Photon-Counting Detector CT: Potential for 75% Reduction in Contrast Medium Amount: A Phantom Study.

IF 0.6 4区 医学 Q4 MEDICINE, RESEARCH & EXPERIMENTAL
Fumiyo Higaki, Yusuke Morimitsu, Toshihiro Iguchi, Hayato Saito, Haruhiko Takaki, Ayako Nakagoshi, Maki Wada, Mayu Uka, Noriaki Akagi, Toshiharu Mitsuhashi, Yusuke Matsui, Takao Hiraki
{"title":"Photon-Counting Detector CT: Potential for 75% Reduction in Contrast Medium Amount: A Phantom Study.","authors":"Fumiyo Higaki, Yusuke Morimitsu, Toshihiro Iguchi, Hayato Saito, Haruhiko Takaki, Ayako Nakagoshi, Maki Wada, Mayu Uka, Noriaki Akagi, Toshiharu Mitsuhashi, Yusuke Matsui, Takao Hiraki","doi":"10.18926/AMO/66916","DOIUrl":null,"url":null,"abstract":"<p><p>This study aimed to evaluate the potential reduction in contrast medium utilization using photon-counting detector computed tomography (PCD-CT). One PCD-CT scan (CT1) and three conventional (non-PCD-CT) CT scans (CT2-CT4) were performed using a multi-energy CT phantom that contained eight rods with different iodine concentrations (0.2, 0.5, 1, 2, 5, 10, 15, and 20 mg/ml). The CT values of the seven groups (CT1 for 40, 50, 60, and 70 keV; and CT2-4) were measured. Noise and contrast-to-noise ratio (CNR) were assessed for the eight rods at various iodine concentrations. CT2 and CT1 (40 keV) respectively required 20 mg/ml and 5 mg/ml of iodine, indicating that a comparable contrast effect could be obtained with approximately one-fourth of the contrast medium amount. The standard deviation values increased at lower energy levels irrespective of the iodine concentration. The CNR exhibited a decreasing trend with lower iodine concentrations, while it remained relatively stable across all iodine levels (40-70 keV). This study demonstrated that virtual monochromatic 40 keV images offer a similar contrast effect with a reduced contrast medium amount when compared to conventional CT systems at 120 kV.</p>","PeriodicalId":7017,"journal":{"name":"Acta medica Okayama","volume":"78 2","pages":"135-142"},"PeriodicalIF":0.6000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta medica Okayama","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.18926/AMO/66916","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

Abstract

This study aimed to evaluate the potential reduction in contrast medium utilization using photon-counting detector computed tomography (PCD-CT). One PCD-CT scan (CT1) and three conventional (non-PCD-CT) CT scans (CT2-CT4) were performed using a multi-energy CT phantom that contained eight rods with different iodine concentrations (0.2, 0.5, 1, 2, 5, 10, 15, and 20 mg/ml). The CT values of the seven groups (CT1 for 40, 50, 60, and 70 keV; and CT2-4) were measured. Noise and contrast-to-noise ratio (CNR) were assessed for the eight rods at various iodine concentrations. CT2 and CT1 (40 keV) respectively required 20 mg/ml and 5 mg/ml of iodine, indicating that a comparable contrast effect could be obtained with approximately one-fourth of the contrast medium amount. The standard deviation values increased at lower energy levels irrespective of the iodine concentration. The CNR exhibited a decreasing trend with lower iodine concentrations, while it remained relatively stable across all iodine levels (40-70 keV). This study demonstrated that virtual monochromatic 40 keV images offer a similar contrast effect with a reduced contrast medium amount when compared to conventional CT systems at 120 kV.

光子计数探测器 CT:将造影剂用量减少 75% 的潜力:模型研究。
本研究旨在评估使用光子计数探测器计算机断层扫描(PCD-CT)减少造影剂使用量的可能性。研究人员使用一个多能量 CT 模型进行了一次 PCD-CT 扫描(CT1)和三次常规(非 PCD-CT)CT 扫描(CT2-CT4),该模型包含八根不同碘浓度(0.2、0.5、1、2、5、10、15 和 20 毫克/毫升)的碘棒。测量了七组(CT1 为 40、50、60 和 70 keV;CT2-4)的 CT 值。评估了不同碘浓度下八种棒材的噪声和对比度-噪声比(CNR)。CT2 和 CT1(40 千伏)分别需要 20 毫克/毫升和 5 毫克/毫升的碘,这表明只需约四分之一的造影剂量就能获得类似的造影效果。无论碘的浓度如何,标准偏差值在较低能量水平时都会增加。随着碘浓度的降低,CNR 呈下降趋势,而在所有碘浓度水平(40-70 千伏)下,CNR 保持相对稳定。这项研究表明,与 120 千伏的传统 CT 系统相比,40 千伏的虚拟单色图像在减少造影剂用量的情况下提供了相似的对比效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Acta medica Okayama
Acta medica Okayama 医学-医学:研究与实验
CiteScore
1.00
自引率
0.00%
发文量
110
审稿时长
6-12 weeks
期刊介绍: Acta Medica Okayama (AMO) publishes papers relating to all areas of basic and clinical medical science. Papers may be submitted by those not affiliated with Okayama University. Only original papers which have not been published or submitted elsewhere and timely review articles should be submitted. Original papers may be Full-length Articles or Short Communications. Case Reports are considered if they describe significant and substantial new findings. Preliminary observations are not accepted.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信