{"title":"ANALISIS SUHU PADA ANALYZER INKUBATOR BAYI BERBASIS FORMULA MEAN","authors":"Kholid Al Sulaimi, Wisnu Kartika, Kuat Supriyadi","doi":"10.18196/MT.010101","DOIUrl":null,"url":null,"abstract":"Baby incubator is an equipment to keep the temperature stability of premature babies. Temperature is an important parameter. Therefore, its stability should be monitored based on the standard required by BPFK (Health Facilities Safety Center) hence the need for calibration. The writer aims to make a calibration tool for baby incubator temperature parameter with four measurement spots. The main sensor from the temperature parameter of incubator analyzer is LM35 as temperature sensor of T1, T2, T3, and T4. The range of temperature that can be detected is between 20.00 o C – 50.00 o C. The temperature inside the baby incubator has to be between 35.5 o C – 37 o C. The test was conducted in the baby incubator starting from 32 o C – 37 o C. The data was collected from the test point value on each temperature spot, the temperature comparison between TA modules used fluke temperature meter as comparison tool and the calibration of baby incubator. The output power for LM35 temperature sensor was done as much as ± 5 times. The biggest error value was found on T4 temperature sensor at ± 1.09762824% and at the temperature of 37 o C, while the smallest error value was found on T1 temperature sensor at ± 0.005633803% at the temperature of 35 o C. The measurement of baby incubator temperature calibration shows the margin of baby incubator temperature meter ± 1 o C at the temperature setting of 32 o C, 33 o C, and 34 o C.","PeriodicalId":441327,"journal":{"name":"Medika Teknika : Jurnal Teknik Elektromedik Indonesia","volume":"57 3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medika Teknika : Jurnal Teknik Elektromedik Indonesia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18196/MT.010101","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Baby incubator is an equipment to keep the temperature stability of premature babies. Temperature is an important parameter. Therefore, its stability should be monitored based on the standard required by BPFK (Health Facilities Safety Center) hence the need for calibration. The writer aims to make a calibration tool for baby incubator temperature parameter with four measurement spots. The main sensor from the temperature parameter of incubator analyzer is LM35 as temperature sensor of T1, T2, T3, and T4. The range of temperature that can be detected is between 20.00 o C – 50.00 o C. The temperature inside the baby incubator has to be between 35.5 o C – 37 o C. The test was conducted in the baby incubator starting from 32 o C – 37 o C. The data was collected from the test point value on each temperature spot, the temperature comparison between TA modules used fluke temperature meter as comparison tool and the calibration of baby incubator. The output power for LM35 temperature sensor was done as much as ± 5 times. The biggest error value was found on T4 temperature sensor at ± 1.09762824% and at the temperature of 37 o C, while the smallest error value was found on T1 temperature sensor at ± 0.005633803% at the temperature of 35 o C. The measurement of baby incubator temperature calibration shows the margin of baby incubator temperature meter ± 1 o C at the temperature setting of 32 o C, 33 o C, and 34 o C.
婴儿保温箱是一种保持早产儿体温稳定的设备。温度是一个重要的参数。因此,应根据BPFK(卫生设施安全中心)要求的标准监测其稳定性,因此需要进行校准。研制了一种具有四个测量点的婴儿恒温箱温度参数校准工具。培养箱分析仪温度参数的主要传感器为LM35,作为T1、T2、T3、T4的温度传感器。温度能被探测到的范围在20.00 o C - 50.00 o C .婴儿孵化器内的温度必须在35.5 o 37 C - o C测试是在婴儿培养箱进行从32 o 37 C - o C的数据收集测试点值在每个温度点,温度比较TA模块使用侥幸温度计比较工具和婴儿培养箱的校准。LM35温度传感器输出功率达到±5倍。T4温度传感器在37℃时误差最大,误差值为±1.09762824%,T1温度传感器在35℃时误差最小,误差值为±0.005633803%。婴儿恒温箱温度校准测量显示,在32℃、33℃、34℃的温度设置下,婴儿恒温箱温度计的裕度为±1℃。