Five days serum glucose stability at room-temperature in centrifuged fast-clotting serum tubes and the comparability with glucose in heparin-plasma and plasma containing citrate-stabilizer.

IF 1.3 4区 医学 Q4 MEDICINE, RESEARCH & EXPERIMENTAL
Hanne Bakkebø, Karen Lund Haaland, Kristin Solli Hoff, Lutz Schwettmann
{"title":"Five days serum glucose stability at room-temperature in centrifuged fast-clotting serum tubes and the comparability with glucose in heparin-plasma and plasma containing citrate-stabilizer.","authors":"Hanne Bakkebø, Karen Lund Haaland, Kristin Solli Hoff, Lutz Schwettmann","doi":"10.1080/00365513.2024.2318617","DOIUrl":null,"url":null,"abstract":"<p><p>Glucose measurement plays a central role in the diagnosis of gestational diabetes mellitus (GDM). Because of earlier reports of overestimation of glucose in the widely used tubes containing granulated glycolysis inhibitor, the study assessed the performance of fast-clotting serum tubes as an alternative sample for the measurement of glucose. Glucose concentration in fast-clotting serum was compared to lithium-heparin plasma placed in an ice-water slurry after sample collection and glucose stability at room-temperature was studied. Blood samples from 30 volunteers were drawn in four different types of tubes (serum separator tubes, fast-clotting serum tubes, lithium-heparin tubes and sodium fluoride, EDTA and a citrate buffer (NaF-EDTA-citrate) tubes, all from Greiner Bio-One). Lithium-heparin tubes were placed in an ice-water slurry until centrifugation in accordance with international recommendations and centrifuged within 10 min. After centrifugation, glucose was measured in all tubes (timepoint <i>T</i><sub>0</sub>) and after 24, 48, 72, 96 and 120 h of storage at 20-22 °C. NaF-EDTA-citrate plasma showed significant overestimation of glucose concentration by 4.7% compared to lithium-heparin plasma; fast-clotting serum showed glucose concentrations clinically equivalent to lithium-heparin plasma. In fast-clotting serum tubes, mean bias between glucose concentration after 24, 48, 72, 96 and 120 h and <i>T</i><sub>0</sub> was less than 2.4%. All individual differences compared to <i>T</i><sub>0</sub> were less than 6.5%. The results fulfill the acceptance criteria for sample stability based on biological variation. Fast-clotting serum tubes can be an alternative for the measurement of glucose in diagnosis and management of GDM and diabetes mellitus, especially when prolonged transportation is necessary.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"62-67"},"PeriodicalIF":1.3000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2024.2318617","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/7 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

Abstract

Glucose measurement plays a central role in the diagnosis of gestational diabetes mellitus (GDM). Because of earlier reports of overestimation of glucose in the widely used tubes containing granulated glycolysis inhibitor, the study assessed the performance of fast-clotting serum tubes as an alternative sample for the measurement of glucose. Glucose concentration in fast-clotting serum was compared to lithium-heparin plasma placed in an ice-water slurry after sample collection and glucose stability at room-temperature was studied. Blood samples from 30 volunteers were drawn in four different types of tubes (serum separator tubes, fast-clotting serum tubes, lithium-heparin tubes and sodium fluoride, EDTA and a citrate buffer (NaF-EDTA-citrate) tubes, all from Greiner Bio-One). Lithium-heparin tubes were placed in an ice-water slurry until centrifugation in accordance with international recommendations and centrifuged within 10 min. After centrifugation, glucose was measured in all tubes (timepoint T0) and after 24, 48, 72, 96 and 120 h of storage at 20-22 °C. NaF-EDTA-citrate plasma showed significant overestimation of glucose concentration by 4.7% compared to lithium-heparin plasma; fast-clotting serum showed glucose concentrations clinically equivalent to lithium-heparin plasma. In fast-clotting serum tubes, mean bias between glucose concentration after 24, 48, 72, 96 and 120 h and T0 was less than 2.4%. All individual differences compared to T0 were less than 6.5%. The results fulfill the acceptance criteria for sample stability based on biological variation. Fast-clotting serum tubes can be an alternative for the measurement of glucose in diagnosis and management of GDM and diabetes mellitus, especially when prolonged transportation is necessary.

离心快速凝固血清管中血清葡萄糖在室温下五天的稳定性,以及与肝素血浆和含柠檬酸盐稳定剂血浆中葡萄糖的可比性。
葡萄糖测量在妊娠糖尿病(GDM)的诊断中起着核心作用。由于早前有报告称,广泛使用的含有颗粒状糖酵解抑制剂的试管中的葡萄糖被高估,因此本研究评估了快速凝固血清试管作为葡萄糖测量替代样本的性能。将快速凝固血清中的葡萄糖浓度与样本采集后置于冰水浆中的锂肝素血浆进行了比较,并研究了葡萄糖在室温下的稳定性。30 名志愿者的血样被抽取到四种不同类型的试管中(血清分离试管、快速凝血血清试管、锂肝素试管以及氟化钠、乙二胺四乙酸和柠檬酸缓冲液(NaF-EDTA-柠檬酸)试管,均来自 Greiner Bio-One)。按照国际建议,将锂肝素试管置于冰水浆中直至离心,并在 10 分钟内离心。离心后,测量所有试管(时间点 T0)以及在 20-22 °C 下储存 24、48、72、96 和 120 小时后的葡萄糖。与锂肝素血浆相比,NaF-EDTA-柠檬酸钠血浆的葡萄糖浓度被明显高估了 4.7%;快速凝血血清的葡萄糖浓度在临床上与锂肝素血浆相当。在快速凝血血清试管中,24、48、72、96 和 120 小时后的葡萄糖浓度与 T0 之间的平均偏差小于 2.4%。与 T0 相比,所有个体差异均小于 6.5%。结果符合基于生物变异的样本稳定性验收标准。快速凝固血清管可作为诊断和管理 GDM 和糖尿病时测量葡萄糖的替代方法,尤其是在需要长时间运输的情况下。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.50
自引率
4.80%
发文量
85
审稿时长
4-8 weeks
期刊介绍: The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry. The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信