Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography

Nivedita Rana, D. Rawat, Madan Parmar, D. Dhawan, A. K. Bhati, B. Mittal
{"title":"Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography","authors":"Nivedita Rana, D. Rawat, Madan Parmar, D. Dhawan, A. K. Bhati, B. Mittal","doi":"10.4103/0971-6203.170790","DOIUrl":null,"url":null,"abstract":"This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using CT (120 kV, 2.5 mA) component of SPECT/CT system, ensuring fixation of filter on X-ray collimator. Different thickness of filters of Al and Cu (1 mm, 2 mm, 3 mm, and 4 mm) and filter combinations Cu 1 mm, Cu 2 mm, Cu 3 mm each in combination with Al (1 mm, 2 mm, 3 mm, and 4 mm), respectively, were used. All image sets were visually analyzed for streak artifacts and contrast to noise ratio (CNR) was derived. Similar acquisition was done using Philips CT quality control (QC) phantom and CNR were calculated for its lexan, perspex, and teflon inserts. Attenuation corrected SPECT/CT images of Jaszack phantom filled with 444–555 MBq (12–15 mCi) of 99mTc were obtained by applying attenuation correction map generated by hardened X-ray beam for different filter combination, on SPECT data. Uniformity, root mean square (rms) and contrast were calculated in all image sets. Less streak artifacts at iodine water interface were observed in images acquired using external filters as compared to those without a filter. CNR for syringes, spheres, and inserts of Philips CT QC phantom was almost similar to Al 2 mm, Al 3 mm, and without the use of filters. CNR decreased with increasing copper thickness and other filter combinations. Uniformity and rms were lower, and value of contrast was higher for SPECT/CT images when CT was acquired with Al 2 mm and 3 mm filter than for images acquired without a filter. The study suggests that for Infinia Hawkeye 4, SPECT/CT system, Al 2 mm, and 3 mm are the optimum filters for improving image quality of SPECT/CT images of Jaszack or Philips CT QC phantom keeping other parameters of CT constant.","PeriodicalId":143694,"journal":{"name":"Journal of Medical Physics / Association of Medical Physicists of India","volume":"7 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics / Association of Medical Physicists of India","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/0971-6203.170790","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

Abstract

This study was undertaken to evaluate the effect of external metal filters on the image quality of computed tomography (CT) and single photon emission computed tomography (SPECT)/CT images. Images of Jaszack phantom filled with water and containing iodine contrast filled syringes were acquired using CT (120 kV, 2.5 mA) component of SPECT/CT system, ensuring fixation of filter on X-ray collimator. Different thickness of filters of Al and Cu (1 mm, 2 mm, 3 mm, and 4 mm) and filter combinations Cu 1 mm, Cu 2 mm, Cu 3 mm each in combination with Al (1 mm, 2 mm, 3 mm, and 4 mm), respectively, were used. All image sets were visually analyzed for streak artifacts and contrast to noise ratio (CNR) was derived. Similar acquisition was done using Philips CT quality control (QC) phantom and CNR were calculated for its lexan, perspex, and teflon inserts. Attenuation corrected SPECT/CT images of Jaszack phantom filled with 444–555 MBq (12–15 mCi) of 99mTc were obtained by applying attenuation correction map generated by hardened X-ray beam for different filter combination, on SPECT data. Uniformity, root mean square (rms) and contrast were calculated in all image sets. Less streak artifacts at iodine water interface were observed in images acquired using external filters as compared to those without a filter. CNR for syringes, spheres, and inserts of Philips CT QC phantom was almost similar to Al 2 mm, Al 3 mm, and without the use of filters. CNR decreased with increasing copper thickness and other filter combinations. Uniformity and rms were lower, and value of contrast was higher for SPECT/CT images when CT was acquired with Al 2 mm and 3 mm filter than for images acquired without a filter. The study suggests that for Infinia Hawkeye 4, SPECT/CT system, Al 2 mm, and 3 mm are the optimum filters for improving image quality of SPECT/CT images of Jaszack or Philips CT QC phantom keeping other parameters of CT constant.
外束硬化滤光片对计算机断层和单光子发射计算机断层/计算机断层成像质量的评价
本研究旨在评估外部金属滤光片对计算机断层扫描(CT)和单光子发射计算机断层扫描(SPECT)/CT图像质量的影响。使用SPECT/CT系统的CT (120 kV, 2.5 mA)组件获取充水和含碘造影剂注射器的Jaszack幻影图像,确保滤光片固定在x射线准直仪上。采用不同厚度的Al和Cu过滤器(1mm、2mm、3mm和4mm),以及分别与Al过滤器(1mm、2mm、3mm和4mm)组合的Cu 1 mm、Cu 2 mm、Cu 3 mm过滤器。对所有图像集进行了条纹伪影的视觉分析,并推导了噪声比(CNR)。使用飞利浦CT质量控制(QC)幻影进行类似的采集,并计算其lexan, perspex和teflon插入的CNR。将不同滤波组合下硬化x射线束生成的衰减校正图应用于SPECT数据,得到99mTc填充量为444 - 555mbq (12 - 15mci)的Jaszack幻影的衰减校正SPECT/CT图像。计算所有图像集的均匀性、均方根(rms)和对比度。与不使用滤光片的图像相比,使用外部滤光片获得的图像在碘水界面处观察到较少的条纹伪影。Philips CT QC幻影的注射器、球和插片的CNR几乎与Al 2 mm、Al 3 mm相似,且不使用过滤器。CNR随铜厚度和其他过滤器组合的增加而降低。使用Al 2mm和3mm滤波器获取的CT图像的均匀性和均方根值较低,对比度值高于未使用滤波器获取的图像。研究表明,对于Infinia Hawkeye 4, SPECT/CT系统,Al 2mm和3mm是提高Jaszack或Philips CT QC幻影的SPECT/CT图像质量的最佳滤波器,同时保持CT的其他参数不变。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信