{"title":"Application of the cell-based RT-qPCR assay (C-QPA) for potency detection of the novel trivalent rotavirus vaccine in China.","authors":"Yunjin Wang, YueYue Liu, Hong Bao, Yueru Chen, Guiying Kou, Mingqiang Wang, Shengfang Fu, Wen Huo, Wenzhu Guan, Yahui Cheng, Xu Zhou, Xiongxiong Li","doi":"10.1002/jcla.24989","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Because of the deficiencies of traditional methods in multivalent rotavirus vaccine potency detection, a cell-based quantitative RT-qPCR assay (C-QPA) was established and validated for specificity, precision, and accuracy.</p><p><strong>Methods: </strong>In order to further validate the robustness of this method in actual titer detection, the linear range and the practical application under different conditions were tested using monovalent and trivalent rotavirus samples and standards.</p><p><strong>Results: </strong>Results showed that the linear range was 2.0-6.5, 3.9-8.3, and 3.5-8.1 UI (unit of infectivity) for G2, G3, and G4, respectively. Besides, unknown sample with high titer exceeding the linear range can be calculated by dilution. The UIs of serotypes G2, G3, and G4 in monovalent and trivalent rotavirus samples showed a relative deviation ≤4.10%, and the monovalent samples of the same serotype with or without protective agents showed a relative deviation ≤4.28%; the coefficient of variation (CV) of at least 176 tests (548 individual runs) of 3 in vitro-transcribed RNA standards with certain concentrations was not higher than 6.50%; the results of the trivalent samples tested by more than 149 times in 5 years (467 individual runs) showed the CVs lower than 12.66%; 15 samples detected by one laboratory showed a CV lower than 9.83%, while other three samples tested by two independent laboratories showed a CV lower than 6.90%.</p><p><strong>Conclusion: </strong>In summary, the C-QPA has good linearity, durability, repeatability, and reproducibility in practical application and has been proved by the authority to be widely used in the production, quality control and release of the recently licensed trivalent vaccine in China.</p>","PeriodicalId":15509,"journal":{"name":"Journal of Clinical Laboratory Analysis","volume":" ","pages":"e24989"},"PeriodicalIF":2.6000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10756945/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Laboratory Analysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/jcla.24989","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/11/17 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Because of the deficiencies of traditional methods in multivalent rotavirus vaccine potency detection, a cell-based quantitative RT-qPCR assay (C-QPA) was established and validated for specificity, precision, and accuracy.
Methods: In order to further validate the robustness of this method in actual titer detection, the linear range and the practical application under different conditions were tested using monovalent and trivalent rotavirus samples and standards.
Results: Results showed that the linear range was 2.0-6.5, 3.9-8.3, and 3.5-8.1 UI (unit of infectivity) for G2, G3, and G4, respectively. Besides, unknown sample with high titer exceeding the linear range can be calculated by dilution. The UIs of serotypes G2, G3, and G4 in monovalent and trivalent rotavirus samples showed a relative deviation ≤4.10%, and the monovalent samples of the same serotype with or without protective agents showed a relative deviation ≤4.28%; the coefficient of variation (CV) of at least 176 tests (548 individual runs) of 3 in vitro-transcribed RNA standards with certain concentrations was not higher than 6.50%; the results of the trivalent samples tested by more than 149 times in 5 years (467 individual runs) showed the CVs lower than 12.66%; 15 samples detected by one laboratory showed a CV lower than 9.83%, while other three samples tested by two independent laboratories showed a CV lower than 6.90%.
Conclusion: In summary, the C-QPA has good linearity, durability, repeatability, and reproducibility in practical application and has been proved by the authority to be widely used in the production, quality control and release of the recently licensed trivalent vaccine in China.
期刊介绍:
Journal of Clinical Laboratory Analysis publishes original articles on newly developing modes of technology and laboratory assays, with emphasis on their application in current and future clinical laboratory testing. This includes reports from the following fields: immunochemistry and toxicology, hematology and hematopathology, immunopathology, molecular diagnostics, microbiology, genetic testing, immunohematology, and clinical chemistry.