Evaluation of Hemolysis Caused by a Miniature Heart Catheter Pump

C. Janeczek, M. Gföhler, M. Harasek, W. Mohl
{"title":"Evaluation of Hemolysis Caused by a Miniature Heart Catheter Pump","authors":"C. Janeczek, M. Gföhler, M. Harasek, W. Mohl","doi":"10.1109/BMEICON.2018.8609994","DOIUrl":null,"url":null,"abstract":"An important performance criterion for blood pumps is blood damage due to the non-physiological loads on the blood cells. Especially for miniature pumps that are inserted into blood vessels in catheter devices and – due to their small diameters - operate at high rotational speeds, hemolysis is an important design issue. The Assistocor heart catheter pump is a left ventricular assist device which is inserted into the Aorta and has an outer diameter of 6.8 mm, it operates at rotational speeds up to 40000 rpm. In-vitro tests with animal blood and different pump prototypes of the Assistocor heart pump showed that hemolysis values stayed below the recommended values for five out of nine of the tested prototypes and are comparable with results from literature, indicating that the design process is adequate to keep the damage of blood cells within acceptable limits despite the high rotational speeds.","PeriodicalId":232271,"journal":{"name":"2018 11th Biomedical Engineering International Conference (BMEiCON)","volume":"7 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2018 11th Biomedical Engineering International Conference (BMEiCON)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BMEICON.2018.8609994","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

An important performance criterion for blood pumps is blood damage due to the non-physiological loads on the blood cells. Especially for miniature pumps that are inserted into blood vessels in catheter devices and – due to their small diameters - operate at high rotational speeds, hemolysis is an important design issue. The Assistocor heart catheter pump is a left ventricular assist device which is inserted into the Aorta and has an outer diameter of 6.8 mm, it operates at rotational speeds up to 40000 rpm. In-vitro tests with animal blood and different pump prototypes of the Assistocor heart pump showed that hemolysis values stayed below the recommended values for five out of nine of the tested prototypes and are comparable with results from literature, indicating that the design process is adequate to keep the damage of blood cells within acceptable limits despite the high rotational speeds.
微型心脏导管泵致溶血的评价
血泵的一个重要性能指标是由非生理负荷引起的血液损伤。特别是对于导管装置中插入血管的微型泵,由于其直径小,在高转速下运行,溶血是一个重要的设计问题。Assistocor心脏导管泵是一种左心室辅助装置,插入主动脉,外径6.8毫米,转速可达40000转/分钟。用动物血液和不同的Assistocor心脏泵原型进行的体外测试表明,在9个测试原型中,有5个的溶血值低于推荐值,并且与文献结果相当,这表明设计过程足以使血细胞的损伤保持在可接受的范围内,尽管转速很高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信