Factors Related to Proximal Junctional Kyphosis and Device Failure in Patients with Early-Onset Scoliosis Treated with a Traditional Dual Growing Rod: A Single Institution Study.

IF 2.3 Q2 ORTHOPEDICS
Asian Spine Journal Pub Date : 2024-04-01 Epub Date: 2024-03-08 DOI:10.31616/asj.2023.0286
Mohammadreza Chehrassan, Farshad Nikouei, Mohammadreza Shakeri, Javad Moeini, Behnam Jafari, Ebrahim Ameri Mahabadi, Hasan Ghandhari
{"title":"Factors Related to Proximal Junctional Kyphosis and Device Failure in Patients with Early-Onset Scoliosis Treated with a Traditional Dual Growing Rod: A Single Institution Study.","authors":"Mohammadreza Chehrassan, Farshad Nikouei, Mohammadreza Shakeri, Javad Moeini, Behnam Jafari, Ebrahim Ameri Mahabadi, Hasan Ghandhari","doi":"10.31616/asj.2023.0286","DOIUrl":null,"url":null,"abstract":"<p><strong>Study design: </strong>Observational study.</p><p><strong>Purpose: </strong>Investigation of factors related to proximal junctional kyphosis (PJK) and device failure in patients with early-onset scoliosis.</p><p><strong>Overview of literature: </strong>The use of growth-friendly devices, such as traditional dual growing rod (TDGR) for the treatment of earlyonset scoliosis (EOS), may be associated with important complications, including PJK and device failure.</p><p><strong>Methods: </strong>Thirty-five patients with EOS and treated with TDGR from 2014 to 2021 with a minimum follow-up of 2 years were retrospectively evaluated. Potential risk factors, including demographic factors, disease etiology, radiological measurements, and surgical characteristics, were assessed.</p><p><strong>Results: </strong>PJK was observed in 19 patients (54.3%), and seven patients (20%) had device failure. PJK was significantly associated with global final kyphosis change (p=0.012). No significant correlation was found between the rod angle contour, type of implant, connector design, and the risk of PJK or device failure.</p><p><strong>Conclusions: </strong>Treatment of EOS with TDGR is associated with high rates of complications, particularly PJK and device failure. The device type may not correlate with the risk of PJK and device failure. The progression of thoracic kyphosis during multiple distractions is an important risk factor for PJK.</p>","PeriodicalId":8555,"journal":{"name":"Asian Spine Journal","volume":" ","pages":"236-243"},"PeriodicalIF":2.3000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11065516/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Asian Spine Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31616/asj.2023.0286","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/8 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Study design: Observational study.

Purpose: Investigation of factors related to proximal junctional kyphosis (PJK) and device failure in patients with early-onset scoliosis.

Overview of literature: The use of growth-friendly devices, such as traditional dual growing rod (TDGR) for the treatment of earlyonset scoliosis (EOS), may be associated with important complications, including PJK and device failure.

Methods: Thirty-five patients with EOS and treated with TDGR from 2014 to 2021 with a minimum follow-up of 2 years were retrospectively evaluated. Potential risk factors, including demographic factors, disease etiology, radiological measurements, and surgical characteristics, were assessed.

Results: PJK was observed in 19 patients (54.3%), and seven patients (20%) had device failure. PJK was significantly associated with global final kyphosis change (p=0.012). No significant correlation was found between the rod angle contour, type of implant, connector design, and the risk of PJK or device failure.

Conclusions: Treatment of EOS with TDGR is associated with high rates of complications, particularly PJK and device failure. The device type may not correlate with the risk of PJK and device failure. The progression of thoracic kyphosis during multiple distractions is an important risk factor for PJK.

使用传统双生长棒治疗的早发性脊柱侧凸患者近端交界处后凸和装置失效的相关因素:单机构研究。
研究设计目的:调查与早发性脊柱侧凸患者近端交界性脊柱侧凸(PJK)和装置失效相关的因素:文献综述:使用有利于生长的器械(如传统双生长棒(TDGR))治疗早发脊柱侧凸(EOS)可能与重要的并发症有关,包括PJK和器械失效:对2014年至2021年期间接受TDGR治疗的35例EOS患者进行了回顾性评估,随访至少2年。评估了潜在的风险因素,包括人口统计学因素、疾病病因、放射学测量和手术特征:19名患者(54.3%)观察到PJK,7名患者(20%)出现设备故障。PJK与整体最终椎体后凸变化有明显相关性(P=0.012)。在杆角轮廓、植入物类型、连接器设计与 PJK 或装置失效风险之间没有发现明显的相关性:结论:使用 TDGR 治疗 EOS 的并发症发生率很高,尤其是 PJK 和装置故障。设备类型可能与 PJK 和设备故障的风险无关。多次牵引过程中胸椎后凸的发展是 PJK 的一个重要风险因素。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Asian Spine Journal
Asian Spine Journal ORTHOPEDICS-
CiteScore
5.10
自引率
4.30%
发文量
108
审稿时长
24 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信