Mobile Clinical Gait Analysis Using Orient Specks

Smita S. Pochappan, D. Arvind, Jennifer Walsh, A. Richardson, J. Herman
{"title":"Mobile Clinical Gait Analysis Using Orient Specks","authors":"Smita S. Pochappan, D. Arvind, Jennifer Walsh, A. Richardson, J. Herman","doi":"10.1109/BSN.2012.34","DOIUrl":null,"url":null,"abstract":"This paper explores the use of Orient specks as on-body network of wireless inertial-magnetic sensors to capture the parameters of the human gait for mobile clinical gait analysis. A range of kinematic and temporal parameters were measured for normal humans using Orient specks and compared to values obtained from a commercial Vicon optical motion capture system. There was a good correlation between the joint angle data from the two systems, most notably in the sagittal plane. Hip flexion graphs showed the highest correlation value of 0.973, with knee flexion at 0.855 and pelvic rotation at 0.943, followed by pelvic obliquity at 0.689 and ankle flexion at 0.626. We conclude that the Orient specks have the potential for obtaining gait parameters outside the laboratory environment by measuring temporal parameters, and detecting the shape and trend of kinematic parameters of the patients when they are out and about during their everyday lives.","PeriodicalId":101720,"journal":{"name":"2012 Ninth International Conference on Wearable and Implantable Body Sensor Networks","volume":"4 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2012-05-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2012 Ninth International Conference on Wearable and Implantable Body Sensor Networks","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BSN.2012.34","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

Abstract

This paper explores the use of Orient specks as on-body network of wireless inertial-magnetic sensors to capture the parameters of the human gait for mobile clinical gait analysis. A range of kinematic and temporal parameters were measured for normal humans using Orient specks and compared to values obtained from a commercial Vicon optical motion capture system. There was a good correlation between the joint angle data from the two systems, most notably in the sagittal plane. Hip flexion graphs showed the highest correlation value of 0.973, with knee flexion at 0.855 and pelvic rotation at 0.943, followed by pelvic obliquity at 0.689 and ankle flexion at 0.626. We conclude that the Orient specks have the potential for obtaining gait parameters outside the laboratory environment by measuring temporal parameters, and detecting the shape and trend of kinematic parameters of the patients when they are out and about during their everyday lives.
移动临床步态分析的东方斑点
本文探索利用东方斑点作为无线惯性磁传感器的体上网络,捕捉人体步态参数,用于移动临床步态分析。使用Orient specks测量了正常人的一系列运动学和时间参数,并与商业Vicon光学运动捕捉系统获得的值进行了比较。两个系统的关节角度数据之间有很好的相关性,尤其是在矢状面。髋屈曲图相关值最高,为0.973,膝关节屈曲为0.855,骨盆旋转为0.943,其次是骨盆倾斜为0.689,踝关节屈曲为0.626。我们得出的结论是,东方斑点有潜力通过测量时间参数来获得实验室环境之外的步态参数,并检测患者在日常生活中外出走动时运动学参数的形状和趋势。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信