{"title":"The Usability Assessment of the Self-Produced Table for the Quality Control of Mobile CT","authors":"D. Jang, S. Choi, Dan-Bi Lee","doi":"10.31320/jksct.2022.24.1.1","DOIUrl":null,"url":null,"abstract":"Since there are no tables customized for mobile CT examinations, it is difficult to place the AAPM phantom to the isocenter during the quality control. This produced a phantom table for the quality control of Mobile CT, assessing its utility through conducting a quantitative evaluation through CNR value comparison between with and without the use of the table on low contrast resolution images among 20 phantom images evaluation, along with a qualitative assessment conducted on the scale of 1 to 5 by 5 quality control operators. \nThe obtained average CNR values from low contrast resolution images in the order of the table being used and not used are -0.01069 ± 0.00149 vs -0.01485 ± 0.00208, -0.01277 ± 0.00131 vs -0.01526 ± 0.00215, -0.01419 ± 0.00144 vs -0.01671 ± 0.00277, -0.01651 ± 0.00140 vs -0.01836 ± 0.00199, -0.01542 ± 0.0188 vs -0.01836 ± 0.00199 from A-row, -0.01132 ± 0.00134 vs -0.01300 ± 0.00222, -0.1202 ± 0.00129 vs -0.01369 ± 0.00145, -0.001391 ± 0.00163 vs -0.01460 ± 0.00139, -0.01607 ± 0.00185 vs -0.01683 ± 0.00182, -0.01502 ± 0.00170 vs -0.01656 ± 0.00125 from B-row, indicating higher absolute values of CNR when the table is being used than not being used, which corresponds to more suitable images. Furthermore, the average score of all items in qualitative assessment has increased from 3.64 to 4.24, suggesting statistically significant margin of increase especially on the items of efficiency and the quality of images. In conclusion, placing the phantom on a more accurate isocenter through the exploitation of the table produced through this research provides phantom images with higher quality along with the decrease of unnecessary efforts from the operators.","PeriodicalId":272693,"journal":{"name":"Korean Society of Computed Tomographic Technology","volume":"119 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-03-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Korean Society of Computed Tomographic Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31320/jksct.2022.24.1.1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Since there are no tables customized for mobile CT examinations, it is difficult to place the AAPM phantom to the isocenter during the quality control. This produced a phantom table for the quality control of Mobile CT, assessing its utility through conducting a quantitative evaluation through CNR value comparison between with and without the use of the table on low contrast resolution images among 20 phantom images evaluation, along with a qualitative assessment conducted on the scale of 1 to 5 by 5 quality control operators.
The obtained average CNR values from low contrast resolution images in the order of the table being used and not used are -0.01069 ± 0.00149 vs -0.01485 ± 0.00208, -0.01277 ± 0.00131 vs -0.01526 ± 0.00215, -0.01419 ± 0.00144 vs -0.01671 ± 0.00277, -0.01651 ± 0.00140 vs -0.01836 ± 0.00199, -0.01542 ± 0.0188 vs -0.01836 ± 0.00199 from A-row, -0.01132 ± 0.00134 vs -0.01300 ± 0.00222, -0.1202 ± 0.00129 vs -0.01369 ± 0.00145, -0.001391 ± 0.00163 vs -0.01460 ± 0.00139, -0.01607 ± 0.00185 vs -0.01683 ± 0.00182, -0.01502 ± 0.00170 vs -0.01656 ± 0.00125 from B-row, indicating higher absolute values of CNR when the table is being used than not being used, which corresponds to more suitable images. Furthermore, the average score of all items in qualitative assessment has increased from 3.64 to 4.24, suggesting statistically significant margin of increase especially on the items of efficiency and the quality of images. In conclusion, placing the phantom on a more accurate isocenter through the exploitation of the table produced through this research provides phantom images with higher quality along with the decrease of unnecessary efforts from the operators.
由于没有为移动CT检查定制的工作台,在质量控制中难以将AAPM假体放置到等中心。这就产生了一个用于Mobile CT质量控制的幻像表,通过对20个低对比度分辨率幻像评估中使用与未使用幻像表的CNR值进行定量评估,并由5个质量控制操作员以1到5的等级进行定性评估,从而评估其实用性。从低对比度分辨率图像中获得的平均CNR值按已使用和未使用的表格顺序依次为-0.01069±0.00149 vs -0.01485±0.00208、-0.01277±0.00131 vs -0.01526±0.00215、-0.01419±0.00144 vs -0.01671±0.00277、-0.01651±0.00140 vs -0.01836±0.00199、-0.01542±0.0188 vs -0.01836±0.00199、-0.01132±0.00134 vs -0.01300±0.00222、-0.1202±0.00129 vs -0.01369±0.00145、-0.001391±0.00163 vs -0.01460±0.00139、b排-0.01607±0.00185 vs -0.01683±0.00182,-0.01502±0.00170 vs -0.01656±0.00125,说明使用表时的CNR绝对值比不使用表时高,对应的图像更合适。此外,所有定性评估项目的平均得分从3.64分提高到4.24分,特别是在效率和图像质量方面的提高幅度具有统计学意义。综上所述,通过利用本研究产生的表,将幻像放置在更精确的等心位置,可以提供更高质量的幻像图像,同时减少了操作人员不必要的工作量。