Measurement of deionized water density based on single silicon sphere

Jin-tao Wang, Jingyue Zhang, Xin-yu Ma, Lin Tong, X. Bao
{"title":"Measurement of deionized water density based on single silicon sphere","authors":"Jin-tao Wang, Jingyue Zhang, Xin-yu Ma, Lin Tong, X. Bao","doi":"10.1117/12.2512096","DOIUrl":null,"url":null,"abstract":"Due to perfect stability, deionized water is usually used as one important reference material for density and volume calibration. To evaluate suitability of pure water density formulas for deionized water, one measurement system based on Archimedes’s principle was designed, and silicon ring was used for density standard to measure the density of deionized at 20°C. The experimental result shows that The Tanaka equation can be used to calculate the density of deionized water with deviation 5ppm. The uncertainty budget of measurement is analyzed, and the combined relative uncertainty is 4.09ppm (k=2).","PeriodicalId":115119,"journal":{"name":"International Symposium on Precision Engineering Measurement and Instrumentation","volume":"109 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Symposium on Precision Engineering Measurement and Instrumentation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2512096","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Due to perfect stability, deionized water is usually used as one important reference material for density and volume calibration. To evaluate suitability of pure water density formulas for deionized water, one measurement system based on Archimedes’s principle was designed, and silicon ring was used for density standard to measure the density of deionized at 20°C. The experimental result shows that The Tanaka equation can be used to calculate the density of deionized water with deviation 5ppm. The uncertainty budget of measurement is analyzed, and the combined relative uncertainty is 4.09ppm (k=2).
基于单硅球的去离子水密度测量
由于去离子水具有良好的稳定性,通常被用作密度和体积标定的重要参考物质。为了评价纯水密度公式对去离子水的适用性,设计了一套基于阿基米德原理的测量系统,以硅环为密度标准,测量20℃下去离子水的密度。实验结果表明,Tanaka方程可用于计算去离子水的密度,误差为5ppm。分析了测量的不确定度预算,综合相对不确定度为4.09ppm (k=2)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信