[Accuracy of 2 electronic canal length measuring instruments].

F Ducoin
{"title":"[Accuracy of 2 electronic canal length measuring instruments].","authors":"F Ducoin","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Two electronic root canal length measuring instruments were used to measure the working length to \"the apex\" in 88 teeth scheduled for extraction. After extraction, the canal length was measured using X-Ray and microscopic observation after section of the teeth. The accuracy of the RCM apex locator in determining canal measurement within +/- 0.5 mm from the apical foramen varied around 62%. The \"NO'X\" allows a success rate of 41% in the same conditions. The conditions of this experiment shows that electronic root-canal length measuring instruments do not work in case of hemorrhage or swelling.</p>","PeriodicalId":77616,"journal":{"name":"Revue francaise d'endodontie : publication officielle de la Societe francaise d'endodontie","volume":"10 1","pages":"27-33"},"PeriodicalIF":0.0000,"publicationDate":"1991-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revue francaise d'endodontie : publication officielle de la Societe francaise d'endodontie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Two electronic root canal length measuring instruments were used to measure the working length to "the apex" in 88 teeth scheduled for extraction. After extraction, the canal length was measured using X-Ray and microscopic observation after section of the teeth. The accuracy of the RCM apex locator in determining canal measurement within +/- 0.5 mm from the apical foramen varied around 62%. The "NO'X" allows a success rate of 41% in the same conditions. The conditions of this experiment shows that electronic root-canal length measuring instruments do not work in case of hemorrhage or swelling.

[2种电子式管长测量仪的精度]。
采用两台电子式根管长度测量仪对88颗预备拔牙的根管工作长度进行测量。拔牙后x线测量根管长度,切面后显微镜观察根管长度。在离根尖孔+/- 0.5 mm范围内,RCM根尖定位仪测定根管测量的准确度约为62%。在相同的条件下,“NO”允许41%的成功率。本实验的条件表明,电子式根管长度测量仪在出血或肿胀的情况下不起作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信