Vanessa Franziska Schmidt, Olaf Dietrich, Max Seidensticker, Moritz Wildgruber, Bernd Erber, Jens Ricke, Sophia Samira Goller
{"title":"Artifact characterization of Nitinol needles in magnetic resonance imaging-guided musculoskeletal interventions at 3.0 tesla: a phantom study.","authors":"Vanessa Franziska Schmidt, Olaf Dietrich, Max Seidensticker, Moritz Wildgruber, Bernd Erber, Jens Ricke, Sophia Samira Goller","doi":"10.4274/dir.2023.232262","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To characterize the artifacts of an 18-gauge coaxial nickel-titanium needle using a balanced steady-state free precession sequence in magnetic resonance imaging-guided interventions at 3.0 tesla.</p><p><strong>Methods: </strong>The influence of flip angle (FA), bandwidth, matrix, slice thickness (ST), and read-out direction on needle artifact behavior was investigated for different intervention angles (IA). Artifact diameters were rated at predefined positions. Subgroup differences were assessed using Bonferroni-corrected non-parametric tests and correlations between continuous variables were expressed using the Bravais-Pearson coefficient. Interrater reliability was quantified using intraclass correlation coefficients (ICCs), and a contrast-enhanced target lesion to non-enhanced muscle tissue contrast ratio was quantified.</p><p><strong>Results: </strong>The artifact diameters decreased with an increase in FA for all IAs (<i>P</i> < 0.001) and with an increase in ST for IAs of 45°-90° (all P < 0.05). Tip artifacts occurred at low IAs (0°-45°) and gradually increased in size with a decrease in IA (<i>P</i> = 0.022). The interrater reliability was high (ICC: 0.994-0.999). The contrast-enhanced target lesion to non-enhanced muscle tissue contrast ratio presented positive correlations with increasing FAs and matrices (<i>P</i> < 0.001; <i>P</i> = 0.003) and a negative correlation with increasing STs (<i>P</i> = 0.007).</p><p><strong>Conclusion: </strong>To minimize needle artifacts, it is recommended to use FAs of 40°-60°, a ST of >7 mm, and, if possible, an IA of 45°-60°. The visibility of the target lesion and the needle's artifact behavior must be weighed up against each other when choosing the ST, while higher FAs (40°-60°) and matrices (224 × 224/256 × 256) are associated with low artifacts and sufficient lesion visibility.</p>","PeriodicalId":11341,"journal":{"name":"Diagnostic and interventional radiology","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2023-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10679549/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostic and interventional radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4274/dir.2023.232262","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/8/18 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To characterize the artifacts of an 18-gauge coaxial nickel-titanium needle using a balanced steady-state free precession sequence in magnetic resonance imaging-guided interventions at 3.0 tesla.
Methods: The influence of flip angle (FA), bandwidth, matrix, slice thickness (ST), and read-out direction on needle artifact behavior was investigated for different intervention angles (IA). Artifact diameters were rated at predefined positions. Subgroup differences were assessed using Bonferroni-corrected non-parametric tests and correlations between continuous variables were expressed using the Bravais-Pearson coefficient. Interrater reliability was quantified using intraclass correlation coefficients (ICCs), and a contrast-enhanced target lesion to non-enhanced muscle tissue contrast ratio was quantified.
Results: The artifact diameters decreased with an increase in FA for all IAs (P < 0.001) and with an increase in ST for IAs of 45°-90° (all P < 0.05). Tip artifacts occurred at low IAs (0°-45°) and gradually increased in size with a decrease in IA (P = 0.022). The interrater reliability was high (ICC: 0.994-0.999). The contrast-enhanced target lesion to non-enhanced muscle tissue contrast ratio presented positive correlations with increasing FAs and matrices (P < 0.001; P = 0.003) and a negative correlation with increasing STs (P = 0.007).
Conclusion: To minimize needle artifacts, it is recommended to use FAs of 40°-60°, a ST of >7 mm, and, if possible, an IA of 45°-60°. The visibility of the target lesion and the needle's artifact behavior must be weighed up against each other when choosing the ST, while higher FAs (40°-60°) and matrices (224 × 224/256 × 256) are associated with low artifacts and sufficient lesion visibility.
期刊介绍:
Diagnostic and Interventional Radiology (Diagn Interv Radiol) is the open access, online-only official publication of Turkish Society of Radiology. It is published bimonthly and the journal’s publication language is English.
The journal is a medium for original articles, reviews, pictorial essays, technical notes related to all fields of diagnostic and interventional radiology.