{"title":"Testing 144- and 256-crystal BGO block detectors","authors":"J. Rogers, R. Nutt, M. Andreaco, C. Williams","doi":"10.1109/NSSMIC.1993.373611","DOIUrl":null,"url":null,"abstract":"New block detectors have been fabricated incorporating large numbers of small crystals. The authors evaluate and compare the performance of the new detectors with a standard 64-crystal block detector from Siemens-CTI. The new detectors demonstrate greatly improved imaging capability for 511 keV gamma rays. Future PET tomographs incorporating such detectors should produce substantially better volume images with little increase in tomograph manufacturing costs. The detectors will require a new type of automatic calibration procedure. Various such procedures have been tested and are discussed. A technique using higher energy gamma rays has shown special promise.<<ETX>>","PeriodicalId":287813,"journal":{"name":"1993 IEEE Conference Record Nuclear Science Symposium and Medical Imaging Conference","volume":"26 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1993-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"21","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1993 IEEE Conference Record Nuclear Science Symposium and Medical Imaging Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.1993.373611","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 21
Abstract
New block detectors have been fabricated incorporating large numbers of small crystals. The authors evaluate and compare the performance of the new detectors with a standard 64-crystal block detector from Siemens-CTI. The new detectors demonstrate greatly improved imaging capability for 511 keV gamma rays. Future PET tomographs incorporating such detectors should produce substantially better volume images with little increase in tomograph manufacturing costs. The detectors will require a new type of automatic calibration procedure. Various such procedures have been tested and are discussed. A technique using higher energy gamma rays has shown special promise.<>