Sensitivity and specificity of the uniform field electroretinogram in glaucoma detection in comparison to the pattern electroretinogram.

IF 2.6 4区 医学 Q2 OPHTHALMOLOGY
Documenta Ophthalmologica Pub Date : 2024-04-01 Epub Date: 2024-01-20 DOI:10.1007/s10633-023-09960-9
Asma Hermas, Catherine Tsilfidis, Rustum Karanjia, Garfield Miller, Ange-Lynca Kantungane, Amanda El-Salibi, Irfan N Kherani, Shveta Bali, Stuart Coupland
{"title":"Sensitivity and specificity of the uniform field electroretinogram in glaucoma detection in comparison to the pattern electroretinogram.","authors":"Asma Hermas, Catherine Tsilfidis, Rustum Karanjia, Garfield Miller, Ange-Lynca Kantungane, Amanda El-Salibi, Irfan N Kherani, Shveta Bali, Stuart Coupland","doi":"10.1007/s10633-023-09960-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To determine the ability of the photopic negative response (PhNR) of the uniform field electroretinogram (UF-ERG) to identify early glaucomatous changes in comparison to the checkerboard and bar stimuli of the pattern electroretinogram (PERG).</p><p><strong>Methods: </strong>Forty-nine glaucoma patients were classified into two groups: glaucoma-suspect (23 eyes) and early to moderate glaucoma (30 eyes), based on their clinical examination and the results of standard automated perimetry. Thirty patients (30 eyes) with intraocular pressures (IOP) of 21 mmHg or less, with no history of reported high IOP, were included as controls. PERG and UF-ERG recordings were obtained on a Diagnosys D-341 Attaché-Envoy System. Visual field testing was done only for glaucoma-suspect and glaucoma patients.</p><p><strong>Results: </strong>All three tests (PERG bar stimulus, PERG checkerboard stimulus and PhNR) displayed significantly prolonged peak times for glaucoma and glaucoma-suspect patients, with delays ranging from 7.8 to 14.8%, depending on the test. The PERG bar stimulus also showed a significantly lower N95 amplitude for both glaucoma groups (with reductions of 26.0% and 33.0% for glaucoma-suspect and glaucoma groups, respectively). The PERG checkerboard N95 amplitude component had high sensitivity for detecting glaucoma patients but a low specificity (97% and 37%, respectively; AUC = 0.61). Overall, the PhNR peak time showed the highest sensitivity and specificity (77% and 90%, respectively; AUC = 0.87).</p><p><strong>Conclusions: </strong>PERG bar stimuli and the PhNR of the UF-ERG can be used in the clinical setting to detect glaucoma-related changes in glaucoma-suspect and glaucoma patients. However, our data confirm that the PhNR peak time has the best combined sensitivity and specificity.</p>","PeriodicalId":11207,"journal":{"name":"Documenta Ophthalmologica","volume":" ","pages":"97-106"},"PeriodicalIF":2.6000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Documenta Ophthalmologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10633-023-09960-9","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/20 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To determine the ability of the photopic negative response (PhNR) of the uniform field electroretinogram (UF-ERG) to identify early glaucomatous changes in comparison to the checkerboard and bar stimuli of the pattern electroretinogram (PERG).

Methods: Forty-nine glaucoma patients were classified into two groups: glaucoma-suspect (23 eyes) and early to moderate glaucoma (30 eyes), based on their clinical examination and the results of standard automated perimetry. Thirty patients (30 eyes) with intraocular pressures (IOP) of 21 mmHg or less, with no history of reported high IOP, were included as controls. PERG and UF-ERG recordings were obtained on a Diagnosys D-341 Attaché-Envoy System. Visual field testing was done only for glaucoma-suspect and glaucoma patients.

Results: All three tests (PERG bar stimulus, PERG checkerboard stimulus and PhNR) displayed significantly prolonged peak times for glaucoma and glaucoma-suspect patients, with delays ranging from 7.8 to 14.8%, depending on the test. The PERG bar stimulus also showed a significantly lower N95 amplitude for both glaucoma groups (with reductions of 26.0% and 33.0% for glaucoma-suspect and glaucoma groups, respectively). The PERG checkerboard N95 amplitude component had high sensitivity for detecting glaucoma patients but a low specificity (97% and 37%, respectively; AUC = 0.61). Overall, the PhNR peak time showed the highest sensitivity and specificity (77% and 90%, respectively; AUC = 0.87).

Conclusions: PERG bar stimuli and the PhNR of the UF-ERG can be used in the clinical setting to detect glaucoma-related changes in glaucoma-suspect and glaucoma patients. However, our data confirm that the PhNR peak time has the best combined sensitivity and specificity.

均匀视野视网膜电图与模式视网膜电图在青光眼检测中的灵敏度和特异性比较。
目的:与模式视网膜电图(PERG)的棋盘和条形刺激相比,确定均匀视野视网膜电图(UF-ERG)的光性负反应(PhNR)识别早期青光眼病变的能力:根据临床检查和标准自动视网膜测图的结果,将 49 名青光眼患者分为两组:青光眼疑似患者(23 眼)和早期至中度青光眼患者(30 眼)。对照组包括眼压(IOP)在 21 mmHg 或以下、无高眼压病史报告的 30 名患者(30 只眼)。通过 Diagnosys D-341 Attaché-Envoy 系统获得 PERG 和 UF-ERG 记录。仅对青光眼疑似患者和青光眼患者进行了视野测试:所有三项测试(PERG 条形刺激、PERG棋盘刺激和 PhNR)均显示青光眼和青光眼疑似患者的峰值时间明显延长,延迟时间从 7.8%到 14.8%不等,视测试而定。在 PERG 条形刺激中,两组青光眼患者的 N95 振幅也明显降低(青光眼疑似组和青光眼组分别降低了 26.0% 和 33.0%)。PERG 检查板 N95 振幅成分检测青光眼患者的灵敏度较高,但特异性较低(分别为 97% 和 37%;AUC = 0.61)。总体而言,PhNR 峰值时间显示出最高的灵敏度和特异性(分别为 77% 和 90%;AUC = 0.87):PERG 条形刺激和 UF-ERG 的 PhNR 可用于临床检测青光眼疑似患者和青光眼患者的青光眼相关变化。然而,我们的数据证实,PhNR 峰值时间具有最佳的综合灵敏度和特异性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Documenta Ophthalmologica
Documenta Ophthalmologica 医学-眼科学
CiteScore
3.50
自引率
21.40%
发文量
46
审稿时长
>12 weeks
期刊介绍: Documenta Ophthalmologica is an official publication of the International Society for Clinical Electrophysiology of Vision. The purpose of the journal is to promote the understanding and application of clinical electrophysiology of vision. Documenta Ophthalmologica will publish reviews, research articles, technical notes, brief reports and case studies which inform the readers about basic and clinical sciences related to visual electrodiagnosis and means to improve diagnosis and clinical management of patients using visual electrophysiology. Studies may involve animals or humans. In either case appropriate care must be taken to follow the Declaration of Helsinki for human subject or appropriate humane standards of animal care (e.g., the ARVO standards on Animal Care and Use).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信