R. Carta, D. Turgis, B. Hermans, P. Jourand, R. Onclin, R. Puers
{"title":"A Differential Pressure Approach to Spirometry","authors":"R. Carta, D. Turgis, B. Hermans, P. Jourand, R. Onclin, R. Puers","doi":"10.1109/BIOCAS.2007.4463295","DOIUrl":null,"url":null,"abstract":"This paper presents the design, implementation and measurements of a spirometer based on differential pressure sensing. A system which fulfills the last medical standard specifications has been designed exploiting the Venturi tube principle. A fully operating prototype has been tested, and data have been subsequently analyzed. Air flows up to 14 L/s can be measured with an accuracy of 0.2 L/s. Data acquisition software and user interface were developed.","PeriodicalId":273819,"journal":{"name":"2007 IEEE Biomedical Circuits and Systems Conference","volume":"155 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2007-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"20","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2007 IEEE Biomedical Circuits and Systems Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BIOCAS.2007.4463295","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 20
Abstract
This paper presents the design, implementation and measurements of a spirometer based on differential pressure sensing. A system which fulfills the last medical standard specifications has been designed exploiting the Venturi tube principle. A fully operating prototype has been tested, and data have been subsequently analyzed. Air flows up to 14 L/s can be measured with an accuracy of 0.2 L/s. Data acquisition software and user interface were developed.