292-9","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/7 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The spinopelvic reconstruction poses significant challenges following total sacrectomy in patients with malignant or aggressive benign bone tumours encompassing the entire sacrum. In this study, we aim to assess the functional outcomes and complications of an integrated 3D-printed sacral endoprostheses featuring a self-stabilizing design, eliminating the requirement for supplemental fixation.
Methods: We retrospectively analyzed patients with sacral tumours who underwent total sacrectomy followed by reconstruction with 3D-printed self-stabilizing endoprosthesis. Clinically, we evaluated functional outcomes using the 1993 version of the musculoskeletal tumour society (MSTS-93) score. Perioperative and postoperative complications were also documented.
Results: 10 patients met final inclusion criteria. The median age was 49 years (range, 31-64 years). The median follow-up time was 26.5 months (range, 15-47 months). Median postoperative functional MSTS-93 was 22.5 (range, 13-25). The median operation time was 399.5 min (305-576 min), and the median intraoperative blood loss was and 3200 ml (2400-7800 ml). Complications include wound dehiscence in one patient, bowel, bladder, and sexual dysfunction in four patients, cerebrospinal fluid leak in one patient, and tumour recurrence in one patient. There were no mechanical complications related to the endoprosthesis at the last follow-up.
Conclusion: The utilization of 3D-printed self-stabilizing endoprosthesis proved to be a viable approach, yielding satisfactory short-term outcomes in patients undergoing total sacral reconstruction without supplemental fixation.
期刊介绍:
"European Spine Journal" is a publication founded in response to the increasing trend toward specialization in spinal surgery and spinal pathology in general. The Journal is devoted to all spine related disciplines, including functional and surgical anatomy of the spine, biomechanics and pathophysiology, diagnostic procedures, and neurology, surgery and outcomes. The aim of "European Spine Journal" is to support the further development of highly innovative spine treatments including but not restricted to surgery and to provide an integrated and balanced view of diagnostic, research and treatment procedures as well as outcomes that will enhance effective collaboration among specialists worldwide. The “European Spine Journal” also participates in education by means of videos, interactive meetings and the endorsement of educative efforts.