Quantitative Assessment of Iodine Intensity of Different Types of Lesions in the Low-Energy (LE) Images of Contrast-Enhanced Spectral Mammography (CESM)

Sachila Niroshani, Tokiko Nakamura, N. Michiru, T. Negishi
{"title":"Quantitative Assessment of Iodine Intensity of Different Types of Lesions in the Low-Energy (LE) Images of Contrast-Enhanced Spectral Mammography (CESM)","authors":"Sachila Niroshani, Tokiko Nakamura, N. Michiru, T. Negishi","doi":"10.31487/j.rdi.2021.02.01","DOIUrl":null,"url":null,"abstract":"Purpose: To assess the iodine enhancement intensity of breast lesions in low energy (LE) images obtained in contrast-enhanced spectral mammography (CESM) with different tissue compositions.\nMaterials and Methods: A 50 mm dedicated phantom with different lesion insert and iodine insert were used to assess the enhancement intensity quantitatively. The target slab of the phantom consists of three lesions + iodine inserts together and 100% adipose equivalent,100% glandular equivalent inserts alone to mimic the adipose and glandular lesion without contrast-enhancement. Each iodine inserts having a concentration of 0.5 mgI/cm3, 1.0 mgI/cm3, 2.0 mgI/cm3. The phantom was exposed under semiautomated function at 28 kV, 30 kV, and 32 kV with Mo/Rh target/filter combination. Iodine intensity was estimated for three types of lesions at three breast equivalent compositions.\nResults: Lesions with fatty tissue had high intensity while lesions with glandular tissues had the minimum intensity. Among fatty lesions, highest mean intensity value (0.972±0.003) observed with minimum iodine concentration (F + 0.5 mgI/cm3). The highest mean intensity value (0.882±0.001) was found related to the glandular lesion with maximum iodine concentration (G + 2.0 mgI/cm3). The one-way ANOVA statistical test confirmed that mean intensity values were significantly varied among different lesions (P < 0.05).\nConclusion: LE images obtained in CESM can be used to identify the different types of lesions without performing the full field digital mammography (FFDM) as an additional examination prior to the CESM procedure.","PeriodicalId":308916,"journal":{"name":"Radiology and Medical Diagnostic Imaging","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-12-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiology and Medical Diagnostic Imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31487/j.rdi.2021.02.01","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To assess the iodine enhancement intensity of breast lesions in low energy (LE) images obtained in contrast-enhanced spectral mammography (CESM) with different tissue compositions. Materials and Methods: A 50 mm dedicated phantom with different lesion insert and iodine insert were used to assess the enhancement intensity quantitatively. The target slab of the phantom consists of three lesions + iodine inserts together and 100% adipose equivalent,100% glandular equivalent inserts alone to mimic the adipose and glandular lesion without contrast-enhancement. Each iodine inserts having a concentration of 0.5 mgI/cm3, 1.0 mgI/cm3, 2.0 mgI/cm3. The phantom was exposed under semiautomated function at 28 kV, 30 kV, and 32 kV with Mo/Rh target/filter combination. Iodine intensity was estimated for three types of lesions at three breast equivalent compositions. Results: Lesions with fatty tissue had high intensity while lesions with glandular tissues had the minimum intensity. Among fatty lesions, highest mean intensity value (0.972±0.003) observed with minimum iodine concentration (F + 0.5 mgI/cm3). The highest mean intensity value (0.882±0.001) was found related to the glandular lesion with maximum iodine concentration (G + 2.0 mgI/cm3). The one-way ANOVA statistical test confirmed that mean intensity values were significantly varied among different lesions (P < 0.05). Conclusion: LE images obtained in CESM can be used to identify the different types of lesions without performing the full field digital mammography (FFDM) as an additional examination prior to the CESM procedure.
乳腺造影(CESM)低能量(LE)影像中不同类型病变碘强度的定量评估
目的:评价不同组织组成的乳腺造影(CESM)低能量(LE)图像对乳腺病变的碘增强强度。材料与方法:采用50 mm专用假体,置入不同病变植入物和碘植入物,定量评估增强强度。幻体靶板由三个病变+碘植入物和100%脂肪当量、100%腺体当量植入物组成,模拟脂肪和腺体病变,无需增强。每个碘片的浓度分别为0.5毫克/立方厘米、1.0毫克/立方厘米和2.0毫克/立方厘米。模体在28 kV、30 kV和32 kV的Mo/Rh靶/滤波器组合下进行半自动化功能暴露。对三种乳腺等效成分中三种类型病变的碘强度进行了估计。结果:脂肪组织病变强度最高,腺体组织病变强度最低。在脂肪性病变中,碘浓度最低(F + 0.5 mgI/cm3)时,平均强度值最高(0.972±0.003)。最高平均强度值(0.882±0.001)与碘浓度最高(G + 2.0 mgI/cm3)的腺体病变有关。单因素方差分析证实,不同病变间平均强度值差异有统计学意义(P < 0.05)。结论:CESM中获得的LE图像可用于识别不同类型的病变,而无需在CESM手术前进行全视野数字乳房x线摄影(FFDM)作为额外的检查。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信