Investigation of the applicability of a tensile testing machine for measuring mucoadhesive strength.

Acta pharmaceutica Nordica Pub Date : 1992-01-01
K Dyvik, C Graffner
{"title":"Investigation of the applicability of a tensile testing machine for measuring mucoadhesive strength.","authors":"K Dyvik,&nbsp;C Graffner","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The applicability of a tensile testing machine (M30K, JJ Lloyd Instruments Ltd, GB) is investigated for measuring mucoadhesive strengths. A sample of an aqueous dispersion of a polymer with expected mucoadhesive properties is placed between two homemade discs of polyoxymethylene. The upper disc is mounted on a movable part of the machine while the lower disc is fixed on the stationary frame. A tensile force is submitted and the maximum detachment force at fracture and the adhesion work are estimated from the force displacement curve recorded. In some experiments, native mucous tissue of the large intestine of pigs was glued to the upper disc. Four polymers polycarbophil (Carbopol EX-55), carboxypolymethylene (Carbopol 934P), hydroxypropylmethylcellulose (Methocel K4M), and sodium alginate, are used in five different concentrations. At least three measurements are made of each polymer and concentration. Viscosity and osmolality are determined. By standardizing the time of sample equilibration and the run rate before measurement, it is possible to get good reproducibility of the tensile values. Based on the maximum nominal breaking force and the work consumed, it is concluded that the tensile strength is dependent both on the concentration and the type of polymer. The conclusions are the same independent of whether mucous pig tissue is used, or not. The same rank order in adhesive properties of the polymers is achieved as from using modified surface tensiometers.</p>","PeriodicalId":7082,"journal":{"name":"Acta pharmaceutica Nordica","volume":"4 2","pages":"79-84"},"PeriodicalIF":0.0000,"publicationDate":"1992-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta pharmaceutica Nordica","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The applicability of a tensile testing machine (M30K, JJ Lloyd Instruments Ltd, GB) is investigated for measuring mucoadhesive strengths. A sample of an aqueous dispersion of a polymer with expected mucoadhesive properties is placed between two homemade discs of polyoxymethylene. The upper disc is mounted on a movable part of the machine while the lower disc is fixed on the stationary frame. A tensile force is submitted and the maximum detachment force at fracture and the adhesion work are estimated from the force displacement curve recorded. In some experiments, native mucous tissue of the large intestine of pigs was glued to the upper disc. Four polymers polycarbophil (Carbopol EX-55), carboxypolymethylene (Carbopol 934P), hydroxypropylmethylcellulose (Methocel K4M), and sodium alginate, are used in five different concentrations. At least three measurements are made of each polymer and concentration. Viscosity and osmolality are determined. By standardizing the time of sample equilibration and the run rate before measurement, it is possible to get good reproducibility of the tensile values. Based on the maximum nominal breaking force and the work consumed, it is concluded that the tensile strength is dependent both on the concentration and the type of polymer. The conclusions are the same independent of whether mucous pig tissue is used, or not. The same rank order in adhesive properties of the polymers is achieved as from using modified surface tensiometers.

粘接强度拉伸试验机的适用性研究。
研究了拉伸试验机(M30K, JJ Lloyd Instruments Ltd ., GB)在测定黏合剂强度方面的适用性。将具有预期粘接性能的聚合物的水分散体样品置于两个自制的聚甲醛圆盘之间。上盘安装在机器的可移动部分上,下盘固定在固定框架上。施加拉伸力,根据记录的力位移曲线估计断裂处的最大剥离力和附着功。在一些实验中,将猪大肠的天然粘液组织粘在上盘上。四种聚合物聚碳酸酯(Carbopol EX-55)、羧甲基(Carbopol 934P)、羟丙基甲基纤维素(Methocel K4M)和海藻酸钠以五种不同的浓度使用。对每种聚合物和浓度至少进行三次测量。测定粘度和渗透压。通过规范样品平衡时间和测量前的运行速率,可以获得良好的拉伸值再现性。根据最大名义断裂力和所消耗的功,得出拉伸强度取决于聚合物的浓度和类型的结论。结论与是否使用猪黏液组织无关。通过使用改良的表面张力计,聚合物的粘合性能达到了相同的等级顺序。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信