Final report on EURAMET key comparison EURAMET.M.D-K2.1 (1522)

IF 2.1 3区 工程技术 Q2 INSTRUMENTS & INSTRUMENTATION
L Prochaska, J Rauch, A Furtado, F Benga, P Bartos, J Bebic, E Lenard, S Alisic, A Alic, P T Neuvonen, G S Sariyerli, Ü Y Akcadag
{"title":"Final report on EURAMET key comparison EURAMET.M.D-K2.1 (1522)","authors":"L Prochaska, J Rauch, A Furtado, F Benga, P Bartos, J Bebic, E Lenard, S Alisic, A Alic, P T Neuvonen, G S Sariyerli, Ü Y Akcadag","doi":"10.1088/0026-1394/61/1a/07006","DOIUrl":null,"url":null,"abstract":"<title>Main text</title>The key comparison EURAMET.M.D-K2.1 (1522) on the density of liquids determined by hydrostatic weighing was organized and carried out within the framework of the EMPIR project 17RPT02 \"rhoLiq\". Density measurements of liquids are mainly performed by laboratories calibrating or checking liquid density measuring instruments such as oscillation-type density meters. Another aim of this comparison was to establish traceability for liquid density measurements at emerging National Metrology Institutes (NMIs) to enable them to provide high-level measurement and calibration services for national stakeholders, e.g. food, chemical, pharmaceutical and petroleum industries. The results of this comparison can be used to claim calibration and measurement capabilities (CMCs).Most European emerging NMIs in this project had not yet participated in comparisons or obtained unsatisfactory results, or the uncertainty had a magnitude that was not fit for purpose.BEV (Austria) piloted this comparison, supported by PTB (Germany).Samples of deuterated water, tetrachloroethylene and oil of high viscosity were measured in the temperature range from 5 °C to 20 °C at atmospheric pressure by hydrostatic weighing for this comparison.The coronavirus pandemic caused delays in this key comparison's proposed initial schedule. Measurements were performed in 2021. The reference values of the key comparison were linked to CCM.D-K2 for PTB and EURAMET.M.D-K2 for BEV.To reach the main text of this paper, click on <ext-link xlink:href=\"https://www.bipm.org/documents/d/guest/euramet-m-d-k2-1\" xlink:type=\"simple\">Final Report</ext-link>. Note that this text is that which appears in Appendix B of the BIPM key comparison database <ext-link xlink:href=\"https://www.bipm.org/kcdb/\" xlink:type=\"simple\">https://www.bipm.org/kcdb/</ext-link>.The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"44 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-03-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/0026-1394/61/1a/07006","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

Main textThe key comparison EURAMET.M.D-K2.1 (1522) on the density of liquids determined by hydrostatic weighing was organized and carried out within the framework of the EMPIR project 17RPT02 "rhoLiq". Density measurements of liquids are mainly performed by laboratories calibrating or checking liquid density measuring instruments such as oscillation-type density meters. Another aim of this comparison was to establish traceability for liquid density measurements at emerging National Metrology Institutes (NMIs) to enable them to provide high-level measurement and calibration services for national stakeholders, e.g. food, chemical, pharmaceutical and petroleum industries. The results of this comparison can be used to claim calibration and measurement capabilities (CMCs).Most European emerging NMIs in this project had not yet participated in comparisons or obtained unsatisfactory results, or the uncertainty had a magnitude that was not fit for purpose.BEV (Austria) piloted this comparison, supported by PTB (Germany).Samples of deuterated water, tetrachloroethylene and oil of high viscosity were measured in the temperature range from 5 °C to 20 °C at atmospheric pressure by hydrostatic weighing for this comparison.The coronavirus pandemic caused delays in this key comparison's proposed initial schedule. Measurements were performed in 2021. The reference values of the key comparison were linked to CCM.D-K2 for PTB and EURAMET.M.D-K2 for BEV.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
EURAMET 关键比较的最终报告 EURAMET.M.D-K2.1 (1522)
正文 在 EMPIR 项目 17RPT02 "rhoLiq "的框架内,组织并开展了关于通过静力学称重确定液体密度的 EURAMET.M.D-K2.1 (1522) 关键比较。液体密度测量主要由校准或检查液体密度测量仪器(如振荡型密度计)的实验室进行。此次比对的另一个目的是在新兴的国家计量研究院(NMI)建立液体密度测量的溯源性,使其能够为国家利益相关方(如食品、化工、制药和石油行业)提供高水平的测量和校准服务。该项目的大多数欧洲新兴国家计量院尚未参加过比对,或获得的结果不令人满意,或不确定度的大小不符合目的。BEV(奥地利)在 PTB(德国)的支持下试行了这一比对。为进行此次比对,在 5 °C 至 20 °C 的温度范围内,在常压下通过静压称重法测量了氘化水、四氯乙烯和高粘度油的样品。测量于 2021 年进行。关键对比的参考值与 PTB 的 CCM.D-K2 和 BEV 的 EURAMET.M.D-K2 相关联。请注意,此文本为 BIPM 关键比较数据库附录 B 中的文本 https://www.bipm.org/kcdb/.The 根据 CIPM 互认安排 (CIPM MRA) 的规定,最终报告已通过同行评审,并批准由 CCM 发布。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Metrologia
Metrologia 工程技术-物理:应用
CiteScore
2.80
自引率
25.00%
发文量
137
审稿时长
12 months
期刊介绍: Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them. The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics. In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信