Study of reading errors when calibrating analog ohmmeters

IF 0.1 Q4 INSTRUMENTS & INSTRUMENTATION
Igor Zakharov, O. Botsiura, V. Semenikhin
{"title":"Study of reading errors when calibrating analog ohmmeters","authors":"Igor Zakharov, O. Botsiura, V. Semenikhin","doi":"10.24027/2306-7039.1.2024.300870","DOIUrl":null,"url":null,"abstract":"The features of calibration of analog ohmmeters are considered. Two measurement schemes for calibration were studied: using a multivalued standard measure, which makes it possible to set the ohmmeter readings to the calibrated scale mark, and by a direct measurement of the resistance value of a standard single-valued measure using a calibrated ohmmeter. It is shown that in the first case, the reading error includes two components: the error due to the phenomenon of parallax and the error in aligning the ohmmeter needle with the calibrated scale mark. In the second case, instead of the last component, it is necessary to take into account the interpolation error. Expressions for the uncertainty evaluation of corrections for all components of the reading error for linear and nonlinear ohmmeter scales are given. Formulas have been obtained that make it possible to calculate the measured resistance value in the event that the ohmmeter needle falls between the marks of a nonlinear scale.","PeriodicalId":40775,"journal":{"name":"Ukrainian Metrological Journal","volume":null,"pages":null},"PeriodicalIF":0.1000,"publicationDate":"2024-03-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ukrainian Metrological Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.24027/2306-7039.1.2024.300870","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

The features of calibration of analog ohmmeters are considered. Two measurement schemes for calibration were studied: using a multivalued standard measure, which makes it possible to set the ohmmeter readings to the calibrated scale mark, and by a direct measurement of the resistance value of a standard single-valued measure using a calibrated ohmmeter. It is shown that in the first case, the reading error includes two components: the error due to the phenomenon of parallax and the error in aligning the ohmmeter needle with the calibrated scale mark. In the second case, instead of the last component, it is necessary to take into account the interpolation error. Expressions for the uncertainty evaluation of corrections for all components of the reading error for linear and nonlinear ohmmeter scales are given. Formulas have been obtained that make it possible to calculate the measured resistance value in the event that the ohmmeter needle falls between the marks of a nonlinear scale.
研究校准模拟欧姆表时的读数误差
研究了模拟欧姆表校准的特点。研究了两种用于校准的测量方案:使用多值标准量具,这使得将欧姆表读数设置为校准刻度标记成为可能;以及使用校准欧姆表直接测量单值标准量具的电阻值。结果表明,在第一种情况下,读数误差包括两部分:视差现象造成的误差和欧姆表指针与校准刻度线对准时产生的误差。在第二种情况下,必须考虑内插误差,而不是最后一个部分。本文给出了线性和非线性欧姆表刻度读数误差所有分量修正的不确定度评估表达式。在欧姆表指针落在非线性刻度标记之间的情况下,可以通过公式计算出测量的电阻值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Ukrainian Metrological Journal
Ukrainian Metrological Journal INSTRUMENTS & INSTRUMENTATION-
自引率
0.00%
发文量
21
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信