Analyses with the KODAK-Ektachem. Accuracy control using reference method values and the influence of protein concentration. Part II. Substrates.

W R Külpmann, P Maibaum, O Sonntag, G Schumann, L Siekmann
{"title":"Analyses with the KODAK-Ektachem. Accuracy control using reference method values and the influence of protein concentration. Part II. Substrates.","authors":"W R Külpmann,&nbsp;P Maibaum,&nbsp;O Sonntag,&nbsp;G Schumann,&nbsp;L Siekmann","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The reliability of the determination of the most common substrates with the Ektachem 700 was evaluated. Accuracy control was performed in various ways including the comparison of results with reference method values. The influence of protein concentration was investigated systematically using sera containing varying amounts of protein obtained by ultracentrifugation. Bilirubin. Deviation from the reference method values was within limits of the guidelines (Dt. Arztebl. 85 (1988) B517-B532). The influence of protein concentration was negligible. Cholesterol. The results agreed well with the reference method values. The method was not influenced by different protein concentrations. Creatinine. The results were in good agreement with reference method values, when the method was calibrated with the primary assigned values of the calibrators. At high protein concentrations, the results were higher than those of the comparative method. Glucose. The mean deviation from the reference method values was 1.0%. There was a clear positive bias at high protein concentrations. Protein. The results from the Ektachem deviated from the reference method values by -6.8%. Total glycerol (triacylglycerols). In the analysis of control sera, the Ektachem values differed greatly (+32.1%) from the reference method values. This difference was not found in a comparative study with native sera. At high protein concentrations the Ektachem results were higher than those of the comparative method. Urea. The results were lower than the method-dependent assigned values (-16.6%). This deviation was not observed in a comparative study with native sera. Interference by protein was not observed. Uric acid. In accuracy control with reference method values, a small bias was observed (+3.3%), which increased at high protein concentrations.</p>","PeriodicalId":15649,"journal":{"name":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","volume":"28 11","pages":"835-43"},"PeriodicalIF":0.0000,"publicationDate":"1990-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The reliability of the determination of the most common substrates with the Ektachem 700 was evaluated. Accuracy control was performed in various ways including the comparison of results with reference method values. The influence of protein concentration was investigated systematically using sera containing varying amounts of protein obtained by ultracentrifugation. Bilirubin. Deviation from the reference method values was within limits of the guidelines (Dt. Arztebl. 85 (1988) B517-B532). The influence of protein concentration was negligible. Cholesterol. The results agreed well with the reference method values. The method was not influenced by different protein concentrations. Creatinine. The results were in good agreement with reference method values, when the method was calibrated with the primary assigned values of the calibrators. At high protein concentrations, the results were higher than those of the comparative method. Glucose. The mean deviation from the reference method values was 1.0%. There was a clear positive bias at high protein concentrations. Protein. The results from the Ektachem deviated from the reference method values by -6.8%. Total glycerol (triacylglycerols). In the analysis of control sera, the Ektachem values differed greatly (+32.1%) from the reference method values. This difference was not found in a comparative study with native sera. At high protein concentrations the Ektachem results were higher than those of the comparative method. Urea. The results were lower than the method-dependent assigned values (-16.6%). This deviation was not observed in a comparative study with native sera. Interference by protein was not observed. Uric acid. In accuracy control with reference method values, a small bias was observed (+3.3%), which increased at high protein concentrations.

用柯达- ektachem进行分析。准确度控制采用参考方法值及蛋白质浓度的影响。第二部分。基板。
对Ektachem 700测定最常见底物的可靠性进行了评价。通过各种方法进行精度控制,包括将结果与参考方法值进行比较。系统地研究了蛋白质浓度的影响,使用由超离心获得的含有不同数量蛋白质的血清。胆红素。与参考方法值的偏差在指南的限制范围内(Dt。中华医学杂志。85 (1988)B517-B532)。蛋白质浓度的影响可以忽略不计。胆固醇。结果与参考方法值吻合较好。该方法不受不同蛋白浓度的影响。肌酐。当用校准器的主要指定值对该方法进行校准时,结果与参考方法值吻合良好。在高蛋白质浓度下,结果高于对照法。葡萄糖。与参考方法值的平均偏差为1.0%。在高蛋白质浓度下存在明显的正偏倚。蛋白质。Ektachem的结果与参考方法值偏差-6.8%。总甘油(三酰甘油)。在对照血清分析中,Ektachem值与参考方法值相差较大(+32.1%)。在与天然血清的比较研究中没有发现这种差异。在高蛋白质浓度下,Ektachem结果高于比较方法。尿素。结果低于方法相关的赋值(-16.6%)。在与天然血清的比较研究中没有观察到这种偏差。未见蛋白干扰。尿酸。在参考方法值的准确度控制中,观察到较小的偏差(+3.3%),在高蛋白质浓度下增加。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信