Setup time analysis for stereotactic body radiotherapy in O-ring linear accelerator without rotational correction.

IF 1.7 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Radiological Physics and Technology Pub Date : 2024-06-01 Epub Date: 2024-03-25 DOI:10.1007/s12194-024-00791-0
Biplab Sarkar, Anirudh Pradhan
{"title":"Setup time analysis for stereotactic body radiotherapy in O-ring linear accelerator without rotational correction.","authors":"Biplab Sarkar, Anirudh Pradhan","doi":"10.1007/s12194-024-00791-0","DOIUrl":null,"url":null,"abstract":"<p><p>This study analyse setup time (ST) and frequency of on-board imaging for stereotactic abdomen (liver, stomach), lung, and spine radiotherapy in the absence of automatic rotational correction. Total 53 stereotactic body radiotherapy (SBRT) patients, 28 of abdomen, 19 lung, and 6 spine treated for 230 sessions in O-ring gantry accelerator were evaluated for ST analysis. The mean setup time for all patients, abdomen, lung, and spine cases were 7.7 ± 7.4 min, 9.2 ± 9.2 min, 6.3 ± 4.1 min, and 5.5 ± 3.3 min, respectively. Median number CBCT was 2. 96% of cases had a CBCT between 1 and 3, and 9 (4%) had ≥ 4 CBCTs. Overall, 38.1%, 35.5%, 22.1%, 2.2%, and 2.2% of setup time fall into window of 0-5 min, 5-10 min, 10-20 min, 20-30 min, and > 30 min. Most difficult challenge is to negotiate with unknown rotational errors. It will be easy to dealt with them without automatic rotational correction if values are known.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-024-00791-0","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/25 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

This study analyse setup time (ST) and frequency of on-board imaging for stereotactic abdomen (liver, stomach), lung, and spine radiotherapy in the absence of automatic rotational correction. Total 53 stereotactic body radiotherapy (SBRT) patients, 28 of abdomen, 19 lung, and 6 spine treated for 230 sessions in O-ring gantry accelerator were evaluated for ST analysis. The mean setup time for all patients, abdomen, lung, and spine cases were 7.7 ± 7.4 min, 9.2 ± 9.2 min, 6.3 ± 4.1 min, and 5.5 ± 3.3 min, respectively. Median number CBCT was 2. 96% of cases had a CBCT between 1 and 3, and 9 (4%) had ≥ 4 CBCTs. Overall, 38.1%, 35.5%, 22.1%, 2.2%, and 2.2% of setup time fall into window of 0-5 min, 5-10 min, 10-20 min, 20-30 min, and > 30 min. Most difficult challenge is to negotiate with unknown rotational errors. It will be easy to dealt with them without automatic rotational correction if values are known.

无旋转校正的 O 型环直线加速器立体定向体放射治疗的设置时间分析。
这项研究分析了在没有自动旋转校正的情况下,腹部(肝脏、胃)、肺部和脊柱立体定向放射治疗的设置时间(ST)和机载成像频率。共有 53 名立体定向体放射治疗(SBRT)患者接受了 ST 分析评估,其中 28 名腹部患者、19 名肺部患者和 6 名脊柱患者在 O 型环龙门加速器上接受了 230 次治疗。所有患者、腹部、肺部和脊柱病例的平均设置时间分别为 7.7 ± 7.4 分钟、9.2 ± 9.2 分钟、6.3 ± 4.1 分钟和 5.5 ± 3.3 分钟。中位数 CBCT 为 2,96% 的病例 CBCT 在 1 到 3 之间,9 例(4%)病例 CBCT ≥ 4。总体而言,38.1%、35.5%、22.1%、2.2% 和 2.2% 的设置时间分别为 0-5 分钟、5-10 分钟、10-20 分钟、20-30 分钟和大于 30 分钟。最困难的挑战是处理未知的旋转误差。如果数值已知,则无需自动旋转校正即可轻松应对。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信