Timebase Offset Calibration of Analog-To-Digital Converters for Mains Frequency Measurement

J. Kitzig, G. Bumiller
{"title":"Timebase Offset Calibration of Analog-To-Digital Converters for Mains Frequency Measurement","authors":"J. Kitzig, G. Bumiller","doi":"10.23919/SMAGRIMET48809.2020.9263998","DOIUrl":null,"url":null,"abstract":"The timebase inaccuracy of an Analog-to-Digital Converter (ADC) is an important systemic error to consider when it comes to fast and precise mains frequency measurement. In this paper, a new calibration method to determine the timebase offset of an ADC in real time to correct the mains frequency measurement is proposed and evaluated in tests. The methods uncertainty with regard to different calibration signal frequencies is assessed.","PeriodicalId":272673,"journal":{"name":"2020 3rd International Colloquium on Intelligent Grid Metrology (SMAGRIMET)","volume":"15 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2020 3rd International Colloquium on Intelligent Grid Metrology (SMAGRIMET)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.23919/SMAGRIMET48809.2020.9263998","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The timebase inaccuracy of an Analog-to-Digital Converter (ADC) is an important systemic error to consider when it comes to fast and precise mains frequency measurement. In this paper, a new calibration method to determine the timebase offset of an ADC in real time to correct the mains frequency measurement is proposed and evaluated in tests. The methods uncertainty with regard to different calibration signal frequencies is assessed.
用于主频测量的模数转换器时基偏置校准
模数转换器(ADC)的时基误差是快速、精确测量主频时需要考虑的一个重要系统误差。本文提出了一种实时确定模数转换器时基偏置以校正主频测量的校准方法,并在测试中进行了验证。评估了该方法在不同校准信号频率下的不确定度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信