Pingki Setyowati Dewi, N. Ratini, N. L. Trisnawati
{"title":"Effect of x-ray tube voltage variation to value of contrast to noise ratio (CNR) on computed tomography (CT) Scan at RSUD Bali Mandara","authors":"Pingki Setyowati Dewi, N. Ratini, N. L. Trisnawati","doi":"10.53730/ijpse.v6n2.9656","DOIUrl":null,"url":null,"abstract":"Research has been conducted on the effect of variations in X-ray tube voltage to value of Contrast to Noise Ratio (CNR) on CT Scan at Bali Mandara Hospital using a phantom as a patient replacement. This research aims to determine the effect of X-ray tube voltage to the CNR value. Exposure factors used are X-ray tube voltage with variations of 80, 110, 120 and 135 kV, constant X-ray tube current of 150 mA and constant exposure time of 1 s. The readings of Io, Ib, and sb values in phantom images were performed using RadiAnt DICOM Viewer software (64 bit) and analysis of the effect X-ray tube voltage on CNR values was determined by regression test. The results of the analysis show that the variation of the X-ray tube voltage has a significant effect on the CNR value, where the greater X-ray tube voltage, the greater the CNR value. When the X-ray tube voltage is adjusted to 135 kV, the optimal CNR values are 113.52 for air, 35.06 for derlin, 13.93 for acrylic, 10.44 for nylon and 12.19 for polypropylene.","PeriodicalId":303507,"journal":{"name":"International journal of physical sciences and engineering","volume":"9 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"12","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of physical sciences and engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53730/ijpse.v6n2.9656","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 12
Abstract
Research has been conducted on the effect of variations in X-ray tube voltage to value of Contrast to Noise Ratio (CNR) on CT Scan at Bali Mandara Hospital using a phantom as a patient replacement. This research aims to determine the effect of X-ray tube voltage to the CNR value. Exposure factors used are X-ray tube voltage with variations of 80, 110, 120 and 135 kV, constant X-ray tube current of 150 mA and constant exposure time of 1 s. The readings of Io, Ib, and sb values in phantom images were performed using RadiAnt DICOM Viewer software (64 bit) and analysis of the effect X-ray tube voltage on CNR values was determined by regression test. The results of the analysis show that the variation of the X-ray tube voltage has a significant effect on the CNR value, where the greater X-ray tube voltage, the greater the CNR value. When the X-ray tube voltage is adjusted to 135 kV, the optimal CNR values are 113.52 for air, 35.06 for derlin, 13.93 for acrylic, 10.44 for nylon and 12.19 for polypropylene.