{"title":"Determination of the Re-188 Calibration Number for the Capintec CRC- 25PET Dose Calibrator.","authors":"Mitchell Ashley Klenner, Aron Poole","doi":"10.2174/1874471016666230502140224","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>During the development of novel Re-188 radiopharmaceuticals, it was discovered that no calibration settings were published to calibrate Re-188 on the Capintec CRC-25PET dose calibrator.</p><p><strong>Methods: </strong>Sodium [<sup>188</sup>Re]perrhenate was eluted from an OncoBeta <sup>188</sup>W/<sup>188</sup>Re generator to measure activity on a Capintec CRC-25R dose calibrator using established dose calibrator settings provided by the manufacturer. The eluent was then used to tune the calibra on settings on a Capintec CRC-25PET dose calibrator, accounting for geometry. Radionuclidic purity of the [<sup>188</sup>Re]perrhenate source was verified via gamma spectroscopy.</p><p><strong>Results: </strong>The calibrator number for Re-188 was determined to be 469 x 10 for the Capintec CRC-25PET dose calibrator, which differed from the manufacturer provided calibra on number of 496 x 10 for the Capintec CRC-25R dose calibra on model. W-188 breakthrough was characterised as < 0.01%.</p><p><strong>Conclusion: </strong>This previously unreported calibration number can be used to determine the activity of Re- 188 labelled radiopharmaceuticals using the Capintec CRC-25PET dose calibrator model.</p>","PeriodicalId":10991,"journal":{"name":"Current radiopharmaceuticals","volume":"16 4","pages":"292-299"},"PeriodicalIF":1.5000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current radiopharmaceuticals","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2174/1874471016666230502140224","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: During the development of novel Re-188 radiopharmaceuticals, it was discovered that no calibration settings were published to calibrate Re-188 on the Capintec CRC-25PET dose calibrator.
Methods: Sodium [188Re]perrhenate was eluted from an OncoBeta 188W/188Re generator to measure activity on a Capintec CRC-25R dose calibrator using established dose calibrator settings provided by the manufacturer. The eluent was then used to tune the calibra on settings on a Capintec CRC-25PET dose calibrator, accounting for geometry. Radionuclidic purity of the [188Re]perrhenate source was verified via gamma spectroscopy.
Results: The calibrator number for Re-188 was determined to be 469 x 10 for the Capintec CRC-25PET dose calibrator, which differed from the manufacturer provided calibra on number of 496 x 10 for the Capintec CRC-25R dose calibra on model. W-188 breakthrough was characterised as < 0.01%.
Conclusion: This previously unreported calibration number can be used to determine the activity of Re- 188 labelled radiopharmaceuticals using the Capintec CRC-25PET dose calibrator model.